Page 2
LIMITED WARRANTY AND LIMITATION OF LIABILITY Each Fluke product is warranted to be free from defects in material and workmanship under normal use and service. The warranty period is one year and begins on the date of shipment. Parts, product repairs, and services are warranted for 90 days.
Page 7
List of Figures Figure Title Page Fuse Replacement ..................15 Battery Location ..................16 DC Voltage Test Connections ..............21 4-Wire Ohms Test Equipment Setup ............23 PRT Test Equipment Setup ............... 24 Thermistor Test Equipment Setup ............. 25 mA DC Current Equipment Setup ............. 26 Password Reset Jumper ................
See the Specifications section for information on the types and ranges of the measurement inputs the Product can accept. Contact Fluke Calibration To contact Fluke Calibration, call one of the following telephone numbers: • Technical Support USA: 1-877-355-3225 •...
1586A Calibration Manual Safety Information A Warning identifies conditions and procedures that are dangerous to the user. A Caution identifies conditions and procedures that can cause damage to the Product or the equipment under test. See Table 1 for a list of symbols used in this manual and on the Product. Table 1.
Page 11
SUPER-DAQ Precision Temperature Scanner Safety Information Warning To prevent possible electrical shock, fire, or personal injury: • Read all safety information before you use the Product. • Carefully read all instructions. • Use the Product only as specified, or the protection supplied by the Product can be compromised.
Page 12
1586A Calibration Manual • Do not touch voltages >30 V ac rms, 42 V ac peak, or 60 V dc. • Limit operation to the specified measurement category, voltage, or amperage ratings. • Measure a known voltage first to make sure that the Product operates correctly.
SUPER-DAQ Precision Temperature Scanner General Specifications General Specifications Mains Voltage 100 V Setting ............90 V to 110 V 120 V Setting ............108 V to 132 V 220 V Setting ............198 V to 242 V 240 V Setting ............216 V to 264 V Frequency ..............
1586A Calibration Manual USB Host Port Connector type ........... Type A Function .............. Memory File system ............FAT32 Memory capacity..........32 GB USB Device Port Connector type ........... Type B Class ..............Instrument Function .............. Control and data transfer Command protocol ..........SCPI Function ..............
SUPER-DAQ Precision Temperature Scanner Measurement Specifications PRT/RTD Resistance Accuracy Accuracy is given as % of measurement or ohms, whichever is greater. Basic accuracy is for 4-wire PRT/RTD. When using 3-wire PRT/RTD add 0.013 Ω to the accuracy specification for internal resistance mismatch and voltage offset if using Channel 1, or add 0.05 Ω...
1586A Calibration Manual Thermistor Temperature Range ..........−200 °C to 400 °C (depending on the sensor) Resistance Range ..........0 Ω to 1 MΩ Thermistor Resistance Accuracy Accuracy is given as ± (% of measurement + Ω). The basic accuracy specification is for 4-wire thermistor, slow sample rate.
SUPER-DAQ Precision Temperature Scanner Measurement Specifications Accuracy 10 kΩ Thermistor Temperature Slow Medium Fast 2-wire Sample Rate Sample Rate Sample Rate −40 °C 0.003 °C 0.004 °C 0.01 °C add 0.001 °C 0 °C 0.002 °C 0.002 °C 0.01 °C add 0.002 °C 25 °C 0.003 °C...
1586A Calibration Manual Thermocouple Temperature Accuracy Accuracy specifications apply using medium or slow sample rate. When using fast sample rate, increase the accuracy specification by 25 %. If the environment temperature is outside the specified range, increase the accuracy specification by 12 % for every 1 °C outside the specified environment temperature range.
SUPER-DAQ Precision Temperature Scanner Measurement Specifications Thermocouple Measurement Characteristics Temperature Display Resolution Range Slow / Medium Fast Sample Rate Sample Rate −270 °C to 2315 °C 0.01 °C 0.1 °C DC Voltage Maximum Input ............50 V on any range Common Mode Rejection ........
1586A Calibration Manual DC Current Accuracy Accuracy is given as ± (% measurement + % of range). Basic accuracy specification is for medium or slow sample rate. When using a fast sample rate, add the number given in the table to the accuracy specification. If the environment temperature is outside the specified range, multiply the temperature coefficient numbers by the temperature deviation and add to the accuracy specification.
SUPER-DAQ Precision Temperature Scanner 1586-2588 DAQ-STAQ Input Module Specifications Digital I/O Absolute Voltage Range ......... -4 V to 30 V Input Minimum Logic High ........2.0 V Input Maximum Logic Low ........0.7 V Output Type ............open drain active low Output Logic Low (<1 mA) ........
3. Make sure the power source that the Product is plugged into is energized. 4. Make sure that the line power fuse is good. See Replace the Line Power Fuse. If these steps do not solve the problem, then contact Fluke Calibration. See Contact Fluke Calibration. Replace the Line Power Fuse The Product has a fuse that protects from overcurrent.
