Page 2
Reproduction, adaptation, or translation of this manual is prohibited without prior written consent from InBody Co., Ltd. under the copyright laws. This manual may be printed incorrectly and subject to change without notice. InBody Co., Ltd. shall not be liable for any errors, incidental, or consequential damages that occurred by not complying with the content of the User᾽s Manual.
Page 3
InBody Europe B.V. Gyroscoopweg 122, 1042 AZ, Amsterdam, The Netherlands TEL: +31 (0)20 238 6080 FAX: +31 (0)6 5734 1858 Website: https://nl.inbody.com E-mail: info.eu@inbody.com Australian Sponsor. [AUSTRALIA] Emergo AUSTRALIA. Level 20, Tower II, Darling Park, 201 Sussex Street, Sydney, NSW 2000, AUSTRALIA TEL: +61-2-9006-1662 FAX: +61-2-9006-1010 Website: https://www.emergogroup.com E-mail: Sponsor@emergogroup.com...
BPBIO250 User’s Manual Contents I. Product Introduction II. Blood pressure measurement A. Non-Invasive Pressure Measurment B. Precautions Before Measurement C. Preparation For Measurement D. Checking Blood Pressure Records E. Check Saved Blood Pressure F. User Settings III. Storage IV. Troubleshooting A.
17 cm and 42 cm. - Do not use this blood pressure monitor on pregnant women and newborn babies. * Product Components BPBIO250 ❶ M-sized Cuff ❷...
Page 7
* Main Unit Parts <Front> <Rear> Handle [M] Button [MODE] Button Cuff Rack AC Adapter Connection Cuff Connection [START/STOP] Button [▼] Button [▲] Button <Bottom> Battery Cover...
Page 8
* Cuff Parts Checking mark for the available Arm Circumference Range Indicator Artery Indicator When Worn Air Inlet (To Cuff Connector) Cuff Tube * Display Parts Bluetooth Date Time Memory Mode AC Adapter Connection Battery Level Indicator Sequence Number Mute Measurement Mode Blood Pressure Cuff Status Detection...
Page 9
* Installation for optional stand ∙ The Stand can be installed on the Main Unit. ∙ Various size Cuff can be stored in the Stand Box. ∙ The Stand can be installed in the following sequence. Fix the Main Unit with the bracket. ❶ Fix the Bracket with the Stand using four screws. ❷ ❶ ❷ Note ∙ Make sure the lithium-ion battery is inserted before installing the Stand. ∙ The Bracket and six set screws are enclosed in the Stand Box. ∙ Please check the assembly manual for more details.
Blood Pressure Measurement Method A. Non-Invasive Blood Pressure Measurement Principles Oscillometric method Blood pressure is a measurement of changes in the pressure on the blood vessels. Blood pressure can be measured by direct or indirect measurement. This equipment supports indirect measurement using a cuff. This equipment tracks and analyzes the blood flow oscillations under the pressure using the cuff from the highest to the lowest pressure as shown by the figure below when the minute pressure speed is determined by the heartbeat.
B. Precautions Before Measurement Warning 1. Do not use blood pressure monitor near Magnetic Resonance Imaging (MRI) equipment. If MRI is to be performed, remove the Cuff from the patient. 2. Do not use the blood pressure monitor near the equipment that generates noise, such as equipments which uses frequencies or electric scalpel.
Page 12
20. Do not give strong impact to the main unit. Otherwise, it may cause failure and malfunction. 21. There is a possibility of death due to injury or suffocation if the air tube is wrapped around the neck. Caution 1. Do not use the blood pressure monitor without wrapping the Cuff on the arm. 2.
Page 13
15. Repair and inspection can be done only by the authorized staff of InBody Co., Ltd. For repair and inspection, please contact InBody Co., Ltd. 16. Operation of a radio equipment may interfere with the measurement. 17. Please note the following when using the Bluetooth function of BPBIO250.
Remove the battery cover on the bottom of the product. ❶ Insert the lithium-ion battery in the right direction. ❷ Close the battery cover. ❸ Note ∙ The lithium-ion battery is provided as the basic component of BPBIO250. AA size battery can also be used if a lithium-ion battery is unavailable. Caution ∙ Do not install batteries with wet hands. ∙ Make sure that the polarities (+) and (-) are correct when lithium-ion battery is installed. ∙ Use only dedicated lithium-ion battery for BPBIO250. ∙ Use battery only for the blood pressure measaurments.
