Adjusting Second Observer Tube; Installing Documentation Accessories - Leica M220 F12 User's Manual & Installation Manual

Table of Contents

Advertisement

7.5

Adjusting second observer tube

Insert stereo attachment/second observer tube into beam
X
splitter.
Tighten the rotary ring (1).
X
Stereo attachment for second observer
The dual stereo attachment can be attached to the left or right on
the beam splitter and turned any direction.
Adjusting the second observer tubes
2
Rotate the monocular second observer tube in the desired
X
direction.
Adjust the diopter setting at the eyepiece.
X
Correct the image with the knurled ring (2).
X
Adjusting the stereo attachment for second observer
Turn the attachment for second observer in the desired
X
direction.
Align the binocular tube horizontally.
X
Adjust the diopter setting at the eyepiece.
X
Correct the image with the knurled ring (2).
X
7.6
1
1 Adjusting the magnification
2 Focusing knob
Fitting the phototube
X
X
Please refer to section 9.2 for a list of video accessories.
Leica M220 F12 / Ref. 10 716 922 / Version 03
Installing documentation
accessories
WARNING
Risk of injury from parts falling down!
Complete all preparations and adjustments on the
X
optics carrier before the operation.
Never rebalance or re-equip the optical components and
X
accessories with the instrument over the operation area.
Before re-equipping, always lock the swing arm.
X
Check that the optical components and accessories fit
X
well and are fastened securely before the operation.
Before changing equipment during an operation, swivel
X
the microscope away from the operation area.
1
2
Fasten the phototube to the documentation port of the 0°
assistant attachment or to the beam splitter.
Secure the camera, complete with adapter, in the phototube.
Tighten the clamping screw.
Preparation before surgery
19