Power Saver; Zeroing With Absolute Pressure Modules - Fluke 718Ex 30G User Manual

Pressure calibrator
Hide thumbs Also See for 718Ex 30G:
Table of Contents

Advertisement

Power Saver

The Calibrator automatically turns off after 30 minutes of
inactivity. To reduce this time or disable this feature:
1.
With the Calibrator OFF, press .
P.S. xx is displayed, where xx is the turn-off time in
minutes. OFF means the power saver is disabled.
2.
Press
to decrease or
time.
3.
To disable, press
The Calibrator resumes normal operation after 2 seconds.
to increase the turn-off
until the display shows OFF.

Zeroing with Absolute Pressure Modules

Zeroing with Absolute Pressure Modules
For zeroing, adjust the Calibrator to read a known
pressure. This can be barometric pressure, if it is
accurately known. An accurate pressure standard can also
apply a pressure within range for any Absolute Pressure
Module. Adjust the Calibrator reading as follows:
1.
Press and hold
.
2.
Press
to increase or
Calibrator reading to equal the applied pressure.
3.
Release
to exit the zeroing procedure.
Press the
button to convert to any convenient
measurement display unit.
Pressure Calibrator
to decrease the
11