Page 15
(2) Inserting/replacing the AA batteries Remove the battery cover on the bottom of the product. ❶ Insert four AA size batteries in the right direction. ❷ Close the battery cover. ❸ Note ∙ AA batteries and lithium-ion batteries cannot be used at the same time. Caution ∙ Do not install batteries with wet hands. ∙ Make sure that the polarities (+) and (-) are correct when batteries are installed. ∙ Use 1.5V AA-size alkaline or manganese batteries. ∙ Do not use new and used batteries together. ∙ Replace all four AA batteries at once when replacing batteries. ∙ Remove the batteries from the battery compartment when it will not be use for long period. ∙ Dispose used batteries at designated locations.
Page 16
Main Unit of the blood pressure monitor. Connect the dedicated AC adapter to the power outlet. ❷ Warning ∙ Do not connect or disconnect the AC Adapter with wet hands. ∙ Do not put heavy objects above the AC Aapter cable. ∙ Do not use the AC Adapter if it is damaged. Caution ∙ Plug the AC Adapter into the wall outlet securely. ∙ When unplugging the AC Adapter, grasp the adapter body and pull it out. ∙ Disconnect the AC Adapter in sequence from wall outlet to BPBIO250. ∙ Do not apply excessive force to AC Adapter body and plug. ∙ Keep the AC Adapter Connector away from foreign objects. ∙ Use only the AC Adapter provided by InBody Co., Ltd. ∙ Do not use it for other purposes except the blood pressure measurement. ∙ Do not place the BPBIO250 in an area where it is difficult to disconnect the AC Adapter. * AC Adapter’s Specification Manufacturer XP Power Limited Model Name...
Page 17
(4) Connecting cuff Insert the Air Inlet of the Cuff into the Cuff Connection on the Main Unit. ❶ Press lightly as shown in A and turn it clock-wist as shown in B to tighten the ❷ connection. Note ∙ To remove the cuff, press the air inlet of the cuff slightly in the direction of the arrow A and turn it in the opposite direction of the arrow B.
Page 18
D. Blood Pressure Measurement (1) Wearing the Cuff Wrap the Cuff around the arm, about 2-3 cm INDEX(▲) ❶ above the elbow. Make sure that the Index (▲) on the Cuff ❷ RANGE is within the RANGE bar, and replace the cuff with the proper size if the index deviates from the RANGE.
Page 19
- The blood pressure monitor will be turned off by pressing and holding the [START/STOP] button. - BPBIO250 provides power saving function If there is no operation within 5 minutes, the blood pressure monitor automatically turns off. The power is turned on by The power is turned off by pressing and holding the button.
Page 20
(5) Charging the Lithium-Ion Batteries - Lithium-Ion Batteries can be charged with Dedicated AC Adapter. - When the Lithium-Ion Batteries are drained, charge it (4 hours for full charge) before using it. Make sure the Lithium-Ion Batteries are inserted in the Main Unit. ❶...
Page 21
(7) Pressurization Value Setting - The user can set the desired pressure value by pressing the [▼] or [▲] button before measuring blood pressure. - The pressure value can be set from Auto, 220, 250 or 280. Changing the pressure value by pressing the [▼] or [▲] button. : Pressurizing automatically : Pressurizing to 220 mmHg : Pressurizing to 250 mmHg : Pressurizing to 280 mmHg Caution...
Page 22
(8) Five measurement modes - Five measurement modes can be selected according to various medical environments in BPBIO250. - The measurement mode can be checked in the upper left corner of the LCD screen. - Each measurement mode can be used in the following situations.
Page 23
* Changing Measurement Mode - Measurement mode can be changed by pressing the [MODE] button. Changing measurement mode by pressing the [MODE] button - Measurement Mode Changes in Sequence: NORMAL → AUSCULT → CYCLE → CYCLE RANGE → CYCLE HIDE → NORMAL. NORMAL MODE AUSCULT MODE CYCLE MODE CYCLE HIDE MODE CYCLE RANGE MODE...
Page 24
* NORMAL Mode - NORMAL Mode is used for normal blood pressure measurement. <Measurement Order> Make sure that the Cuff is correctly worn and the measurement posture is correct again. ❶ Press the [START/STOP] button of the blood pressure monitor to start blood pressure ❷...
Page 25
* AUSCULT Mode - This mode is used when measuring blood pressure with stethoscope. - This mode can replace aneroid or mercury blood pressure monitor. - The systolic and diastolic blood pressure can be recorded through marking during exhaust. - Manual pressurization is possible during exhaust. <Measurement Order>...
