An expandable vertebral body replacement device
Table of Contents

Advertisement

Implant repositioning (optional)
Instrument
389.201
Disconnecting Instrument
If it is necessary to reposition the implant, use the disconnecting
instrument to collapse the expanded implant to its starting
position. Introduce the disconnecting instrument into the slot
between the two ends of the locking ring and turn it
one-quarter (1/4) of a revolution. Remove the implant using
the implant holder.
Note: Do not reuse Synex implants once they have been
implanted or expanded.