5.2.11 SET ALARM LEVEL FOR CPS/CPM MODES
When the user intends to use the meter to make contamination measurements in steel raw material, scrap
or finished steel goods, alarm level can be set to detect contamination by means of audio/visual alarms
Preset level can be chosen by going to below menu option by pressing PROG button
Alarm level is chosen as a factor of background radiation level.
When acquiring in CPS mode, choose alarm level in the range of 1.1 to 1.5 of BG keep alarm level as low as
possible at a level where there is no occurrence of false alarms when acquiring for natural radiation levels
(without scrap/material)
For CPM mode, choose a factorin the range of 1.5 to 1.3 of BG. Select criteria for alarm level is same as that for CPS
mode of acquisition.
5.2.12 ERASE MEMORY?
ALARM LEVEL
DISABLED
Now using or buttons enable/disable the alarm level
If user wants to erase the stored memory press stop key & then PROG to get
ERASE parameter then press increment or decrement
ERASE
OK
MEMORY?
Then again go for START to acquire in selected mode.