Page 26
Note ∙ Month / Day Hour: Minute is displayed only when date and time are set. ∙ For more information on how to set the date and time, refer to “F. User Settings”. ∙ AUSCULT mode does not automatically measure blood pressure. ∙ AUSCULT mode does not detect heartbeats. ∙ In AUSCULT mode, pulse is not displayed. ∙ In AUSCULT mode, motion and irregular pulse wave are not detected. ∙ If no marking is performed, the measurement is automatically terminated at a certain pressure or less. - The AUSCULT mode allows manual pressurization to be performed at the desired pressure during exhaust in case the patient cannot hear. Press and hold the [▲] button during exhaust. ❶ The manual pressure indicator is displayed while pressing this button, and pressurization begins. ❷...
Page 27
* CYCLE Mode - In CYCLE mode, the patient can measure blood pressure repeatedly. - The number of measurements and standby time can be set. - If the measurement count is set to 3 times and the standby time is set to 60 seconds, the operation in the CYCLE mode will be performed shown below.
Page 28
1st Measurement Start Pressurize Exhaust Completed Setting Mode Start Reducing Standby Time Start and Finish 2nd Measurement Last Measurement Start Pressurize Exhaust Completed Note ∙ The next measurement can be started immediately by pressing the [START / STOP] button when the standby time is reduced. ∙ All remaining measurements can be canceled during measurement by pressing the [START/STOP] button.
Page 29
- In the CYCLE Mode, the previous measurement results can be checked. - If all measurements are completed for the set number of times, the previous measurement result can be checked by pressing the [▼] or [▲] button. Measurement Completed Check the Previous Measurements. Note ∙ The previous measurements can only be checked when all the measurements have been completed for the set number of times.
Page 30
* CYCLE RANGE Mode - In the CYCLE RANGE Mode, the patient can measure his / her blood pressure repeatedly by utilizing the RANGE function. (RANGE function notifies the user when the measured value is out of the average measurement range.) - The previous measurement result can be checked after the measurement is completed.
Page 31
Note ∙ The RANGE value and the current pressure value will alternately displayed on the screen. ∙ The current number of measurements / the total number of measurements and the standby time are displayed in date and time display field. For more information on how to set the number of measurements, standby time, and RANGE, refer to “F. User Settings”. [Measurement Completed Screen] In Case the Blood Pressure is Out of RANGE Note ∙ The previous measurement can be checked by pressing the [▼] or [▲] button after all measurements are completed.
Page 32
* CYCLE HIDE Mode - In the CYCLE HIDE mode, the patient can measure his / her blood pressure repeatedly by utilizing the HIDE function. (HIDE function: It does not display measurements on screen.) - The measurement operation is the same as the CYCLE mode. - The measurement result can be checked in the MEMORY Mode after the measurement is completed.
Page 33
(9) Checking Measurement Results Checking the Blood Pressure Basic Information ❶ Measurement Date and Time Systolic Blood Pressure Measurement Mode Diastolic Blood Pressure Pulse Pressurization Value Checking the Status of the Cuff, Movement, and Irregular Pulse ❷ Detection of Cuff Status: The Cuff was loose.
Page 34
E. Checking Saved Measurements (MEMORY MODE) (1) Press the [M] button to check the saved measurements. (2) Sequence number and measurement date and time are displayed. (3) Press the [▼] or [▲] button to check the saved results one by one. : Go to Past Measured Results : Go to Recent Measured Results Order Based on Date and Time the Sequence (4) Recorded results in CYCLE RANGE mode are displayed as below.
Page 35
(5) Press the [M] button again to exit the MEMORY Mode. Note ∙ Up to 99 measurements can be saved. ∙ Press and hold the [▼] or [▲] button to check quickly the past or recent results. ∙ Month / Day Hour: Minute is displayed only when date and time are set. ∙ ‘--’ is displayed if there is no saved blood pressure. ∙ The saved blood pressure measurement value can be deleted. For more information on how to delete the results, refer to “F. User Settings”.
E. User Settings - Press and hold the [MODE] button to set the measurement mode and function. - The settings are displayed - Press the [▼] or [▲] button to change the settings and press the [START/STOP] button to set it. - Press and hold the [MODE] button again to exit the user setup screen. Press the [▼] or [▲] Button to Change the Settings.
Page 37
(2) F01: Setting Number of Repeated Counts - The number of repeated measurements can be set in the CYCLE (RANGE, HIDE) mode. <Measurement Order> Press and hold the [MODE] button for 3 seconds to enter the User Setting Mode. ❶ Press the [▼] or [▲] button to set .
Page 38
(3) F02: Setting Standby Time - The standby time between the repeated measurements can be set in the CYCLE (RANGE, HIDE) Mode. <Measurement Order> Press and hold the [MODE] button for 3 seconds to enter the user settings mode screen. ❶...
