GE XMO2 User Manual page 80

Thermoparamagnetic oxygen transmitter

Advertisement

PRESSURE CURVE Option
(cont.)
General Programming
If you selected Numeric Entry as the Calibrate Mode:
At the mmHg mV prompt, enter the listed pressure signal mV value
corresponding to the mmHg value entered at the previous prompt.
Obtain this data from your Calibration Sheet (see page 8-6 for a
sample calibration sheet).
Enter Pressure Curve
mmHg mV [-xxx.x]:
The mmHg and mmHg mV prompts repeat until the total number of
data points specified at the # Points prompt have been entered. This
completes the programming of the Pressure Curve option, and the
XMO2 returns to the Calibrate System menu.
If you selected Introduce Gases as the Calibrate Mode:
At the mmHg prompt, enter the pressure that you intend to use.
Enter Pressure Curve
mmHg [xxx.x]:
Introduce xxx.x mmHg
Press [YES] when ready.
Press [YES] when settled.
-xxx.x mmHg mV
The XMO2 displays working... as it records the calibration data, and
then repeats the mmHg and Introduce xxx.x mmHg prompts for the
number of points specified at the # Points prompt. This completes the
programming of the Pressure Curve option, and the XMO2 returns to
the Calibrate System? menu.
September 2007
To accept the current value, press
[Y]
or
[Enter]
. To change the
current value, use the numeric
keys to enter a new value and
press
[Y]
or
[Enter]
twice.
To accept the current value, press
or
. To change the
[Y]
[Enter]
current value, use the numeric
keys to enter a new value and
press
or
twice.
[Y]
[Enter]
Apply the specified pressure,
then press
or
.
[Y]
[Enter]
The mV value corresponding to
the above mmHg value is shown.
When the displayed value
stabilizes, press
[Y]
or
[Enter]
.
6-13