Testing Your Blood - Pogo Press. Test. GO Owner's Manual

Table of Contents

Advertisement

Pass
If the Control Solution result is in range, the
screen displays a pass result, indicating
the meter and test cartridge are working
properly together. You are now ready to
perform a blood glucose test.
Fail
If you receive a fail Control Solution Test
result, this indicates that the value is
outside of the Control Solution target
range, and you should immediately repeat
the test.
A failed Control Solution Test may indicate:
• An error in your testing method—
for example, removing applicator from
the test port before indicated.
• Expired or improperly stored Control
Solution, exposure to extreme
temperatures or humidity.
• Expired, damaged or improperly stored
test cartridges.
• A problem with the meter.
WARNING:
If your results continue to fail, call Intuity Medical
Customer Service immediately at 1-855-IMI-POGO (464-7646).
DO NOT use the meter to test your blood glucose until you obtain
a passing result.
28
Customer Service 1-855-IMI-POGO (464-7646) • www.presspogo.com
Testing Your Blood Glucose
If this is your first time testing, it is recommended that you practice
first (see Practice on page 32).
The POGO Automatic System is a new
way of testing. POGO Automatic has
All-Automatic
lancing and blood collection are combined
in one easy step. Just place your finger
on the test port and press. The meter
automatically lances your finger and
collects your blood sample without the
need for separate supplies.
It's that simple.
IMPORTANT:
Press and hold your finger on the test port until the meter
beeps and arrow appears. This indicates that the meter has successfully
collected enough blood. Performing this step properly will help to ensure
you get a glucose result when you test.
Testing with POGO Automatic requires
Meter needs only .25
a very small blood sample size—only .25
microliters
microliters. This is about the size of a
of blood
pinhead.
To prepare for testing, wash hands with
soap and water and dry thoroughly. Use of
warm water stimulates blood flow. If you
also choose to use alcohol to clean your
Wash hands
finger, make sure you let it dry. It is also a
good idea to wash your hands any time
after you remove the test platform or open
the cartridge door to handle or change
cartridges.
Customer Service 1-855-IMI-POGO (464-7646) • www.presspogo.com