Specifications
Gas:
Coolant:
Sample Tube:
Internal Components:
Temperature:
Humidity:
Dimensions:
Weight:
FlowSorb 2305
FlowSorb 2310
1-16
Feature
————— SUPPLIES —————
Mixtures, with helium, of nitrogen, argon,
krypton, carbon dioxide, ethane, n-butane, and
other non-corrosive gases. A mixture of 30%
N
2
point analyses. Mixtures of He and
approximately 5, 12, 18, and 24% N
suggested for multipoint use.
Liquid nitrogen or argon, solvent slush baths,
ice water as appropriate for adsorbate
————— EXPOSED MATERIALS —————
Borosilicate glass, usable to 400 °C
Type 316 stainless steel, chrome plated brass,
copper, Buna-N
————— ENVIRONMENT —————
15-32 °C (59-90 °F) operating;
0-50 °C (32-122 °F) storing and shipping
20-80% relative (non-condensing)
————— CABINET —————
46.5W x 53H x 30.5D cm
(18.3W x 20.9H x 12D in.)
18 kg (40 lbs)
20 kg (44 lbs)
Specification
and 70% He is recommended for single-
FlowSorb III
are
2
Mar 2012
Need help?