Menu-Measure Settings - Pro's Kit MT-7610 Series User Manual

Optical time domain reflectometer
Hide thumbs Also See for MT-7610 Series:
Table of Contents

Advertisement

2018-06-01
FRI
17:25
Files
Measure Set.
System Set. Module
Wavelength
Scan Mode
Range
Pulse Width
Average Time
Splice Loss
11
Reflection Threshold
End Threshold
Reflective Rate
Scatter Coefficient
Optical Detector
Real time analyse

Menu-Measure settings

Measure settings menu is used to set relative measurement data, which The judgment
About
of events list is based on. Wrong setting might leads to wrong or missing events.
1550nm
Wavelength---wavelength of laser.
Manual
20Km
Scan mode---Manual and auto mode. Under auto mode, it will match the Distance,
range and pulse width .
1us
15s
Range---match with the length of measured optic fiber, usually over one level.
0.05dB
Pulse width---set the pulse width of output laser. Usually, small pulse width can measure
65.0dB
close event, large pulse width can measure remote distance,but enlarge event's blind area.
3.0dB
Average time---can set between 5 second and 180 second as average measurement time.
Splice loss---treat as an event when the loss is higher than setting value.
1.46832
52.1
Reflection threshold---treat as an event when the reflection is higher than setting value.
Open
End threshold---treat as the end of optic fiber when the loss is higher than setting value.
Close
Reflective rate---represent the average refractive index of entire optic fiber.
Scatter coefficient--- the intrinsic value of Rayleigh Scattering.
Optical detector---set whether to check light input in the fiber before measurement or not.
Real time analyse---set whether to analyses events after real time measuring or not.
15