Cardiohelp System
Priming the System
17 Open the Clamp on Blue Line to passively prime the Set.
18 Once the flow stops, set speed to 3000 rpm for 2 min.
5.62
-27
221
16
C A R D I O H E L P S E T U P G U I D E
3000
205
20.8
-----
19 Reduce speed to 0 rpm for at least 5 seconds. Reset the
flow/bubble sensor. Tap the tube and the HLS Module
with your hand to remove gas bubbles.
-0.04
Ven
-0.04
3,000
rpm
2
min
20 Set the speed to 4000 rpm for 1 minute and then reduce
speed to 0 rpm. De-air the 3 Luer lock connectors. If a
bubble is detected, go back to step 18.
7.71
-107
326
Ven
0
77
81
80
--
Art
0
4000
300
--
Art
4,000
rpm
1
min
6