Specifications; Pressure Sensor Input - Fluke 718Ex 30G User Manual

Pressure calibrator
Hide thumbs Also See for 718Ex 30G:
Table of Contents

Advertisement

Specifications

Specifications are based on a one year calibration cycle
and apply for ambient temperature from +18 °C to +28 °C
unless stated otherwise. Counts are the number of
increments or decrements of the least significant digit.

Pressure Sensor Input

Model
Range
-12 psi to 30 psi
30G
(-83 kPa to
207 kPa)
-12 psi to 100 psi
100G
(-83 kPa to
690 kPa)
-12 psi to 300 psi
300G
(-83 kPa to
2068 kPa)
Temperature coefficient: 0.01 % of range per
°
°
ranges -10
C to 18
C and 28
Where in a 3 V/m radiated EM field
Accuracy is 1 % of range.
Max Non-
Accuracy
destructive
Pressure
±0.025 % of
60 psi
range
(413 kPa)
(6 month)
±0.035 % of
200 psi
range
(1.4 mPa)
(1 year)
±0.05 % of
375 psi
range
(2.6 mPa)
(1 year)
°
C for temperature
°
°
C to 55
C.
350 MHz, Pressure
Note
Specifications apply to the Product with firmware
version 2.0 or greater. To verify the firmware
version, push and hold
Calibrator.
Pressure Calibrator
Specifications
and power on the
25

Advertisement

Table of Contents
loading

This manual is also suitable for:

718ex 100g718ex 300g

Table of Contents