10.7 Peak value function

This function displays the highest load value (peak value) of a weighing.
The peak value remains in the display until it will be deleted.
 In weighing mode press MENU button. The first menu item „count" is displayed.
 Press MENU button repeatedly
 Confirm using PRINT button, the taring process is automatically started. „0-t" is
displayed.
From here the balance is in peak value mode, where a „M" pops up.
 Load weighing plate. The maximum load value is displayed.
 The peak value remains in the display until the TARE button is pressed. Then
the balance is ready for further measurements.
Return to weighing mode