Page 1
BPBIO210/220 User's Manual for Measurement Guide and Setup Thank you for purchasing the BPBIO210/220. This user’s manual describes all the features of the BPBIO210/220. Please read before use and keep it in a safe place. By following the manual instructions, you will be able to use the...
Page 3
Reproduction, adaptation, or translation of this manual is prohibited without prior written consent from InBody Co., Ltd. under the copyright laws. This manual may be subject to change without notice. InBody Co., Ltd. shall not be liable for any errors, incidental, or consequential damages that occurred by not complying with the content of the User᾽s Manual.
Page 4
Australian Sponsor. [AUSTRALIA] Emergo AUSTRALIA. Level 20, Tower II, Darling Park, 201 Sussex Street, Sydney, NSW 2000, AUSTRALIA USA CONTACT BIOSPACE INC. dba INBODY 13850 CERRITOS CORPRT DR STE C, Cerritos, California, 90703 TEL: (323) 932-6503 Customer Service Information InBody USA [USA] 13850 Cerritos Corporate Dr.
Page 5
Intended Use: The BPBIO210/220 is a digital device intended for use in measuring blood pressure and pulse rate in user population with upper arm circumference ranging from 17cm to 42cm (7-inch to 17-inch). This device should be used by medical professionals. The systolic blood pressure and diastolic blood pressure shall be measured by the auscultatory method.
BPBIO210/220 User’s Manual Contents 1. BPBIO210 I. Introduction for the BPBIO210 II. Blood Pressure Measurement A. Standard Components A. Precautions before Measurement B. Installation Environment B. Measurement Instructions C. Installation Instructions C. Storage D. Features and Functions 2. BPBIO220 I. Introduction for the BPBIO220 II.
1. BPBIO210 I. Introduction for the BPBIO210 This blood pressure monitor can be used by a medical professional to check blood pressure with auscultatory method by using the pressure indicator bar on the LCD screen. The back-light function allows blood pressure monitoring even under dark conditions.
Page 9
Main Unit Parts 'MARK' button On / Off button Back-light button Battery tray Air bulb jack Cuff jack AC adapter jack Cuff Parts Checking mark for the available arm circumference Artery mark Cuff tube...
Page 10
Air Bulb Parts Exhaust valve Air bulb Air bulb tube Display Screen Parts Pressure guide Pressure display bar AC adapter connection symbol Battery level indicator Pulse symbol Unit for pulse rate Unit for pressure Pressure/Pulse...
(1) Unboxing the BPBIO210 After opening the BPBIO210 packaging box, take out the BPBIO210 from the buff er and the protective vinyl package. Open the BPBIO210 and take out the cuff and the air bulb from the protective vinyl package.
Page 12
(2) Connecting cuff and air bulb 1. Plug the cuff tube into the cuff jack of the main unit. ( 2. Plug the air bulb tube into the air bulb jack of the main unit. ( 3. Push the tubes all the way in to the end. (3) Inserting / replacing batteries 1.
Page 13
• Remove the batteries from the battery compartment when it will not be in use for long period. • Dispose used batteries only in designated locations. • Improper use of batteries may cause degradation of the performance of BPBIO210 and may cause fire.
Page 14
• Keep the adapter away from foreign objects. • Use only the AC adapter provided by InBody. • Do not use the AC adapter for any other purposes other than for the operation of BPBIO210. Attention • Branchez correctement l’adaptateur secteur sur la prise secteur.
Page 15
(5) Installing the cart (Optional) Screw the four bolts into the bracket by using a cross slot screwdriver to fasten the bracket to the cart.
• The BPBIO210 will automatically shut down after 5 minutes has elapsed without pressing any buttons. • The BPBIO210 will not shut down if it is connected to the AC adapter. Main unit will switch off automatically if idle for more than 5 minutes to prevent battery drain.
Page 17
(2) Protection from excessive pressurization • If the pressure applied to the cuff exceeds 320mmHg, the BPBIO210 will automatically exhaust the pressure from the cuff to protect the patient. (3) Residual air detection • If there is any remaining air left in the cuff, the error code 'E1' will appear on the screen when the product power is switched on.
Page 18
• The BPBIO210 must be operated by those who are well trained of the auscultatory method of measuring the blood pressure. • Do not operate the BPBIO210 immediately if it has been left for a long period of time in an environment where the temperature is lower than 10ºC (50ºF) or higher than 40ºC (104ºF).
