EC1000S
AC and DC Power Source
User Manual
Contact Information
Telephone: 800 733 5427 (toll free in North America)
858 450 0085 (direct)
Fax: 858 458 0267
Email:
Domestic Sales: domorders.sd@ametek.com
International Sales: intlorders.sd@ametek.com
Customer Service: service.ppd@ametek.com
Web:
www.programmablepower.com
March 2011
Document No. 4994-975 Rev. B

Advertisement

Table of Contents
loading

Summary of Contents for Ametek EC1000S

  • Page 1 User Manual Contact Information Telephone: 800 733 5427 (toll free in North America) 858 450 0085 (direct) Fax: 858 458 0267 Email: Domestic Sales: domorders.sd@ametek.com International Sales: intlorders.sd@ametek.com Customer Service: service.ppd@ametek.com Web: www.programmablepower.com March 2011 Document No. 4994-975 Rev. B...
  • Page 5: Contact Information

    About AMETEK AMETEK Programmable Power, Inc., a Division of AMETEK, Inc., is a global leader in the design and manufacture of precision, programmable power supplies for R&D, test and measurement, process control, power bus simulation and power conditioning applications across diverse industrial segments.
  • Page 6 This page intentionally left blank.
  • Page 7: Important Safety Instructions

    Neither AMETEK Programmable Power Inc., San Diego, California, USA, nor any of the subsidiary sales organizations can accept any responsibility for personnel, material or inconsequential injury, loss or damage that results from improper use of the equipment and accessories.
  • Page 8 AMETEK will, at its expense, deliver the repaired or replaced Product or parts to the Buyer. Any warranty of AMETEK will not apply if the Buyer is in default under the Purchase Order Agreement or where the Product or any part...
  • Page 9: Preface

    Briefly describes and explains the features, applications, and functions as well as brief operation principles of the EC1000S. 2. PREPARATIONS BEFORE USE Describes various cautions regarding preparations to be made before using the EC1000S, ranging from installation to connection of the power supply. 3. PANEL AND BASIC OPERATIONS Describes the functions and actions of the panel controls and their basic operations.
  • Page 10 User Manual EC1000S 9. SPECIFICATIONS Lists the specifications (functions and performance) of the EC1000S.
  • Page 11: Safety Precautions

    This product uses a line filter and must be grounded to avoid the risk of electric shock. To prevent electric shock, be sure to safely implement grounding according to local electrical safety codes. The EC1000S is automatically grounded when the 3-prong power plug is connected to a 3-prong power outlet with a protective grounding contact.
  • Page 12  Prevention of electrical shock due to output voltage The maximum output of the EC1000S is 400 V. Carefully operate the EC1000S to avoid electrical shock. Directly touching the output or changing cable connections while output is on may cause electrical shock.
  • Page 13 User Manual EC1000S <1> This product contains a lithium battery. <2> The LCD backlight unit contains mercury. <3> Use a professional industrial waste contractor to dispose of this product.
  • Page 14: Table Of Contents

