Tabs

5�1 Overview
The overview tab gives a global summary of information from the connected WheelDrive.
This information is not adjustable and described below.
1.
This field gives system information.
2.
This field gives all hardware and software information of User Interface, Battery pack and Controller.
3.
This field gives information about the availability of new software versions. By clicking on the "!" symbol,
a new software version can be installed on the system. You will be asked if you want to install the new
available software version.
Warning: it is not possible to go back to previous software versions once the installation has been started.
Disconnecting the WheelDrive during update session will result in an unusable system. Please make sure
that internet connection is stable during update process and your pc has enough power.
4.
This field gives the real-time values of Assist rim sensor and Drive rim sensor. The values for a sensor in
rest are "0".
Maximum values Assist rim: Forward = 200, Backward = - 100
Maximum values Drive rim : Forward = 400, Backward = - 400
5.
This field gives information about the battery pack connected to the WheelDrive. The maximum capacity is
3800 mAh.
6.
This field gives date and time from the WheelDrive internal clock.
The clock will be automatically synchronized when if deviates for more than 1 minute.
7.
This field shows when the next periodic check needs to be executed.
PowerTalk2 User Manual - E1801_20181018
Rev.A
11