Using the WT Communication (/E2 option)

Procedure up to Data Collection

2
1.
Connect WTs to the recorder using Ethernet cables.
2.
Configure the WT connection client function.
• Basic settings
Set the WT connection client function to On.
Set the data collection interval and recovery action.
• Connection destination server settings
Set the server names (IP address or host name) and the model names of the serv-
ers (WTs) that the Recorder is to connect to.
• Assignment of collection data to communication channels
Set the WTs that data is to be collected from, collected items, and exponential scal-
ing of the data read from the WT.
3.
Configure communication channels, recording settings, display settings, and so on.
• Communication channel settings
Set the span, unit, etc.
• Recording settings
Assign communication channels to recording channels.
• Display settings
Assign communication channels to display groups.
• Other settings
Set the watchdog timer.
4.
Collect data.
IM 04L51B01-19EN