GE XMO2 User Manual page 69

Thermoparamagnetic oxygen transmitter

Advertisement

September 2007
Entering the General
Menu
General Menu Options
6-2
To switch the XMO2 from Operate Mode to Programming Mode,
proceed as follows:
xx.xx %O2
Note: If the XMO2 is set up for Easy Menu Entry, pressing
switches the XMO2 directly to Programming Mode and enters
the General Menu. See Chapter 7, Advanced Programming,
for instructions on setting up Easy Menu Entry.
Enter Code:
Note: An * is displayed after each digit is entered. If you enter an
incorrect code, press a non-numeric key, or press the keys too
quickly, the XMO2 will return to Operate Mode.
GENERAL MENU
Calibrate System?
The General Menu consists of the following six menu options:
Calibrate System? - The factory calibration data may be accessed
via this menu option.
Calibrate Recorder? - The default analog output high and low
values may be accessed via this menu option.
Test Inputs? - For diagnostic purposes, various input parameters
may be accessed via this menu option.
Test Recorder? - For diagnostic purposes, the analog output values
may be accessed via this menu option.
Set Error Handling? - the manner in which the XMO2 responds to
various error conditions is specified in this menu option.
Resume? - Use this menu option to either return to Operate Mode
or return to the Calibrate System? menu option.
Refer to Figures C-2 through C-6 in Appendix C, Menu Maps, and
proceed to the appropriate section to program the desired General
Menu option.
Press
or
.
[Y]
[ENTER]
[Shift]+[2]
Press the
,
, and
[3]
[6]
[9]
slowly and firmly, in sequence.
You now have access to the
General Menu.
General Programming