718Ex 30G/100G/300G
Users Manual
 Warning
To prevent fire, explosion, or personal injury:
Use the Calibrator only as described in
this User Manual and the Fluke 718Ex
CCD (Concept Control Drawing) or the
protection provided by the Calibrator may
be impaired.
Inspect the Calibrator before use. Do not
use it if it appears damaged.
Check the test leads for continuity,
damaged insulation, or exposed metal.
Replace damaged test leads.
When using probes, keep fingers behind
the finger guards on the probes
Never apply more than 30.0 V between the
input terminals, or between any terminal
and earth ground.
Applying more than 30.0 V to the input
terminals invalidates the Calibrator's Ex
Approval and may result in permanent
damage to the unit so it can no longer be
used.
4
Use the proper terminals, mode, and
range for the measuring or sourcing
application.
To prevent damage to the unit under test,
be sure the Calibrator is in the correct
mode before connecting the test leads.
Never use the Calibrator with the red
holster removed.
Precautions are required to ensure that a
charge-generating mechanism is unlikely
to be present and/or discharge to earthed
metal is improbable. The exposed metal
parts are not earthed and have a
capacitance of more than 3 pF with
respect to an earthed conductor. If a
charge-generating mechanism is present,
an incendive level of charge could
migrate to these metal parts and
subsequently discharge to earthed metal.
Precautions are required to ensure that a
charge-generating mechanism is unlikely
to be present and/or discharge to earthed
metal is improbable.