Page 19
B. Measurement Instructions (1) How to wear the cuff ❶ Wrap the cuff around the arm, about 2-3 cm INDEX(▲) (1 in) above the elbow. ❷ Make sure that the index (▲) on the cuff is RANGE located within the RANGE bar. Replace the cuff with the right size if the index deviates from the RANGE.
Page 20
(2) Proper measurement posture ❶ Place the arm so that the middle of the cuff is at the patient’s heart position. Height of the heart ❷ Be relaxed during the measurement. Caution • The measurement posture may affect the blood pressure values. •...
Page 21
(4) Battery level indicator ❶ The remaining battery level can be checked in the bottom right corner of the screen while using the blood pressure monitor. ❷ When using the AC adapter, the battery level is not displayed, instead the AC adapter connection symbol ( ) appears.
Page 22
(7) Depressurization ❶ Rotate the exhaust valve counterclockwise to open the valve. Continue the blood pressure measurement at a proper exhaust rate. ❷ If the button on the main unit is pressed, a bar symbol will be displayed on the pressure display bar.
• Contact your local InBody representative to purchase the cuff. Remarque • Si le brassard est contaminé, et qu’il ne peut plus être utilisé, veuillez le remplacer par un nouveau brassard. • Communiquez avec le représentant InBody de votre région pour vous procurer des brassards additionnels.
2. BPBIO220 I. Introduction for the BPBIO220 This blood pressure monitor can be used by a medical professional to check blood pressure with auscultatory method by using the pressure indicator bar on the LCD screen. The back- light function allows blood pressure monitoring even under dark conditions. The user may control the speed of the exhaust by using the controller and mark the systolic and the diastolic blood pressure values on the pressure-bar display.
Page 25
Main Unit Parts MARK button On / Off button Back-light button Battery tray Controller 4-Pole plug jack Controller tube jack Cuff jack AC adapter jack Cuff Parts Checking mark for the available arm circumference Artery mark Cuff tube...
Page 26
Controller Parts Exhaust valve Jog switch Air bulb Controller 4-pole plug Controller tube Display Screen Parts Pressure guide Pressure display bar AC adapter connection symbol Battery level indicator Pulse symbol Unit for pulse rate Unit for pressure Pressure/Pulse...
B. Installation Environment Please make sure the environment is appropriate for the BPBIO220 installation. This equipment is designed for indoor use. If installing the equipment outdoors, the following requirements must be met. Temperature range 10 to 40ºC (50 to 104ºF) Relative humidity 30 to 85% RH Atmospheric pressure range...
Page 28
(2) Connecting cuff and air bulb 1. Plug the cuff tube into the cuff tube jack of the main unit ( 2. Plug the controller 4-pole plug and the controller tube into the corresponding jack ( of the main unit. 3.
Page 29
• Remove the batteries from the battery compartment when it will not be in use for long period. • Dispose used batteries only in designated locations. • Improper use of batteries may cause degradation of the performance of BPBIO220 and may cause fire.
Page 30
• Do not apply excessive force to the AC adapter and the plug. • Keep the AC adapter from foreign objects. • Use only the AC adapter provided by InBody. • Do not use the AC adapter for any other purposes other than for the operation of BPBIO220.
Page 31
(5) Installing the cart (Optional) Screw the four bolts into the bracket by using a cross slot screwdriver to fasten the bracket to the cart.
D. Features and Functions (1) Automatic power off • The BPBIO220 will automatically shut down after 5 minutes has elapsed without pressing any buttons. • The BPBIO220 will not shut down if it is connected to the AC adapter. Main unit will switch off automatically if idle for more than 5 minutes to prevent battery drain.
Page 33
(2) Protection from excessive pressurization • If the pressure applied to the cuff exceeds 320mmHg, the BPBIO220 will automatically exhaust the pressure from the cuff to protect the patient. (3) Residual air detection • If there is any remaining air left in the cuff, the error code 'E1' will appear on the screen when the product power is switched on.
Blood Pressure Measurement A. Precautions before Measurement Caution • The BPBIO220 must be operated by those who are well trained of the auscultatory method of measuring the blood pressure. • Do not operate the BPBIO220 immediately if it has been left for a long period of time in an environment where the temperature is lower than 10ºC (50ºF) or higher than 40ºC (104ºF).
