The IX2-RFSPOT has been designed to be used in place of the field iris diaphragm (FS) of a IX2 Series reflected
fluorescence illuminator (IX2-RFAL). When the IX2-RFSPOT is used, a desired pinhole* can be projected on the
specimen plane, which is convenient for canceling the cage on caged specimens.
* No pinhole is provided with the unit. Please fabricate a pinhole yourself (see the figure illustrated later in this
manual) or use the commercially available recommended product (pinhole manufactured by Melles Griot, outer
diameter 16 mm).
1
External View
Clamping ring
Washers
(x 4)
Microscope frame side
Focus adjustment screw
2
Precautions for Operation
· Flare is produced around the pinhole. Ghost may be noticeable when the pinhole position is deviated from the center of
the field of view.
· The pinhole image may not be able to be focused perfectly depending on the objective and wavelength in use.
· Centering of the pinhole guarantees only the movement in the center of the field of view. Movement to the peripheral
section of the field is not possible.
· The pinhole position may be deviated slightly when the pinhole is engaged or disengaged. Perform centering whenever
strict positioning is required.
· The pinhole cannot be set to an intermediate position between the IN and OUT positions.
· The IX2-SFR stage gets in the way of centering the pinhole. To prevent it, remount the IX2-SFR with the front in the back or
replace it with the IX-SVL-2.
PINHOLE UNIT
IX2-RFSPOT
Lamp housing side
Centering screws
Pinhole engagement knob
Pushed in: Pinhole engaged (IN)
Pulled out: Vacant hole (OUT)
Clamping screw
INSTRUCTIONS
A X 7 3 8 3
IX2-RFSPOT

Advertisement

loading

Summary of Contents for Olympus IX2-RFSPOT

  • Page 1 IX2-RFSPOT INSTRUCTIONS The IX2-RFSPOT has been designed to be used in place of the field iris diaphragm (FS) of a IX2 Series reflected fluorescence illuminator (IX2-RFAL). When the IX2-RFSPOT is used, a desired pinhole* can be projected on the specimen plane, which is convenient for canceling the cage on caged specimens.
  • Page 2 Installation CAUTION: To ensure safety, be sure to turn off the reflected light source before attaching. Removing the Field Iris Diaphragm Using the Allen screwdriver provided with the microscope frame, loosen the field iris diaphragm clamping screw on the IX2-RFAL and pull out the field iris diaphragm to remove it. Attaching the Pinhole 1.
  • Page 3 Setting the pinhole diameter The relationship between the actual pinhole diameter and the diameter of its projection on the specimen surface is as expressed below. Pinhole diameter on specimen = 2X (Pinhole diameter/Objective magnification) (Example) When the pinhole diameter is 100 µm and objective magnification is 40X: 2X (100/40) = 5 µm # If strict pinhole diameter setting is required, actually project it on a specimen and measure the diameter using a scale.
  • Page 4 UIS Series Excitation Wavelength & Priority* (Brightness/Contrast) 340 nm 360 nm 480 nm Objective Type** & Magnification Brightness Contrast Brightness Contrast Brightness Contrast Number of Washers Number of Washers Number of Washers UApo20XW3/340 UApo20X3/340 UApo40XOI3/340 UApo40XW3/340 UPlanApo40X UPlanApo40XOI3 PlanApo40XWLSM PlanApo60XO3 UPlanApo60XW3 UPlanApo60XW3/IR PlanApo60XWLSM...