Page 1
User Guide Control Room Unit for Model 9100 Carbon Monoxide Monitor Publication Number: 807248 Language: English Issue: 2 January 2013 Combustion and Environmental Monitoring www.landinst.com...
Page 2
IMPORTANT INFORMATION - PLEASE READ Health and Safety Information Read all of the instructions in this booklet - including all the WARNINGS and CAUTIONS - before using this product. If there is any instruction which you do not understand. DO NOT USE THE PRODUCT. Safety Signs WARNING Indicates a potentially hazardous situation which, if not avoided, could result in death or...
Page 3
Contact Us UK - Dronfield USA - Pittsburgh Land Instruments International AMETEK Land, Inc. Tel: +44 (0) 1246 417691 Tel: +1 412 826 4444 E-Mail: land.combustion@ametek.co.uk E-Mail: combsales@ametek.com Web: www.landinst.com Web: www.ametek-land.com For further details on all LAND/Ametek offices, distributors and representatives, please visit our websites.
Contents User Guide Introduction Applications and Usage Dimensions of Control Room Unit Dimensions of Cutout Optional Rack Mounting Power connections Location, Assembly and Installation Instructions Connecting the Control Room Unit to the Carbon Monoxide Monitor Direct Connection to Transceiver Customer Connections Control Panel Password Entry Menu Tree...
User Guide Control Room Unit for Model 9100 Carbon Monoxide Monitors Introduction Applications and Usage The control room unit is intended for use with the Land Model 9100 Carbon Monoxide Monitor. The control room unit is not suitable for use with any other equipment. The control room unit must be protected from dust and moisture.
Control Room Unit for Model 9100 Carbon Monoxide Monitors User Guide Dimensions of Control Room Unit gasket The diagram shows the optional input/output unit installed. Dimensions of Cutout 260 (10.24) front view earth ø4 (0.15) 35 (1.38) 5 (0.19) earth assembly supplied with control room unit...
User Guide Control Room Unit for Model 9100 Carbon Monoxide Monitors Optional Rack Mounting The rack-mount option allows the Control Room Unit to be fitted into a standard 19” rack. The mounted CRU is 4U (176 mm, 6.9”) high. This option is supplied fully-assembled with its power supply and all dc connections.
Control Room Unit for Model 9100 Carbon Monoxide Monitors User Guide Location, Assembly and Installation Instructions Do NOT install the control room unit where there is conductive dust, corrosive or flammable gas, moisture or rain, excessive heat and shocks or excess vibration Do not allow debris to fall inside the unit during installation.
User Guide Control Room Unit for Model 9100 Carbon Monoxide Monitors Electrical Connections DIGITAL INPUTS ANALOG INPUTS T- PT 24VDC T+ PT AN ANALOG T- PT AOUT0 T+ PT AN AOUT1 ANALOG T- PT OUTPUTS AOUT2 T+ PT AN AOUT3 T- PT 24VDC T+ PT AN...
Control Room Unit for Model 9100 Carbon Monoxide Monitors User Guide Control Panel function keys touch sensitive screen scroll keys escape symbol pqrs wxyz key pad enter Function Keys Password lock. Pressing F8 causes the password to reset. The password must be entered again before any data can be changed.
User Guide Control Room Unit for Model 9100 Carbon Monoxide Monitors Password Entry The password is 417 Enter the password from the key pad and then press the enter key. Please Enter Password: symbol XXXX Press ESC to Return To Main Menu pqrs wxyz Once entered the password remains valid for 10...
Control Room Unit for Model 9100 Carbon Monoxide Monitors User Guide Menu Tree CO Reading Large text display Display units (ppm or mg/m3) match those selected on the Model 9100 Parameters Parameters page 1 Parameters page 2 Parameters page 3 Parameters page 4 Parameters page 5 Faults...
User Guide Control Room Unit for Model 9100 Carbon Monoxide Monitors Technical Specification System Type: (Remote Display) Control Room Unit Compatibility: Model 9100 Communications: Modbus over RS485, 2 Wire Connection Installation: Up to 1200 m / 4000 ft from the analyser Displays Display: Touch-screen LCD with LED Backlight...
Control Room Unit for Model 9100 Carbon Monoxide Monitors User Guide Setting Configurable Parameters The Modbus address on the Model 9100 CRU must be the same as that on the Model 9100 analyser. At the CRU Select SETUP then set the following parameters 9100 Modbus address = 1 O/P1 range O/P2 range...
User Guide Control Room Unit for Model 9100 Carbon Monoxide Monitors Calibration The calibration check and zero calibration functions can be initiated from the Control Room Unit. The password 417 is needed to access the calibration menus. After entering the correct password, an “unlocked”...
Control Room Unit for Model 9100 Carbon Monoxide Monitors User Guide Parameter Table Parameter Table The instrument parameters can be displayed using DIAGNOSTICS > PARAMETERS Parameters page 1 Password number Password Timeout Maintenance Relay Units (ppm or mg/m3) Temperature units (°C or °F) Pressure units Pathlength units (inches or cm) Parameters page 2...
Page 19
Gas Monitoring Products Standard Terms of Warranty Land Instruments International warrant all products of our own manufacture to be within specified limits of calibration, if any, when despatched from our works. Also that they are free from defects in material and workmanship for normal use and service when used for their intended purpose within the limits of our specification.