Replacing The Main Switch - vacuubrand BVC Basic Instructions For Repair

Vacuum aspiration systems
Hide thumbs Also See for BVC Basic:
Table of Contents

Advertisement

Replacing the main switch

14
4
Replacing the main switch
Spare part
Main switch, blue (635239)
Tool
Torx screw driver TX20
DANGER
DANGER
Danger of electric shock.
A defective main switch will result in an electrical shock.
Replace a defective main switch immediately.
>
Replacement of the main switch must be carried out 
>
by a qualified electrician.
DANGER
DANGER
Danger of electric shock.
Improperly executed repairs will result in an electrical 
shock.
Check the electrical safety of the device according to 
>
IEC 61010 and national regulations after repair. 
Check the protective conductor resistance.
>
Check the insulation resistance.
>
Carry out a high voltage test.
>
CAUTION
Electrostatic discharge (ESD).
Damage of electronic components.
Do not touch electronic components.
>
Connect the person in charge to ground 
>
appropriately.