    User Manual EC1000S Contents Page  Preface  ......................5  Safety Precautions  ..................7 1. OVERVIEW ......................21 General ..........................22 Features ..........................22 Applications ........................24 List of Functions ......................25 Operation Principles ......................27 2. PREPARATIONS BEFORE USE ............... 29 Checking Before Use .......................
  • Page 15 User Manual EC1000S Contents 4. ADVANCED OPERATION EXAMPLE ............... 79 Use as a DC Power Source ..................... 80 4.1.1 Connecting output terminal to load during DC output ........... 80 4.1.2 Selecting an output mode (AC+DC-INT mode) ..............81 4.1.3 Setting the output voltage range ..................83 4.1.4...
  • Page 16 User Manual EC1000S Contents 5.2.6 Shortcut keys ........................156 SET Menu ........................157 5.3.1 Output mode setting ......................159 5.3.2 Output voltage range setting ..................160 5.3.3 Setting output voltage ..................... 161 5.3.4 Setting output frequency ....................162 5.3.5 Output initial phase settings ................... 162 5.3.6...
  • Page 17 User Manual EC1000S Contents 6.3.2 Detailed command descriptions ..................203 Command Tree ....................... 234 Status System ......................... 236 6.5.1 Status byte ........................237 6.5.2 Standard event status ...................... 238 6.5.3 Operation status ......................240 6.5.4 Warning status ......................... 241 Error Message List ......................242 Programming Precautions ....................
  • Page 18 User Manual EC1000S Contents 9.17 External Dimensions and Weight ................. 286 California Instruments...
  • Page 19 Figure 2-1. Ambient Temperature and Humidity Ranges ..................33 Figure 2-2. Power Source Inlet ..........................36 Figure 2-3. Operation Check ............................. 39 Figure 3-1. EC1000S Operation Panel ........................44 Figure 3-2. EC1000S Front Panel ..........................45 Figure 3-3. EC1000S Rear Panel ..........................46 Figure 3-4.
  • Page 20 User Manual EC1000S Figures Figure 4-16. Example of Settings from Memory ....................133 Figure 5-1. LCD Screen (When Normal) ......................... 145 Figure 5-2. LCD Screen (When Warning Is Displayed) ..................145 Figure 5-3. Menu Tree .............................. 149 Figure 5-4. Modification Box ........................... 155 Figure 5-5.
  • Page 21 Table 0-1. Current Limiter Setting Ranges....................... 68 Table 0-2. Voltage and Frequency Setting Range Limiter ..................70 Table 4-1. Panel Settings When Using EC1000S as a DC Power Source .............. 80 Table 4-2. Setting Range Options for Various Output Voltage Ranges ..............83 Table 4-3.
  • Page 22 User Manual EC1000S Tables Table 5-8. Items in SYSTEM Menu .......................... 153 Table 5-9. Items in LIMIT Menu ..........................153 Table 5-10. Shortcut Keys ............................156 Table 5-11. SET Menu Items and Output Modes ....................157 Table 5-12. Output Modes ............................159 Table 5-13.
  • Page 23 User Manual EC1000S Tables Table 7-9. When Fault Is Suspected (Problem when Switching Power on/off) ........... 257 Table 7-10. When Fault Is Suspected (Problem During Key Operation).............. 257 Table 7-11. When Fault Is Suspected (Problem Related to Output Voltage or Output Voltage Range Setting) 258 Table 7-12.
  • Page 25: Overview

    User Manual EC1000S 1. OVERVIEW General ..........................22 Features ..........................22 Applications ........................24 List of Functions ......................25 Operation Principles ......................27 California Instruments...
  • Page 26: General

    User Manual EC1000S 1.1 General The EC1000S programmable AC/DC power source is a power source that can output AC and DC power, features a compact design for convenient desktop use, and provides a wealth of measurement functions. Rated output voltage is 140 Vrms (100 V range) or 280 Vrms (200 V range), with maximum output capacity of 1 kVA (during AC 200 V input).
  • Page 27 User Manual EC1000S  Frequency (only in the external synchronization mode)  Load power factor  Load crest factor  Harmonic current (up to 40th harmonic, 50/60 Hz fundamental only)  Enables AC-superimposed DC output When in the AC+DC mode, an AC wave (sine wave, square wave, or arbitrary waveform) can be superimposed on the DC output.
  • Page 28: Applications

    User Manual EC1000S 1.3 Applications  Research, development, and testing of various small-capacity built-in power source units  Research, development, and testing of various compact consumer electronics devices  Testing of battery-powered modules  As power source for testing characteristics of relays and switches ...
  • Page 29: List Of Functions

    User Manual EC1000S 1.4 List of Functions The main functions of the EC1000S are listed below. Table 1-1. List of Functions (1/2) Function Description Total of 8 modes (combinations of operation mode and signal source mode) Alternate current – internal signal source (AC-INT mode) Alternate current –...
  • Page 30: Table 1-2. List Of Functions (2/2)

    User Manual EC1000S Table 1-2. List of Functions (2/2) Function Description External control operation modes: Enable, disable Control input: Input level: High level: +4.0 V or higher Low level: +1.0 V or lower Non-destructive max. input: +10 V/5 V Input impedance: Pulled up to +5 V with 47 k...
  • Page 31: Operation Principles

    User Manual EC1000S 1.5 Operation Principles Figure 1-1 shows a block diagram of the EC1000S. PANEL <2> CONTROLLER SYSTEM CONTROL CONTROL DISPLAY EXT SIG IN 2.2V MAX CONTROL Zin=10k EXT SYNC IN <3> ISOLATION <1> PFC <4> AMP INVERTER INVERTER INVERTER ~LINE 100V-230V...
  • Page 32 User Manual EC1000S California Instruments...
  • Page 33: Preparations Before Use

    User Manual EC1000S 2. PREPARATIONS BEFORE USE Checking Before Use ....................... 30 Installation Environment ....................32 Grounding and Power Supply Connection..............35 Simple Operation Checks ....................38 Calibration ........................41 California Instruments...
  • Page 34: Checking Before Use

    EC1000S 2.1 Checking Before Use Before installing and using the EC1000S, make sure that it has not been damaged during shipment, and check that all the accessories and all parts of the main unit are included. If anything is missing, contact the California Instruments distributor from which the product was purchased.
  • Page 35 User Manual EC1000S This product contains high-voltage parts. Never remove the cover. All internal inspections of this product are to be performed only by service technicians qualified by California Instruments. California Instruments...
  • Page 36: Installation Environment