Page 23
SUPER-DAQ Precision Temperature Scanner General Maintenance To replace the fuses, see Figure 1: 1. Remove any High-Capacity Modules or test leads from the Product where hazardous voltage may be present. 2. Disconnect the mains-power cord from the power-entry module. 3. Insert a small, flat screwdriver blade into the narrow recess to the left of the fuse holder and pry to the right until the holder pops out.
1586A Calibration Manual Battery Replacement The battery in the Product is used for Product memory. To replace the battery, see Figure 2: 1. Disassemble the Product as described in the Disassembly section. 2. Remove the battery and replace with one rated appropriately for the selected voltage.
SUPER-DAQ Precision Temperature Scanner General Maintenance Memory and Factory Reset The Product has two memory reset functions: • Clear all Files • Factory Reset See Table 3 for a comparison of these functions. Note All memory reset functions require the Admin password. Table 3.
1586A Calibration Manual Required Equipment Table 4 lists the equipment required for performance tests and calibration of the Product. Table 4. Required Test Equipment Function Instrument Type Model Comments Standard Fluke 5720A Used to characterized the 8½ digit meter Fluke 8508A 5522A Fluke 884X-SHORT, low Volts DC...
Page 27
SUPER-DAQ Precision Temperature Scanner Required Equipment Table 4. Required Test Equipment (cont.) Function Instrument Type Model Comments Must be characterized with Standard Fluke 5522A 8508A Used to characterized the Current DC 8½ digit meter Fluke 8508A 5522A Alternate standard Fluke 5720A Fluke 9171 with insert Metrology Drywell block...
1586A Calibration Manual Test Considerations For optimum performance, all test procedures should comply with these recommendations: • ) is stable and between 18 °C Assure the calibration ambient temperature (T and 28 °C. Ideally, the calibration should be done at 23 °C ±2 °C. •...
Table 5. Note For the 0 V tests, use the 4-wire short to short the Hi/Lo and Sense inputs. 5522A Calibrator 5520A CALIBRATOR 1586A SUPER-DAQ NORMAL SCOPE V, , ,RTD A, -SENSE, AUX V FUSED TRIG GUARD...
SUPER-DAQ Precision Temperature Scanner Performance Tests 4-Wire Ohms Verification Connect the Product to the test equipment as shown in Figure 4 and apply the resistance listed in Table 6. Note Ω For the 0 tests, use the 4-wire short to short the Hi/Lo and Sense inputs.
1586A Calibration Manual PRT Verification Connect the Product as in Figure 5 for the 1 kΩ value listed in Table 7. For all other values in the table, use standard resistors. Configure Ch001 as a 4-wire PRT function with resistance display. 5522A Calibrator 1 ke hcn312.eps...
SUPER-DAQ Precision Temperature Scanner Performance Tests Thermistor Verification Steps Connect the Product to the test equipment as shown in Figure 6 and apply the resistance listed in Table 8. Configure Ch001 as a 4-wire thermistor function with resistance display. 5522A Calibrator hcn204.eps Figure 6.
When use 5720A/5522A as source, connect Hi (source) to Lo (Product) and Lo (source) to mA (Product). For the 0 A tests, open Lo inputs. 5522A Calibrator 5520A CALIBR ATOR 1586A SUPER-DAQ NORMAL SCOPE V, , , RTD AUX V A, -SENSE,...
SUPER-DAQ Precision Temperature Scanner Performance Tests Table 9. DC Current Verification Steps 1-Year Test Limits Nominal Range Input (mA) High 3.5 nA -3.5 nA 100.0185 μA 99.9815 μA -99.9815 μA -100.0185 μA -0.1 11.0 nA -11.0 nA 1.000161 mA 0.999839 mA -0.999839 mA -1.000161 mA 0.35 μA...
1586A Calibration Manual 1586-2588 Module CJC Accuracy Verification 1. Connect 10 calibrated E-type thermocouples with TC adapors to channels 1 through 10 on the Module TC input connectors. 2. Insert all thermocouples into a drywell calibrator which is set and stabilized at 25 °C.
SUPER-DAQ Precision Temperature Scanner Calibration Adjustment Unlock the Product with a Remote Interface To unlock the Product with a remote interface, send the command: CAL:SEC:STAT OFF,“1586” To relock the Product, send the command: CAL:SEC:STAT ON Reset the Admin Password If the admin password is lost or forgotten, the password can be reset to 1586 with these steps: Note Before doing these steps, try to use the factory default password:...
Page 38
1586A Calibration Manual Short J8, then turn on the power. After the instrument is successfully started, the admin password is reset to 1586. Display Battery Cable RST PSW InGuard Shield Assembly hcn103.eps Figure 8. Password Reset Jumper...
SUPER-DAQ Precision Temperature Scanner Calibration Adjustment Disassembly Warning To prevent possible electrical shock or personal injury, disconnect the mains power cord before you remove the Product covers. Only disassemble the Product to reset the password or replace the battery. See Figure 9.
Page 40
1586A Calibration Manual Boot Removal To Remove, Pull End From Unit Battery hcn208.eps Figure 9. Disassembly...
