Evikon PluraSens E2638 Series User Manual page 16

Ethylene oxide detector-transmitter
Hide thumbs Also See for PluraSens E2638 Series:
Table of Contents

Advertisement

E2638-ETO
0x0005
5 / 40006 RW Baud rate *
0x0006
6 / 40007 RW Response delay, ms
0x0007
7 / 40008 RW Stop bits, parity bit *
0x0008
8 /40009
0x0011
17 / 40018 RW
165 /
0x00A5
40166
166 /
0x00A6
40167 RW
167 /
0x00A7
40168 RW
168 /
0x00A8
40169 RW
Address
Reg / MHR RW Description
201 /
0x00C9
40202
202 /
0x00CA
40203
203 /
0x00CB
40204
204 /
0x00CC
40205
R
Last error code
Technological:
age of last data in seconds
(read) / restart (write)
Zero adjustments for gas
RW
data, ADC
Slope adjustment for gas
data
Change rate limit for gas
units/ s
Integrating filter time
constant, s
Parameter tied to analog
RW
output 1
Parameter tied to analog
RW
output 2
Forced value for analog
RW
output 1***
Forced value for analog
RW
output 2***
1200, 2400, 4800, 9600, 19200,
38400, 57600
1...255
1 – No parity bit, 1 stop bit
(default after factory reset)
2 – No parity bit, 2 stop bits
3 – Odd parity, 1 stop bit
4 – Even parity, 1 stop bit
NOTE: 3 and 4 are available
starting from the Software
version 0x218 (dec. 536)
1...255
0...65535 s (read),
writing 42330 restarts the
instrument
-32000...+32000 ADC units
1...65535
1...32000, 0 - no limit
1...32000 (seconds),
0 - no filter
Supported values (dec)
0 – None
2 – Gas concentration
9 – Forced Modbus control,
value set in MHR / 40204
0 – None
2 – Gas concentration
9 – Forced Modbus control,
value set in MHR / 40205
0...1000
(0.0%...100.0% of output
scale)
0...1000
(0.0%...100.0% of output
scale)
16
Rev 10.03.2021
9600
10
1
-
-
0
512
0
0
Default
2
2
0
0