9
Cleaning and Maintenance
Cleaning
To enable your Clinitek Status
analyzer to perform as intended
and provide reliable test results, it
is recommended that you
periodically check the white
calibration bar on the test table.
In normal use, the white calibration
bar should not become dirty or
discolored.
BIOHAZARD Wear personal protective
equipment. Use universal precautions.
Refer to Appendix H for recommended
precautions when working with
biohazardous materials.
Cleaning the White Calibration Bar
1
Remove the insert from the test
table.
2
Remove the test table by pulling
it slowly out of the analyzer.
3
Check the white calibration bar
on the test table for dirt or
discoloration.
9-4
®
White
Calibration Bar
132387 Rev. D, 2006-06