Page 1
IX2-RFSPOT INSTRUCTIONS The IX2-RFSPOT has been designed to be used in place of the field iris diaphragm (FS) of a IX2 Series reflected fluorescence illuminator (IX2-RFAL). When the IX2-RFSPOT is used, a desired pinhole* can be projected on the specimen plane, which is convenient for canceling the cage on caged specimens.
Page 2
Installation CAUTION: To ensure safety, be sure to turn off the reflected light source before attaching. Removing the Field Iris Diaphragm Using the Allen screwdriver provided with the microscope frame, loosen the field iris diaphragm clamping screw on the IX2-RFAL and pull out the field iris diaphragm to remove it. Attaching the Pinhole 1.
Page 3
Setting the pinhole diameter The relationship between the actual pinhole diameter and the diameter of its projection on the specimen surface is as expressed below. Pinhole diameter on specimen = 2X (Pinhole diameter/Objective magnification) (Example) When the pinhole diameter is 100 µm and objective magnification is 40X: 2X (100/40) = 5 µm # If strict pinhole diameter setting is required, actually project it on a specimen and measure the diameter using a scale.
Page 4
UIS Series Excitation Wavelength & Priority* (Brightness/Contrast) 340 nm 360 nm 480 nm Objective Type** & Magnification Brightness Contrast Brightness Contrast Brightness Contrast Number of Washers Number of Washers Number of Washers UApo20XW3/340 UApo20X3/340 UApo40XOI3/340 UApo40XW3/340 UPlanApo40X UPlanApo40XOI3 PlanApo40XWLSM PlanApo60XO3 UPlanApo60XW3 UPlanApo60XW3/IR PlanApo60XWLSM...