Fluke Calibration 1586A Super-DAQ Application Note

Automating temperature sensor calibration with the
Hide thumbs Also See for 1586A Super-DAQ:

Advertisement

Quick Links

Automating temperature
sensor calibration with the
Take your calibration lab
productivity to the next level
Use the 1586A Super-DAQ to automate
temperature sensor calibration
The 1586A Super-DAQ Precision Temperature
Scanner provides a unique feature for automat-
ing sensor calibration that can greatly increase
your lab productivity—without a PC and soft-
ware. When the Super-DAQ is connected to
a Fluke Calibration dry-well or fluid bath, it
can control the temperature source to run the
calibration automatically. You simply program
the number of set-point temperatures and their
values, select a scan sequence, assign a refer-
ence channel, and set the required stability
band.
The Super-DAQ monitors the temperature
source's stability through the reference chan-
nel, collects the data from the reference probe
and the "unit under test" (UUT) once stabi-
lized, and then advances to the next set-point
temperature.
After you configure and start the test, you
can walk away to work on other things. The
Super-DAQ just made your day a whole lot
easier!
One temperature source can be connected to
the Super-DAQ at a time. The Super-DAQ can
control all of the following Fluke Calibration
models:
• Field Metrology Wells: 9142, 9143, 9144,
9190A
• Metrology Wells: 9170, 9171, 9172, 9173
• Handheld Calibrators: 9100S, 9102S
• Field Dry-Well Calibrators: 3125, 9009, 9103,
9140, 9141
F r o m t h e F l u k e C a l i b r a t i o n D i g i t a l L i b r a r y @ w w w . f l u k e c a l . c o m / l i b r a r y
1586A Super-DAQ
• Thermocouple Furnace: 9150, 9118A
• Dual Block Dry-Well: 9011
• Micro Baths: 6102, 7102, 7103
• Compact Baths: 6330, 7320, 7340, 7380
• Deep-Well Compact Baths: 6331, 7321, 7341,
7381
• Triple Point of Water Maintenance Bath: 7312
• High Temperature Oil Baths: 6020, 6022,
6024
• High Temperature Salt Bath: 6050H
• Cold Temperature Baths: 7008, 7040, 7037,
7012, 7060, 7080
• Deep-Well Baths: 6054, 6055, 7007
• Resistor Baths: 7009, 7108, 7015
Application Note

Advertisement

loading

Summary of Contents for Fluke Calibration 1586A Super-DAQ

  • Page 1 PC and soft- ware. When the Super-DAQ is connected to a Fluke Calibration dry-well or fluid bath, it can control the temperature source to run the calibration automatically. You simply program...
  • Page 2 (Ref2). The sequence is then imme- diately repeated in reverse, descending order. For example: Ref1 > Ch102 > Ch103 > Ch104 > Ref2 > Ref2 > Ch104 > Ch103 > Ch102 > Ref1. 2 Fluke Calibration Automating temperature sensor calibration with the 1586A Super-DAQ...
  • Page 3 USB flash drive during a test. You can select Data, Graph, or Monitor view Note: The number and value of setpoint temperatures is dependent modes during data collection. on your source. 3 Fluke Calibration Automating temperature sensor calibration with the 1586A Super-DAQ...
  • Page 4 ©2014 Fluke Calibration. Specifications subject to change without notice. Printed in U.S.A. 6/2014 6002875A_EN Pub-ID 13173-eng Modification of this document is not permitted without written permission from Fluke Calibration. 4 Fluke Calibration Automating temperature sensor calibration with the 1586A Super-DAQ...