inbody BPBIO210 User Manual
Hide thumbs Also See for BPBIO210:

Advertisement

BPBIO210/220
User's Manual

Advertisement

Table of Contents
loading

Summary of Contents for inbody BPBIO210

  • Page 1 BPBIO210/220 User’s Manual...
  • Page 2 Reproduction, adaptation, or translation of this manual is prohibited without prior written consent from InBody Co., Ltd. under the copyright laws. This manual may be printed incorrectly and subject to change without notice. InBody Co., Ltd. shall not be liable for any errors, incidental, or consequential damages that occurred by not complying with the content of the User᾽s Manual.
  • Page 3 InBody Europe B.V. Gyroscoopweg 122, 1042 AZ, Amsterdam, The Netherlands TEL: +31 (0)20 238 6080 FAX: +31 (0)6 5734 1858 Website: https://nl.inbody.com E-mail: info.eu@inbody.com Australian Sponsor. [AUSTRALIA] Emergo AUSTRALIA. Level 20, Tower II, Darling Park, 201 Sussex Street, Sydney, NSW 2000, AUSTRALIA TEL: +61-2-9006-1662 FAX: +61-2-9006-1010 Website: https://www.emergogroup.com E-mail: Sponsor@emergogroup.com...
  • Page 5: Table Of Contents

    A. Precautions before Measurement External Appearance B. Measurement Instructions B. Installation Environment C. Storage C. Installation Instructions D. Features and Functions 3. BPBIO210/220 Common I. FAQ II. Others A. Error code A. Optional Equipment B. Safety Information B. Regarding the Equipment C. Product Classification...
  • Page 6: Product Installation

    I. Product Installation A. Product Components and External Appearance BPBIO210 can be used by medical professionals to check patient's blood pressure by using the auscultation method with stethoscope and pressure indicator bar displayed on the LCD. The back-light function on the LCD enables blood pressure monitoring even in a dark environment.
  • Page 7 Main Unit Parts 'MARK' Button On / Off Button Backlight Button Battery Insertion Air Bulb Tube Connection Cuff Tube Connection Dedicated AC Adapter Connection Cuff Parts Wearable Arm Circumference Indicator Arterial Location When Worn Cuff Tube...
  • Page 8 Air Bulb Unit Parts Exhaust Valve Air Bulb Air Bulb Tube Display Parts Pressure Guide Pressure Display Bar AC Adapter Indicator Battery Level Indicator Pulse Icon Unit for Pulse Rate Unit for Pressure Pressure/Pulse Value...
  • Page 9: Installation Environment

    C. Installation Instructions (1) Unboxing BPBIO210 After opening BPBIO210 from the box package, take it out from the air cap buffer and the vinyl package. Open BP210 and take out the Cuff and Air Bulb from the vinyl package. Buffer...
  • Page 10 (2) Connecting Cuff and Air Bulb Unit 1. Plug the Cuff Tube into the Cuff Tube Connection of the Main Unit. ( 2. Plug the Air Bulb Tube into the Air Bulb Tube Connection of the Main Unit. ( 3. Please plug the tube tightly to the end. (3) Inserting / Replacing Batteries 1.
  • Page 11 • Keep the AC Adapter plug away from foreign objects which can cause any kind of damage. • Use only the AC Adapter provided by InBody. • Do not use the dedicated AC adapter for any other purposes other than for the operation of BPBIO210.
  • Page 12: Features And Functions

