Table 6-9 MS Method Parameters (continued)
Parameter
Definition
Apply Scan
Select to apply the Retention Time window to
Schedule
individual MRM transitions so that each MRM
experiment can be executed at the specified time.
Users can enter up to 1250 Scheduled MRM
transitions when working with complex samples.
Sort by precursor
Apply to sort the precursor masses in ascending
ion
order in the MS Table prior to acquisition. In some
cases, acquiring MRM transitions in ascending order
might improve MRM sensitivity.
Enhance dynamic
For TOF MS scans, extends the linear dynamic range
range
of measured components to higher concentrations.
If this feature is selected, the cycle time might vary.
Note: When this feature is selected,
accumulation time for MSMS experiments cannot
be set to less than 25 ms.
Advanced Experiment Settings
Q1 resolution
Select a resolution to use to separate closely spaced
components. Q1 resolution is the resolving ability
of the quadrupole. When more sensitivity is
required in an MSMS experiment, the resolution
settings can be reduced to allow more ions to be
analyzed. The Q1 resolution setting can be set to
Low or Open to increase the intensity of the
resulting signal or to allow more than just the C12
isotope in any isotope pattern to be shown. Setting
the resolution to Open provides more signal
strength and allows more isotopes to be shown.
Step size
Type the size of the scan interval in Daltons.
Time bins to sum
Select the number of data points to be summed.
Range for small molecule or peptides: 4 to 6.
Starting value for Intact protein analysis ( > 20 k
Da): 40
Channels 1 to 4
Select the time-to-digital converter (TDC) channels.
Each channel counts ions. If all four of the channels
are selected (the default setting), then all four
channels are summed for the total ion count.
System User Guide
RUO-IDV-05-2334-E
Operating Instructions Software
Experiment
MRM HR
HR
MRM HR
TOF MSMS, SWATH, MRM HR,
Scheduled MRM HR
TOF MSMS, TOF MSMS (SWATH), MRM
HR and Scheduled MRM HR
TOF MSMS, Q1, IDA
Q1
TOF MS, TOF MSMS, IDA, MRM HR,
SWATH
TOF MS, TOF MSMS, IDA, MRM HR,
SWATH
SCIEX X500 QTOF System
97 / 234