Advertisement

718 Series
Product Overview

Getting Aquainted with the Calibrator

O
Press
to turn the Calibrator on and off. The Calibrator
displays pressure and current measurements
simultaneously. See Figure 2.
The upper part of the display shows the applied pressure
or vacuum. (Vacuum is shown as a negative value.) Press
U
to select a different unit. When you cycle the power
off and on, the Calibrator retains the unit you last used.
The lower part of the display shows the current (up to 24
mA) applied to the current (mA) inputs.
To source loop voltage, press

Power Saver

The Calibrator automatically turns off after 30 minutes of
inactivity. To reduce this time or disable this feature:
With the Calibrator OFF, press O.
1.
2.
P.S. xx is displayed, where xx is the turn-off time in
minutes. OFF means the power saver is disabled.
Press h to decrease or a to increase the turn-off
3.
time.
To disable, press h until the display shows OFF.
4.
The Calibrator resumes normal operation after 2
seconds.
6
U
O
while pressing
on.
measurement
Current mA
measurement
On/Off Button
Pressure
sensor input
(Install filter here)
Figure 2. Front Panel Features
Pressure
Pressure
module input
Current
input
wh005f.eps