5
Replacing the circuit board
Spare part
Circuit board BVC control / professional (612770)
Tool
Torx screw driver TX20
Tweezers
DANGER
DANGER
Danger of electric shock.
Improperly executed repairs will result in an electrical
shock.
Check the electrical safety of the device according to
>
IEC 61010 and national regulations after repair.
Check the protective conductor resistance.
>
Check the insulation resistance.
>
Carry out a high voltage test.
>
CAUTION
Electrostatic discharge (ESD).
Damage of electronic components.
Do not touch electronic components.
>
Connect the person in charge to ground
>
appropriately.
Replacing the circuit board
19