5.9.1 CBJ Care, CBJ1, CBJ2 control unit

The control unit is equipped with a sound signal. A beep
indicates that the battery has low capacity, but lowering the
patient is still possible. It is recommended to charge the
batteries as soon as the sound signal is heard.
1. Plug the power cord A into a power outlet.
The battery will charge in approximately 4
hours. The charger stops automatically when the
batteries are fully charged.
The right yellow diode B will light continuously
during charging, and switch off when fully
charged.
The left green diode C will light continuously
while the control unit is connected to the mains.
2. Disconnect the power cord A from the power outlet
after the battery has been fully charged.
Battery Indicator
The control unit may be equipped with a battery indicator
D, indicating the remaining battery capacity.
CBJ Care
Battery
Battery
Indicator
State
Type
Full
Charge
Partial
Charge
Low
Charge
Low
Charge
(LED
flashing)
60127405-A
Description
The battery is OK — no need
for charging (100–50%). The
top LED is GREEN.
The battery needs to be
charged (50–25%). The middle
LED is YELLOW.
The battery needs to be
charged (Less than 25%).
Acoustic signal when a button
is pressed. The bottom LED is
YELLOW.
The battery needs to be
charged.
Some of the functionality of
the lift is lost and it is only
possible to lower the boom.
CBJ1 and CBJ2 with LCD
Battery
Battery
Indicator
State
Type
Full
Charge
(100%)
Partial
Charge
(75%)
Partial
Charge
(50%)
Low
Charge
(25%)
Low
Charge
(0%)
Optional hand control
For CBJ Care a optional hand control with battery indicator
is available.
Battery
Battery
Indicator
State
Type
Full
Charge
Partial
Charge
Low
Charge
Low
Charge
(LED
flashing)

5.9.2 Optional battery charger

(for control units with detachable battery only)
The procedure of removing or installing the battery
is the same for the control unit and the charger.
Removing the battery
Description
The battery is OK — no need
for charging (100%).
The battery is OK — no need
for charging (75%).
The battery needs to be
charged (50%).
The battery needs to be
charged (25%). Acoustic signal
when a button is pressed.
The battery needs to be
charged.
Some of the functionality of
the lift is lost and it is only
possible to lower the boom.
Description
The battery is OK — no need
for charging (100–50%). The
right LED is GREEN.
The battery needs to be
charged (50–25%). The middle
LED is YELLOW.
The battery needs to be
charged (Less than 25%).
Acoustic signal when a button
is pressed. The left LED is
YELLOW.
The battery needs to be
charged.
Some of the functionality of
the lift is lost and it is only
possible to lower the boom.
Usage