Maintenance

6.2 Cleaning and care

!
IMPORTANT
All cleaning agents and disinfectants used must be effective,
compatible with one another and must protect the materials
they are used to clean. For further information on decon-
tamination in Healthcare Environments, please refer to 'The
National Institute for Clinical Excellence' guidelines on Infec-
tion Control
www.nice.org.uk/CG139
control policy.
!
IMPORTANT
The Platilon covers are not washable and need to be replaced
in case of heavy soilage.
Cleaning black covers
(Removal of contaminants such as dust and organic matter)
1. Remove cover for laundering.
2. Launder the cover at the recommended
and 80 °C using a diluted detergent solution (instructions on label).
!
IMPORTANT
Washing at higher temperatures may cause shrinkage.
Drying black covers
1. Hang cover from a line or bar and drip dry in a clean environment.
!
IMPORTANT
▸ Tumble drying only for the ingress residence cover as this
ensures that the fluorocarbon finish remains effective.
▸ Tumble dry setting must not exceed 80 °C.
▸ Do not tumble dry for longer than 10 minutes.
▸ Dry covers thoroughly before re-fitting to the Flotech
product.
Disinfecting black covers
(Reducing the number of microorganisms)
Please contact your hygiene specialist in the event of contamination.
and your local infection
temperature between 60
!
IMPORTANT
Ensure that any residual detergent has been removed prior to
disinfection.
Light soilage
1. Wipe down the cover with a 0.1% Chlorine Solution (1,000 ppm).
2. Rinse the cover with clean water using a single use nonabrasive
cloth.
3. Dry the cover thoroughly.
Heavy soilage
Where the cover is badly soiled, we recommend cleaning with a dilute
cleaning detergent at 80 °C in the washing machine.
Large spillages of blood should be absorbed and removed with
i
paper towels first.
1. Clean up all spillages of bodily fluids i.e. blood, urine, faeces, spu-
tum, wound exudater and all other bodily secretions as soon as
possible using a 1% Chlorine Solution (10,000 ppm).
2. Rinse the cover with clean water using a single use nonabrasive
cloth, and dry thoroughly.
3. Dry the cover thoroughly.
!
IMPORTANT
1% Chlorine Solution used on a regular basis can diminish the
life of the cover if not rinsed and dried properly.
▸ Do not use granules.
WARNING
▸ Remove contaminated foams from use.
CAUTION
▸ Keep clear of open heat sources.
!
IMPORTANT
▸ Do not use phenols, alcohols, bleaches, or other abrasive
materials.
Invacare Flotech® Contour Visco
7