#dm0 (E)
#dm1 (E)
#sa (E)
#vm0 (E)
#vm1 (E)
#vm2 (E)
#vm3 (E)
#vm4 (E)
#br4800 (E)
#bm1 (E)
#rc0 (E)
#sv0 (E)
Special commands for emc measurement in combination with zero-span:
#es0 (E)
#es1 (E)
#ss1 (E)
#ss2 (E)
After receiving and processing the command, the instrument transmitts „RD(hex:
0x0d)".
Polling of Parameters:
Syntax:
#xx (E)
1st example:
#uc (uncalibrated): PC transmitts #uc (E), instrument replies:
2nd example:
#tl (E)
3rd example: sequence control via external computer:
#kl1 (E)
Subject to change without notice
=
Detect mode off
=
Detect mode on
=
Save signal A as signal B
=
View mode: Signal A
=
View mode: Signal B (stored signal)
=
View mode: Signal A - B
=
View mode: Average
=
View mode: Max. Hold
=
baudrate 4800 (9600, 19200, 38400, 115200 ) Bd
=
transfer signal in block mode (2048 bytes)
2044 signal bytes, 3 checksum bytes and (0x0D)
=
Recall (0 to 9)
=
Save (0 to 9)
=
1 second sweep disable
=
1 second sweep enable (measuring time 1 second; select
zero-span and bandwidth 9/120/400 kHz)
=
Start 1 second sweep at a selected center frequency
=
Start 1 second sweep at a selected center frequency (i.e
100MHz) + frequency step(s) (i.e. 100.12MHz, 100.24MHz,
...). step = bandwidth: 400, 120, 10 (9) kHz
=
Send parameter xx (xx = tg, tl, rl, vf, at, bw, sp, cf, db, kl,
hm, vn, vm, uc)
uc0 (calibrated), uc1 (uncalibrated)
=
PC query for Tracking Level: PC transmitts #tl(E), instrument
replies: TL-12.4(E)
=
switches to remotes status, RM LED is lit
Appendix A
39