Evikon PluraSens E2638 Series User Manual page 18

Ethylene oxide detector-transmitter
Hide thumbs Also See for PluraSens E2638 Series:
Table of Contents

Advertisement

E2638-ETO
220 /
0x00DC
40221
221 /
0x00DD
40222
222 /
0x00DE
40223
223 /
0x00DF
40224
224 /
0x00E0
40225
255 /
0x00FF
40256
RW Control logic for relay RE2
LOW setpoint for relay
RW
RE1
HIGH setpoint for relay
RW
RE1
LOW setpoint for relay
RW
RE2
HIGH setpoint for relay
RW
RE2
Sensor, analog outputs,
RW
LED and buzzer status
0 – none
1 – relay on at high values
2 – relay on at low values
3 – relay on at values within
the range
4 – relay on for the values out
of the range
0...65535 (gas units)
0...65535 (gas units)
0...65535 (gas units)
0...65535 (gas units)
bit[0]=0/1 – sensor
present/absent, RO
bit[1]=0/1 – analog outputs
deactivated/activated
bit[2]=0/1 – in case the
sensor is absent, turn
signaling off/on analog
output1
bit[3]=0/1 – in case the
sensor is absent, turn on
signaling with low
current/high current on
analog output1; if bit[2]=0 this
bit will be ignored
bit[4]=0/1 – in case of sensor
absent, turn signaling off/on
analog output2
bit[5]=0/1 – in case of sensor
absent, turn on signaling with
low current/high current on
analog output2; if bit[4]==0
this bit will be ignored
bit[6]=0/1 – current/voltage
output detected on output1,
RO
bit[7]=0/1 – current/voltage
output detected on output2,
18
Rev 10.03.2021
0
see
Specifications
see
Specifications
see
Specifications
see
Specifications