5 Inspiratory channel (maintenance)
5.1 Preventive maintenance
Important: It is recommended that a regular
cleaning and an extended cleaning of the
inspiratory channel should be performed
before carrying out preventive maintenance.
5.2 Preparations
•
Remove the expiratory cassette.
•
Remove the front cover.
•
Remove the O
cell.
2
Refer to section Remove the O
page 15.
5.2.1 Remove inspiratory channel and tube
SERVO-air Ventilator System, User's Manual
•
•
5.3 Disinfection
5.3.1 Rinse before disinfection
If rinsing before disinfection is not included in
the washer disinfector program:
•
cell on
2
•
Important: Immediate rinsing in water
(<35°C/95°F) can remove particles and
reduces the risk of cross-contamination
between patients.
|
Inspiratory channel (maintenance)
Press the latch and gently pull out the
inspiratory channel.
Disconnect the tube and remove the
bacterial filter. Discard the filter to
hazardous waste.
Rinse the parts thoroughly in water
(<35°C/95°F).
Let the water flow through the parts.