Performing a Control Solution Test

Ensure that a test cartridge is installed
and that there are tests remaining. Control
Solution testing will not work without the
vacuum feature turned on in the Set-Up
Menu (see Vacuum on page 58).
• TURN ON METER.
Power
Button
• After a screen check, the meter will
prompt you when ready for testing.
• Open foil pouch and remove a new
Control Solution applicator for use.
• Place the POGO Control Solution
applicator tip (with foil-end facing
downward) over the test port. Press
down and hold.
• After a 3-2-1 countdown, the
applicator tip will be lanced.
IMPORTANT: DO NOT LIFT APPLICATOR AFTER LANCE.
Hold in place while meter collects Control Solution.
• The meter beeps and an arrow signals
that you can remove the applicator. A
pass or fail result will be displayed after a
4-second countdown.
• Discard the applicator after use.
IMPORTANT:
If your control result appears as an
redo the control test with a new applicator. If this continues to
happen, call Intuity Medical Customer Service at
1-855-IMI-POGO (464-7646).
26
Customer Service 1-855-IMI-POGO (464-7646) • www.presspogo.com
E-4 , E-2 , or NO

Unexpected Meter Display

If this screen appears, the meter did not
receive enough control solution from
the applicator. You have 90 seconds to
complete the control test. When there are
60 seconds left, the meter will count down
to 0.
Pick up the control solution applicator.
Press back down into test port.
• Wiggle the tip back and forth or side to
side to transfer solution to the cartridge.
REPEAT until you get a result or meter
times out. If meter times out, repeat test
with a new applicator.

Control Solution Results

Expected Results
The meter reads the Pass Range for both Level 1 and Level 2 Control
Solutions automatically from the cartridge barcode and compares
the result to the range for the level of control used.
The range is assigned as the mean value +/- 15%. The table below
gives an example of the expected value for each Control level and
resulting calculated Pass Range.
Avg.
Expected
Control Level
Response
Value -15%
(mg/dL)
(mg/dL)
Level 1
75
Level 2
150
Customer Service 1-855-IMI-POGO (464-7646) • www.presspogo.com
Expected
Pass Range
Value +15%
(mg/dL)
(mg/dL)
64
86
64 - 86
127
173
127-173