Preparation For Storage & Shipment; Disconnecting The Calibrator; Packing Instructions - Fluke Calibration RUSKA 7615 User Manual

Hydraulic pressure controller/calibrator
Table of Contents

Advertisement

Preparation for Storage & Shipment

Disconnecting the Calibrator

The procedures given in this Chapter must be strictly adhered
to in order to prevent damage to the instrument. Failure to
follow these procedures will likely result in damage to the
Calibrator during shipment. This damage is not covered by the
carrier's insurance.
1. Relieve all hydraulic pneumatic pressure from the RUSKA 7615.
2. Turn the 7615 power switch off.
3. Disconnect the power cable from the RUSKA 7615's power receptacle.
4. Disconnect all pneumatic lines and fittings from the 7615 's back panel.
5. Plug all ports.

Packing Instructions

The instructions below must be strictly followed in order to prevent damage to the
instrument.
The main principle behind a successful shipment is that of minimizing shocks. This is
accomplished by cradling the device within two boxes such that the RUSKA 7615 is
restrained but still has resilience. The two most successful materials for this purpose are
rubber foam and flexible polyurethane foams.
Styrofoam, poured "foam in place" mixtures, and other rigid
foams are not suitable.
Even polyfoam or rubber foam should be cut into strips so that it will not present a large
rigid surface to the Calibrator.
Fluke has found that corrugated cardboard boxes provide the best packing. The boxes
sometimes arrive damaged, but the contents are usually intact. A minimum of 3 inches of
foam with at least an N95 impact rating should separate the inner surface of the inner box
and any portion of the Calibrator. The same is true for the inner and outer box.
W Caution
W Caution
Chapter 7
7-1

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the RUSKA 7615 and is the answer not in the manual?

Table of Contents