Application technology
To adjust the
upper basket:
... upper position:
36
Upper basket height adjustment
Height-adjustable upper baskets can be adjusted between three
positions with 2 cm between each position to accommodate items of
different heights.
To adjust the height, the brackets with rollers on the sides of the
upper basket and the water connector at the back of the basket have
to be moved. The roller brackets are each secured to the upper
basket by two screws. The water connector consists of the following
components:
- A stainless steel plate with 2 openings,
- a plastic connection piece and
- 6 screws
Only adjust the upper basket horizontally. The baskets are not
designed to be positioned on a slant (one side up, one side down).
Altering the height will alter loading heights for both the upper and
lower baskets.
Remove the upper basket by pulling it out until a resistance is felt
and lifting it off the runners.
Unscrew the roller brackets and the water connector.
To adjust the upper basket to the ...
Move the roller brackets on both sides to the lowest position and
secure them firmly.
Position the stainless steel plate over the openings in the water
supply pipe so that the upper opening is covered. Secure the
stainless steel plate at the top with 2 screws. Place the water
connector in the lower opening of the stainless steel plate so that
the middle opening is covered. Secure the water connector with 4
screws.