Technical Specifications - ResMed AirFit P10 User Manual

Hide thumbs Also See for AirFit P10:
Table of Contents

Advertisement

Available languages

Available languages

Technical specifications

Pressure-flow curve
Dead space information
Therapy pressure
Resistance
Sound
Environmental
conditions
Gross dimensions
The mask contains passive venting to protect against
rebreathing. As a result of manufacturing variations, the vent
flow rate may vary.
Physical dead space is the empty volume of the mask to the
end of the swivel. Using the large cushions it is 123 mL.
4 to 20 cm H
O
2
Drop in pressure measured (nominal)
at 50 L/min: 0.4 cm H
at 100 L/min: 1.4 cm H
DECLARED DUAL-NUMBER NOISE EMISSION VALUES in
accordance with ISO 4871. The A-weighted sound power
level of the mask is 21 dBA, with uncertainty of 3 dBA. The
A-weighted sound pressure level of the mask at a distance
of 1 m is 13 dBA, with uncertainty of 3 dBA.
Operating temperature: +41°F to 104°F (+5°C to +40°C)
Operating humidity: 15% to 95% non-condensing
Storage and transport temperature: -4°F to 140°F (-20°C to
+60°C)
Storage and transport humidity: up to 95% relative humidity
non-condensing
Large:
23.23" (H) x 5.87" (W) x 1.18" (D)
(590 mm (H) x 149 mm (W) x 30 mm (D))
Mask fully assembled with short tube assembly (no
headgear)
O
2
O
2
Pressure
Flow
(cm H2O)
(L/min)
4
20
8
29
12
37
16
43
20
49