    D. Features and Functions (1) Automatic Power Off • Main Unit will switch off automatically if idle for more than 2 minutes to prevent battery drain. • Main Unit will not switch off if connected to the AC Adapter. (2) Protection from Excessive Pressurization •...
  • Page 13 A. Precautions Before Measurement Caution • BPBIO210 has to be operated by skilled persons who are well trained of the auscultation method to measure blood pressure. • Do not operate BPBIO210 which has been unused for a along time under the environment where the temperature was less then 10˚C or higher then 40˚C.
  • Page 14 Caution • Wear the Cuff properly for accurate measurement. • If the Cuff is too loose, measurement result may not be accurate. (2) Proper Measurement Posture ❶ Place the arm so that the center of the Cuff is leveled with the patient's heart position. Height of the heart ❷...
  • Page 15 (4) Checking the Battery Level ❶ The remaining Battery level can be checked from the lower right area when the product is switched on. ❷ If the AC Adapter is used, the battery level will not be displayed, but instead, AC Adapter Icon ( ) will be displayed.
  • Page 16 (7) Depressurization ❶ Rotate the exhaust valve counterclockwise to open the valve. Proceed with blood pressure measurement at a proper exhaust rate. ❷ If the button on the main body is pressed at a certain pressure value, the pressure marking will be displayed accordingly. * The marking can be displayed up to 5 times.
  • Page 17 C. Storage (1) Wipe the stains on the Main Unit with a smooth cloth. (2) Place the Cuff and Air Bulb unit as shown below. BPBIO210 should be transported or stored under the following conditions. Temperature Range − 20 ~ 60˚C Relative Humidity...
  • Page 18: Bpbio220

    2. BPBIO220 I. Product Installation A. Product Components and External Appearance BPBIO220 can be used by medical professionals to check patient's blood pressure by using auscultation method with stethoscope and pressure indicator bar displayed on the LCD. The back-light function on the LCD enables blood pressure monitoring even in a dark environment.
  • Page 19 Main Unit Parts MARK Button On / Off Button Backlight Button Battery Insertion Controller 4-Pole Plug Connection Controller Tube Connection Cuff Tube Connection Dedicated AC Adapter Connection Cuff Parts Wearable Arm Circumference Indicator Artery Indicator When Worn Cuff Tube...
  • Page 20 Controller Parts Exhaust Valve Wheel Switch Air Bulb Controller 4-Pole Plug Connector Controller Tube Display Parts Pressure Guide Pressure Display Bar AC Adapter Indicator Battery Level Indicator Pulse Icon Unit for Pulse Rate Unit for Pressure Pressure/Pulse Value...
  • Page 21: Installation Environment

    B. Installation Environment Check for proper surrounding environment before installing BPBIO220. This equipment is suitable for indoor use. If installing the equipment outdoors, The following requirements must be fulfi lled If being installed in the outdoors. Temperatures Range 10 ~ 40˚C Relative Humidity 30 ~ 85% RH Atmospheric Pressure Range 70 ~ 106kPa C.
  • Page 22 (2) Connecting Cuff and Air Bulb Unit 1. Plug the Cuff Tube into the Cuff Tube Connection of the Main Unit ( 2. Plug the Controller 4-Pole Plug Connector and Controller Tube into the Connector ( of the Main Unit. 3.
  • Page 23 • Keep the AC Adapter plug away from foreign objects which can cause any kind of damage. • Use only the AC Adapter provided by InBody Co., Ltd. • Do not use the dedicated AC adapter for any other purposes other than for the operation of BPBIO220.
  • Page 24: Features And Functions

    D. Features and Functions (1) Automatic Power Off • Main Unit will switch off automatically if idle for more than 2 minutes to prevent battery drain. • Main Unit will not switch off if connected to the AC Adapter. (2) Protection from Excessive Pressurization •...
  • Page 25: Blood Pressure Measurement

    Blood Pressure Measurement A. Precautions Before Measurement Caution • BPBIO220 has to be operated by skilled persons who are well trained of the auscultation method to measure blood pressure. • Do not operate BPBIO220 which has been unused for a along time under the environment where the temperature was less then 10˚C or higher then 40˚C.
  • Page 26 Caution • Wear the Cuff properly for accurate measurement. • If the Cuff is too loose, measurement result may not be accurate. (2) Proper Measurement Posture ❶ Place the arm so that the center of the Cuff is in level with the patient's heart position. Height of the heart ❷...
  • Page 27 (4) Checking the Battery Level ❶ Battery level can be checked from the lower right area when the product is switched on. ❷ If the AC Adapter is used, the battery level will not be not displayed, and AC Adapter Icon ( will be displayed.
  • Page 28 (7) Depressurization ❶ Rotate the Exhaust Valve counterclockwise to open. Or, rotate the wheel switch to adjust the exhaust speed. Proceed with blood pressure measurement at a proper exhaust rate. ❷ If the button on the Main Body is pressed at a certain pressure value, measured blood pressure value will be displayed accordingly.
  • Page 29: Storage

