Supplemental Information
- Life Expectancy
- Temperature
- Pressure
- Vibration
- Corrosion
- Fire and Explosion
- Humidity
Licensing
Testing
- Leak Test
- Shutter Test
Transfer and Disposal
Handling, Installation,
and Storage
Storage
Physical Inspection
42
20 years
- 40 °C ... + 60 °C (- 40 °F ... + 140 °F)
Atmospheric
Mild (less than 5 g)
Mild
Accident Conditions Only
0% to 100% Relative Humidity
A general license for the use of this source holder
must meet the requirements as outlined in OAC 3701:1-
46-05 (10 CFR 31.5 equivalent). Any other use of the
source holder requires a specific license.
At least once every 36 months
At least once every 6 months
The transfer of this source holder containing the sealed source
is only to a specifically licensed person or as specified in OAC
3701:1-46 (an equivalent to 10 CFR 31 and 32). If you need
to transfer the source holder to another location or dispose of
the source holder, specific licensees are required.
A general licensee may initially mount the device. You may not
put the source holder into service until it is inspected by a
specific licensee. The manufacturer, or other specifically
licensed persons, are the only people that may perform service.
Whenever you remove the device from its mounted
position, mount the cover plate on the device to shield
the beam port and keep foreign material out of the
shutter assembly.
Perform physical inspections of the device and its
labeling at least once every six months for corrosion
prevention and maintenance in accordance with the
manufacturer's instructions.