Training With A Practice Test Is Recommended - Pogo Press. Test. GO Owner's Manual

Table of Contents

Advertisement

Training with a Practice Test is Recommended

While POGO Automatic is simple, its test procedure is different.
Follow these practice instructions step-by-step to train yourself for
successful automated blood collection.
• Load cartridge, close door and TURN
ON meter by pressing the power button
located on the front below display screen.
• After a screen check, the meter will prompt
you when ready to test.
• PRESS FINGER DOWN on test port to
lance. Be sure your finger is completely
covering the test port hole.
• Continue pressing down to start 3-2-1
countdown. At the end of the countdown,
you will be lanced. YOU MUST HOLD
FINGER IN PLACE after lance.
NOTE:
To help remember to hold your finger
down after lance, count 3...2...1...HOLD
FINGER IN PLACE
Customer Service 1-855-IMI-POGO (464-7646) • www.presspogo.com
32
• Lift your finger only when you see
the arrow, indicating your blood has
successfully been collected. If you lift
your finger after lance, it will interrupt
automatic blood collection.
• View results after a 4-second countdown.
2:15 PM
9
2/10/2019
• What if
appears on the meter display instead of your
result? Sometimes, more blood is required to complete the test.
You have 90 seconds to add more blood into the cartridge for
a result. When 60 seconds is left, the meter will start counting
down to 0.
IMPORTANT:
If you lift your finger after lance, it will interrupt
automatic blood collection. The meter will then prompt you to
add more blood into the cartridge. YOU MUST READ the ADD
MORE BLOOD section of this manual to learn how (See Add More
Blood on page 38).
Customer Service 1-855-IMI-POGO (464-7646) • www.presspogo.com
Sample Result Screen