6.2.5 Analog Menu

ANALOG OUT MODE
toggles the analog output mode between 0 – 20 mA and 4 – 20 mA.
OUT Lo LIMIT
defines the temperature for the low end of the analog output range (scaling).
Valid numbers between 0.0 to 9999.0°C
OUT Hi LIMIT
defines the temperature for the high end of the analog output range (scaling).
Valid numbers between 0.0 to 9999.0°C
ANALOG IN MODE
toggles the analog input mode between 0 – 20 mA and 4 – 20 mA.
IN Lo LIMIT
defines the temperature for the low end of the analog input range (scaling).
Valid numbers are within the temperature range of the sensor.
IN Hi LIMIT
defines the temperature for the high end of the analog input range (scaling).
Valid numbers are within the temperature range of the sensor.
54
ANALOG
ANALOG OUT MODE
4 – 20 mA
OUT Lo LIMIT
400.0 °C
OUT Hi LIMIT
3000.0 °C
ANALOG IN MODE
4 – 20 mA
IN Lo LIMIT
400.0 °C
IN Hi LIMIT
3000.0 °C