    The pipettes enable precise and simultaneous delivery of 8 or 12 preset-volume doses of liquid. BIOPETTE PLUS 12 channel BIOPETTE PLUS pipettes are produced in four ranges of SPECIFICATIONS volumes: 1-10 μl, 5-50 μl, 20-200 μl, 50-300 μl. V V o o l l u u m m e e...
  • Page 5: Method Of Pipetting

    MULTICHANNEL The pipette design enables the user to perform the recali- button (Fig. 1A2) or the adjustment knob (Fig. 1B) increas- bration process according to the rules presented in sec- ing the value until the lower figure wheel comes 1/3 of a tion 5.
  • Page 6: Recommendations

    Permissible BIOPETTE PLUS pipettes are calibrated by gravimetric Model pipette volume volumes the calibration method, using Labnet tips and distilled water, at the tem- volumes [μl] [μl] key [μl] perature 20±1°C, according to EN ISO 8655 standard. [μl] (24 increments)
  • Page 7: Cleaning And Sterilization

    MULTICHANNEL • Calculate average aspirated volume in μl multiplying Sterilization: the average aspiration amount [mg] by the distilled The pipette can be sterilized in the autoclave at 121°C for water density coefficient [μl/mg], which depends on 20 minutes. After sterilization, the pipette should be dried temperature and pressure according to the following and cooled to room temperature.
  • Page 8 Tip cone F: Ejector G: Calibration key J: Locking ring can be ordered from your Labnet representative (type of pipette and name of the part for this pipette should be specified). Warning: The replacement of the piston requires con- ducting of calibration procedure according to section 5.