Page 39
(4) F03: Setting Systolic Blood Pressure Range - The systolic blood pressure range can be set in the CYCLE RANGE Mode. <Measurement Order> Press and hold the [MODE] button for 3 seconds to enter user settings. ❶ Press the [▼] or [▲] button to set . ❷ Press the [START/STOP] button to enter the maximum value (S-H) for systolic ❸...
Page 40
(5) F04: Setting Diastolic Blood Pressure Range - The diastolic blood pressure range can be set in the CYCLE RANGE mode. <Measurement Order> Press and hold the [MODE] button for 3 seconds to enter user settings mode screen. ❶ Press the [▼] or [▲] button to set .
Page 41
(6) F05: Setting Mute - If the user doesn't want the sound, it can be muted. <Measurement Order> Press and hold the [MODE] button for 3 seconds to enter under settings. ❶ Press the [▼] or [▲] button to set . ❷ Press the [START/STOP] button to enter the sound settings. ❸...
Page 42
(7) F06: Deleting Memory - The saved blood pressure measurement can be deleted. <Measurement Order> Press and hold the [MODE] button for 3 seconds to enter user settings. ❶ Press the [▼] or [▲] button to set . ❷ Press the [START/STOP] button to enter the delete settings. ❸...
Page 43
(8) F07: Setting Date and time - The user can set the date and time. <Measurement Order> Press and hold the [MODE] button for 3 seconds to enter user settings. ❶ Press the [▼] or [▲] button to set . ❷ Press the [START/STOP] button toenter the Date and Time settings. ❸...
Page 44
(9) F08 : Setting Fast / Slow Precision measurement - Slow Precision measurement mode ensures the measured results even if there was a disturbance such as the movement of a patient. But, the measurement speed will be slower than a Fast measurement mode. <Measurement Order>...
(3) Keep the cuff on the cuff rack in a flat state. Note ∙ To unfold the cuff, hold the both sides of the cuff and stretch them in the opposite direction of the cuff curl. The BPBIO250 should be transported or stored under the following conditions. Temperatures Range - 10 ~ 70˚C Relative Humidity 10 ~ 85% RH (No Condensation)
40 mmHg higher than the expected blood pressurization excessively pressure.) · Severely moves · Measure the blood pressure in a stable state. Please contact the InBody Customer Center It is a failure due to malfunction (+82-80-501-3939). Charge the lithium-ion batteries or replace Low battery voltage...
Page 47
B. Troubleshooting Problem Cause Measure Check lithium-ion battery or The lithium-ion batteries or AA AA battery polarities and reinstall There is no batteries are incorrectly inserted the batteries to match with response even ‘+’ and ‘-’ marks. the button was pressed.
Others A. Optional Equipment - S, L Sized Cuff (S: 17-22 cm only, L: 32-42 cm only) - AA Batteries - Stand B. Safety Information Safety Symbols Power Adapter Connection Terminal BF-Type Equipment Operating instructions CLASS II equipment (Dedicated AC adapter) Other Symbols Disposal Guidelines Disposal Guidelines (Lithium-ion battery)
Page 49
Disposal of old Electrical & Electronic Equipment (Application in the European Union and other European countries with separate collection system.) This symbol indicates that this product shall not be treated as household waste. Instead, it shall be handed over to the applicable collection point for the recycling of electrical and electronic equipment.
- Level of protection against flooding: General Equipment (No special protection against external water infiltration) - Operation mode: Continuous operation - Level of protection against electric shock: BF-type equipment D. Specifications Product Name BPBIO250 Display Method Digital Display Method (Custom Segment LCD) Test Instructions Oscillometric Test range...
Page 51
Manufacturing Country Republic of Korea Manufacturing Company InBody Co., Ltd. * For purposes of improvement, specifications are subject to change without notice. * This product is a ‘medical equipment’. Please read the guide and cautions before using this blood pressure monitor.
E. EMC Declaration This EUT is intended for use in the electromagnetic environment specified below. The customer or the user of the EUT assure that it is used in such an environment. IEC 60601-1-2 Immunity Test Compliance Level Electromagnetic Environment - Guidance Test Level Floors should be wood, concrete or Electrostatic...
Page 53
Portable and mobile RF Conducted 0.15 MHz - 80 MHz 0.15 MHz - 80 MHz communications equipment should be disturbances 6 V in ISM bands 6 V in ISM bands used no closer to any part of the EUT, induced by RF between 0.15 MHz between 0.15 MHz including cables, than the recommended...