Model 220 Current Source Connections; Figure - Keithley 7174A Instruction Manual

8×12 low current matrix card
Table of Contents

Advertisement

Model 220 current source connections

The Model 220 current source can be connected to the matrix
card using the Model 6167 Guarded Adapter, as shown in
Figure 2-10. This configuration guards the output signal to
minimize the effects of distributed capacitance and leakage
current.
NOTE
The Model 6167 adapter must be modified
by internally disconnecting the inner
shield connection of the input jack from
the GUARDED/UNGUARDED selection
switch. Otherwise, instrument LO will be
Model 6167
Guarded Adapter
Connect GUARD OUT
to GUARD
Note: See paragraph 2.5.3 and Figure

Figure 2-10

Model 220 current source connections
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
220 Current Source
2-4 for special triax to banana cable.
connected to chassis ground through the
adapter.
1. Connect the Model 6167 adapter to the Model 220
OUTPUT jack.
2. Connect a Model 7078-TRX-(3, 10 or 20) triax cable
between the guarded adapter and the desired row of the
Model 7174A.
3. Connect the Model 220 GUARD output to GUARD
INPUT terminal of the adapter.
4. Connect the triax end of a triax/banana cable to the
desired row on the Model 7174A.
5. Connect the banana plug end of the triax/banana cable
to the OUTPUT COMMON jack of the Model 220.
7078-TRX Triax
237-BAN-3 or Special Triax
Operation
K E ITHL E Y
7174A
8 x 12 L OW
CUR R E NT MATR IX
S IGNAL
COLUMNS
200V P K
200V P K
GU AR D
200V P K
R OWS
1
A
2
B
3
C
4
D
5
6
7
8
E
9
F
10
G
11
H
12
!
WAR NING:
TIGHTE N MOUNTING S CR E WS
TO E NS UR E P R OP E R
CHAS S IS GR OUND
7174A Matrix Card
2-15