    C. Storage (1) Press the Button to switch off the power. (2) Place the Cuff and Controller as shown below. BPBIO220 should be transported or stored under the following conditions. Temperatures Range - 20 ~ 60˚C Relative Humidity 10 ~ 85% RH (No Condensation) Atmospheric Pressure Range 50 ~ 106 kPa...
  • Page 30 If the same error message occurs continuously after When the Pressure sensor exhausting the air from the Cuff, the pressure sensor malfunction. may be defective. Please contact InBody Customer Center (080-501-3939). Detection of adapter Use the dedicated AC Adapter after disconnecting the abnormality. current adapter.
  • Page 31: Blood Pressure Measurement

    The controller not connected properly. tightly and push it all the way in. does not work, or it works Please contact InBody Customer Center abnormally. The controller is damaged. (080-501-3939). Blood pressure is measured too Keep the decompression speed at rapidly.
  • Page 32: Optional Equipment

    M-sized Cuff For arm circumstance of 22 to 32 cm BP-C-M-01 S-sized Cuff For arm circumstance of 17 to 22 cm BP-C-S-01 Available for BPBIO210 Air-Bulb BP-CO-M-01 Does NOT contain the Wheel Switch Available for BPBIO220 Controller Contains the Wheel Switch...
  • Page 33: Safety Information

    B. Safety Information Safety Symbols Power Adapter Connection Terminal BF-type equipment Operating instructions CLASS II equipment (Dedicated AC adapter) Other Symbols Disposal Guidelines Manufacturer Serial number Authorized representative in the EUROPEAN COMMUNTY European conformity...
  • Page 34: Product Classification

    Caution Storage 1. Please note that the Cuff and the Tube can be damaged by a sharp bite. 2. Do not store the blood pressure monitor in the following places: - Prevent where water or other liquids can reach - Exposure to Direct Sunlight - Prevent from strong vibration or shock - Prevent dust, salt, and sulfer content - Prevent from hot and humid environment...
  • Page 35 Cuff for the Use of 22~32cm in Arm Circumference (M size cuff) Pressurization Manual Pressurization Method by Air Bulb De-pressurization BPBIO210:variable exhaust valve method BPBIO220: Variable Exhaust Valve Method/Variable Wheel Switch Method Rated Power DC 6V, 2A (four 1.5V AA-size batteries or dedicated AC adapter)
  • Page 36: Emc Environment Guidance

    E. EMC environment guidance Guidance and Manufacturers Declaration - Electromagnetic Emissions The BPBPIO210/220 is intended for use in the electromagnetic environment specified below. The customer or the user of the BPBIO210/220 should assure that it is used in such an environment. Emissions Test Compliance Electromagnetic Environment - Guidance The [ME EQUIPMENT or ME SYSTEM] uses RF energy only for its internal function.
  • Page 37 NOTE UT is the a.c. mains voltage prior to application of the test level. Guidance and Manufacturers Declaration - Electromagnetic Emissions The BPBIO210/220 is intended for use in the electromagnetic environment specified below. The customer or the user of the BPBIO210/220 should assure that it is used in such an environment. IEC 60601 Electromagnetic Environment...
  • Page 38 where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and D is the recommended separation distance in metres (m). Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey, should be less than the compliance level in each frequency range.
  • Page 40 ©2019 InBody Co., Ltd. All rights reserved. BM-ENG-K6-A-190812...