SUPER-DAQ Precision Temperature Scanner Calibration Adjustment Change the Calibration Date Remotely The calibration date is automatically updated when the “CALibrate:STORe” command is sent. Equipment for “Mainframe” Calibration The required equipment for calibration is listed in Table 4. “Mainframe” Adjustment Process Table 10 lists the steps to adjust the “mainframe”.
Page 42
1586A Calibration Manual Table10. Adjustment Steps (cont.) Input Entry Step Value Range Description Signal Point Start calibration of FRES R100 100.0 100 Ω range 100.0 Ω R_100_1 19.0 Ω R_19_1 0.0 Ω R_0_1 Start calibration of FRES 1 1.0E+3 kΩ range R_1K_2 1.0 kΩ...
Page 43
SUPER-DAQ Precision Temperature Scanner Calibration Adjustment Table 10. Adjustment Steps (cont.) Input Entry Step Value Range Description Signal Point Start calibration of RES R100M 100.0E+6 100 MΩ range R_100M_7 100.0 MΩ R_19M_7 19.0 MΩ 0.0 Ω R_0_7 Start calibration of PRT P400FR 400.0 400 Ω...
Page 44
1586A Calibration Manual Table 10. Adjustment Steps (cont.) Input Entry Step Value Range Description Signal Point Start calibration of dc DC100UA 100.0E-6 100 μA range 100.0 μA DC_P100UA_1 DC_0A_1 Open -100.0 μA DC_N100UA_1 Start calibration of dc 1 mA DC1MA 1.0E-3 range DC_P1_0MA_2...
SUPER-DAQ Precision Temperature Scanner Calibration Adjustment Remote Commands for Calibration Table 11 alphabetically lists the command set for calibration. Table 11. List of Commands Remote Command Meaning Instruct Product to abort calibration procedure after CALibrate:ABORt present step. Backup to previous entry point in calibration CALibrate:BACKup procedure.
1586A Calibration Manual Remote Programming Examples This section gives examples of command sequences for several likely scenarios. Start a Full Calibration Table 12 shows an example for how to run a full calibration. Table 12. Full Calibration Example Command Action Disable the security for calibration.
SUPER-DAQ Precision Temperature Scanner Calibration Adjustment Calibrate 1 V DC Range only Table 13 shows an example to run a calibration for 1 V dc. Table 13. 1 V DC Calibration Example Command Action Disable the security for calibration. CAL:SEC:STAT OFF,”1586” Start MAIN calibration procedure, and jump to 1 V dc range directly, it will show ask for 1 V signal input, and the reference CAL:STAR MAIN,DC1V...
1586A Calibration Manual Command References CALibrate:ABORt Description: Instruct unit to abort calibration procedure after present step. Example: CAL:ABOR Related Commands: CALibrate:STATe? CALibrate:BACKup Description: Backup to previous entry point in calibration procedure. Example: CAL:BACK Related Commands: CALibrate:SKIP CALibrate:SECTion CALibrate:CONSt? “<cco_name>” Description: Retrieves the value in use of the given calibration constant.
Page 49
SUPER-DAQ Precision Temperature Scanner Calibration Adjustment CALibrate:INFOrmation? Description: Return message or instructions associated with running step Example: CAL:INFO? Response: “Connect calibrator to Volt terminals” Related Commands: CALibrate:STATe? CALibrate:STEP? CALibrate:MODule:DATE <slot>,<year>,<month>,<day> CALibrate:MODule:DATE? <slot> Description: Set and query module calibration date. Example: CAL:MOD:DATE 1,2013,11,1 CAL:MOD:DATE? 1 Response:...
Page 50
1586A Calibration Manual CALibrate:REFerence? Description: Return nominal value expected for reference entry Example: CAL:REF? Response: 3.000000e+00 Related Commands: CALibrate:NEXT CALibrate:STATe? CALibrate:INFOrmation? CALibrate:STEP? CALibrate:SECure:STATe <boolean>, <admin_password> Description: Instruct Product to enable calibration Example: CAL:SEC:STATE OFF,”1586” Related Commands: CALibrate:SECure:STATe? CALibrate:SECure:STATe? Description: Query calibration enable state Example: CAL:SEC:STAT? Response:...
Page 51
SUPER-DAQ Precision Temperature Scanner Calibration Adjustment CALibrate:STARt <procedure>[,<step>] Description: Start a calibration procedure. As parameter, the name of procedure should be provided (MAIN is the procedure for full instrument calibration), an optional parameter <step> can be provided to start from, if it is omitted, it starts at the beginning.
Page 52
1586A Calibration Manual CALibrate:STORe CALibrate:STORe? Description: Store new calibration constants or query whether a cal store is needed. Example: CAL:STOR CAL:STOR? Response: 1 is yes, 0 if no Related Commands: CALibrate:SECure:STATe...
SUPER-DAQ Precision Temperature Scanner User-Replaceable Parts and Accessories User-Replaceable Parts and Accessories Table 15 lists the part numbers of each user-replaceable part or accessory for the Product, see Figure 10. Table 15. User-Replaceable Parts and Accessories Fluke Part Item Quantity Description Number ...