BodiLink Lateral Pelvic/Thigh Support

LPTS

INSTALLATION - SUPPORT PAD TO MOUNTING HARDWARE

All required tools are provided.
Note: In some cases, the support pad will come installed
on the hardware.
1
Unzip and remove the cover and foam insert to
gain access to the metal tracks (A) for hardware
installation. Remove the plastic cap (if installed).
Insert the two square nuts (B) into each of the two
pad tracks (A).
2
Line up the bolts (C), washers (D), and slots on the
mounting bracket (E) with square nuts (B) in track
(A).
Recommendation: When installing a hardware
assembly on a support pad, install the hardware close
to the center of the support pad (depth).
3
Tighten the bolts (C) using the 5mm hex key.
Reinstall the plastic cap (if provided).
4
Replace the cover by inserting the foam insert in
the cover with the gripping side of the foam facing
the support pad, and then add in the support pad
and zip the cover closed.

INSTALLATION - POWER MOUNT HARDWARE

Attaching hardware to the seat frame tracks. All required tools are provided.
Note: In some cases, the support pad will come installed on the hardware.
1
Loosen the screws (A), connecting the hardware
mounting block (B) to the double nut(s) (C) using the
5mm hex key. A gap between the mounting block
and the double nut(s) is needed to slide into the
seat frame track.
2
Attach the hardware mounting block (B) to the seat
frame track by sliding the double nut(s) (C) inside of
the seat frame track (requires temporary removal
of a cap at the end of the track). Slide the hardware
to the general location on the seat frame track for
the desired support pad and hardware placement.
3
Do not fully tighten the screws (A). Continue with
support pad and hardware ADJUSTMENT and USE
instructions in this manual.
Note: For removable hardware, the release lever can be
installed towards the front or rear of the wheelchair,
pointing either up or down, depending on user/caregiver
preferences.
Removable Mounting Block shown
Release Lever
BodiLink on Permobil Corpus Operation Manual
3.5" x 8" (8.9cm x 20.3cm) Support Pad and
Power Mount Hardware shown
Track Slots
11
A
E
B
B
C
A
D
C
LPTS

Advertisement

Table of Contents
loading

This manual is also suitable for:

Bodilink ltsBodilink lpts