9
Cleaning and Maintenance
Cleaning
Disinfecting the Test Table and
Insert
1
Prepare one of the following
solutions in a tall, narrow
container (e.g., empty Multistix
bottle) to a depth of about
4 inches (10 cm):
• Precept, Cidex,
Theracide and Amphyl
solutions - prepare
according to product
directions.
• Household Bleach (5%
sodium hypochlorite) – this
can be used either full
strength or dilute with water
to as much as 1:20
(i.e., mix 5 mL bleach with
95 mL water for a total of
100 mL).
Rinse away all bleach
residue, as remaining bleach
will affect several of the reagent
pad chemistries.
• Isopropyl Alcohol (70% to
85%) - this can be used full
strength.
Any solutions other than those
listed above may damage the test
table and insert.
132387 Rev. D, 2006-06
2
Place the insert and/or test
table into the solution, making
sure the white calibration bar on
the test table remains above
the liquid level.
®
Be sure the solution does not
come in contact with the white
calibration bar. Do not cover the
container while the test table is
soaking.
3
Soak the table and insert for a
minimum of 2 minutes and
maximum of 10 minutes. Do not
soak longer than 10 minutes.
4
Rinse the test table and insert
thoroughly with water.
5
Dry with a soft cloth and
replace test table and table
insert in the analyzer (as
described on the previous
page).