B. Measurement Instructions (1) How to wear the cuff ❶ Wrap the cuff around the arm, about 2-3cm INDEX(▲) (1 in) above the elbow ❷ Make sure that the index (▲) on the cuff is RANGE located within the RANGE bar. Replace the cuff with the right size if the index deviates from the RANGE.
Page 36
(2) Proper measurement posture ❶ Place the arm so that the center of the cuff is level with the patient's heart position. Height of the heart ❷ Be relaxed during the measurement. Caution • The measurement posture may affect the blood pressure values. •...
Page 37
(4) Battery level indicator ❶ The remaining battery level can be checked in the bottom right corner of the screen while using the blood pressure monitor. ❷ When using the AC adapter, the battery level is not displayed, and instead the AC adapter connection symbol ( ) appears.
Page 38
(7) Depressurization ❶ Rotate the exhaust valve counterclockwise to open the valve. Or you may adjust the exhaust speed by operating the jog switch. Maintain an adequate exhaust rate when conducting a blood pressure test. ❷ When the systolic blood pressure and the diastolic blood pressure are examined by auscultation, and the jog switch or the on the main body is pressed, marking will be displayed on the display bar corresponding to the instantaneous...
• Contact your local InBody representative to purchase the cuff. Remarque • Si le brassard est contaminé, et qu’il ne peut plus être utilisé, veuillez le remplacer par un nouveau brassard. • Communiquez avec le représentant InBody de votre région pour vous procurer des brassards additionnels.
3. BPBIO210/220 Common I. FAQ A. Error code Cause Solution Excess air detected in the Exhaust all the air from the cuff completely, and cuff restart. If the same error message occurs continuously after Pressure sensor exhausting the air from the cuff, the pressure sensor malfunction may be defective.
B. Regarding the Equipment Status Cause Solution Check the battery polarities and reinstall the batteries to match with Check the inserted batteries. ‘+’ and ‘-’ marks. Polarities may not be matching. Refer to the User's Manual (page Power does not 10, page 26), (3) Inserting/replacing turn on.
For arm circumference of 22 to 32 cm (9 to 13 in) BP-C-M-01 S-size cuff For arm circumference of 17 to 22 cm (7 to 9 in) BP-C-S-01 Available for BPBIO210 Air bulb BP-CO-M-01 Does NOT contain the jog switch Available for BPBIO220...
B. Safety Information Safety Symbols Warning/Caution Note Power adapter connection terminal BF-type equipment Operating instructions CLASS II equipment (AC adapter) Other Symbols Disposal guidelines Manufacturer Serial number Authorized representative in the EUROPEAN COMMUNITY European conformity NRTL mark (MET) Unique Device Identifi cation * MET symbols apply to products sold in the United States and Canada.
Page 44
3. Turn off the device completely by taking the batteries out or disconnecting the AC adapter if the device malfunctions. 4. For inspection / repairs, please contact the local InBody representative. 5. Do not disassemble or modify the equipment arbitrarily. Any disassembly or modification will void the warranty and may cause a fire.
4. Pour une inspection / réparation, veuillez contacter le lieu d'achat ou le centre clients InBody (080-501-3939). 5. Ne démontez pas et ne modifiez pas l'équipement de façon arbitraire. Tout démontage ou modification annulera la garantie et pourrait provoquer un incendie.
E. EMC Declaration This EUT is intended for use in the electromagnetic environment specified below. The customer or the user of the EUT assure that it is used in such an environment. Compliance Level Electromagnetic environment - guidance Immunity Test IEC 60601-1-2 Test Level Electrostatic...
Page 48
3 V ͫ ⁾ Portable and mobile RF Conducted 0,15 MHz - 80 MHz communications equipment should be disturbances IEC 61000-4-6 6 V ͫ ⁾ in ISM bands used no closer to any part of the EUT induced by RF between 0,15 MHz image intensifier, including cables, than fields ᶜ⁾...
Page 49
Test specifications for ENCLOSURE PORT IMMUNITY to RF wireless communications equipment Test Maximum Immunity Band Distance power frequency Service Modulation test level (MHz) (MHz) (V/m) Pulse 380 –390 TETRA 400 modulation 18 Hz GMRS 460, ± 5 kHz 430 – 470 FRS 460 deviation 1 kHz sine...