QUANTiC readhead installation drawing

Reference mark selector magnet
P limit magnet
Selected IN-TRAC
reference mark
Reference mark selector sensor position
R > 30 Dynamic bend radius
R >   10 Static bend radius
4.75
*
 
  Extent of mounting faces.
Dimension from substrate.
QUANTiC RKLC40-S linear installation guide
18
29
7.8
Forward direction of readhead
relative to scale
6 min
35
23
11.5
Optical centreline marker
Optical centreline
(Incremental and reference mark)
Offset 3.75 ±0.5
7.8
Set-up LED
P and Q limit switch sensor position
A
(Pitch tol. ±1°)
2 mounting holes M2.5 through, counterbored Ø3 × 2.3 deep both sides.
NOTE: The recommended thread engagement is 5 min (7.5 including counterbore)
and the recommended tightening torque is between 0.25 and 0.4 Nm.
(Yaw tol. ±0.9°)
0.55
(Roll tol. ±0.8°)
0.12
Mounting faces
Q limit magnet
Detail A
Calibration rideheight: 2.1 ±0.2
Operating rideheight: 2.1 ±0.3
0.6
Dimensions and tolerances in mm
*
8.75
13.5
4.25
4.15
10
Scale reading surface
Scale thickness 0.15
(including adhesive)
3