Auto Power Off Disable - Lutron Electronics TM-946 Manual

4 channels thermometer, t1, t2, t3, t4, t1-t2, type j/k, pt 100 ohm
Table of Contents

Advertisement

* Use the "
Fig. 1 ) to adjust the desiring value on right bottom
display.
* Pressing " Enter button " ( 3-4, Fig. 1 ) once, both
bottom display will flashing 4 times then return to
normal measuring screen and finish the offset
adjustment procedure.
* The above offset adjustment for Pt 100 ohm is
valid for PT1, PT2 individually.

8. AUTO POWER OFF DISABLE

The meter is built the " Auto power shut off " to saves
battery life. If any buttons not be pushed within 10
minutes, then the power will be off automatically.
If the user is not intend to execute the " Auto Power off "
function, it should take the following procedures :
During the measurement, push the " Record
Button" ( 3-4, Fig. 1 ) to execute the memory
record function, the auto power function will
disable.
button "( 3-5, Fig. 1 ) "
18
button " ( 3-7,