    If the EC1000S is exposed to sudden shifts in ambient temperature and/or humidity during transportation in winter, internal condensation may occur. In such a case, leave the EC1000S until the condensation has cleared before connecting the EC1000S to a power supply.
  • Page 37: Figure 2-1. Ambient Temperature And Humidity Ranges

     The EC1000S should be used indoors, and at an elevation of 2,000 meters.  The EC1000S uses a fan for forced-air cooling. To allow for ample air flow, be sure to maintain a gap of at least 50 cm between the air inlets and outlets along the sides and rear of this product and walls or other obstructions.
  • Page 38 User Manual EC1000S  Do not install the EC1000S in the following locations:  Location with flammable gas  An explosion may occur. Never install and use this product in such a location.  Outdoors, or location exposed to direct sunlight or near a fire or heat source ...
  • Page 39: Grounding And Power Supply Connection

    ! This product uses a line filter. Be sure to ground this product. Otherwise, an electric shock may occur. To prevent an electric shock from occurring, be sure to ground the EC1000S according to local electrical safety codes. When the input power line is 100 V AC, use power cord set 1. When a 3-prong power plug that includes a protective ground contact is connected to a 3-prong power supply outlet, this product is grounded automatically.
  • Page 40: Figure 2-2. Power Source Inlet

    User Manual EC1000S  Connect the power in this order. Make sure that the power source voltage to be used is within the specified range. Set the power switch to off. 3. Insert the power cord into the inlet on the rear panel.
  • Page 41 User Manual EC1000S The product itself has a withstand voltage of 1500 V AC. California Instruments...
  • Page 42: Simple Operation Checks

     Operation steps See “3.4.1 Power on/off”. 1. Set the EC1000S power source switch to on.  Operation starts once the power is on. <1> Press switch to up ( | ) position. <2> The LCD goes on and displays the initial screen.
  • Page 43: Figure 2-3. Operation Check

    User Manual EC1000S 2. Panel settings See “3.4 Main Operation Example”, for descriptors of Set the EC1000S as shown below. the settings methods. Table 2-1. Panel Settings for Operation Check Item Setting Output mode AC-INT mode Output voltage range 100 V...
  • Page 44 User Manual EC1000S Before turning off the power, make sure that the output on/off LED is off. California Instruments...
  • Page 45: Calibration

    User Manual EC1000S 2.5 Calibration To calibrate the EC1000S, contact NF Corporation or an NF Corporation representative. California Instruments...
  • Page 47: Panel And Basic Operations

    User Manual EC1000S 3. PANEL AND BASIC OPERATIONS Panel Components and Operations ................44 3.1.1 Operation panel ........................44 3.1.2 Front panel ......................... 45 3.1.3 Rear panel ........................... 46 Display and Initial Settings at Power-on ................ 47 I/O Terminals ........................49 3.3.1...
  • Page 48: Panel Components And Operations

    Numbers shown after names in this figure correspond to item numbers in the detailed descriptions below. CANCEL key ENTER key Output ON/OFF LED 3.4.8 MODIFY dial DIGIT key OUTPUT key 3.4.8 UP/DOWN keys Shortcut keys 5.2.6 Figure 3-1. EC1000S Operation Panel California Instruments...
  • Page 49: Front Panel

    User Manual EC1000S 3.1.2 Front panel Numbers shown after names in the following two figures correspond to item numbers in the detailed descriptions below. Power switch 3.4.1 Exhaust vent Output outlets (universal type) 3.3.1 Figure 3-2. EC1000S Front Panel California Instruments...
  • Page 50: Rear Panel

    User Manual EC1000S 3.1.3 Rear panel External signal input/external sync signal input connector 3.3.4 Exhaust vent Output terminals 3.3.1 LINE INPUT USB connector 3.3.2 External control I/O connector 3.3.3 Figure 3-3. EC1000S Rear Panel California Instruments...
  • Page 51: Display And Initial Settings At Power-On

    See “7.2.1 Error at power-on”, for description of error messages and recommended responses.  Initial settings Table 3-1 and Table 3-2 list the initial settings for the EC1000S. Table 3-1. Settings in Memory (1/2) Function Setting Initial Setting...
  • Page 52: Table 3-2. Settings In Memory (2/2)

    User Manual EC1000S AC voltage frequency 50.0 Hz Phase when output goes on 0.0 Sine wave AC voltage waveform (SIN) Table 3-2. Settings in Memory (2/2) Function Setting Initial setting At Power-on Output peak current 100 V range +42.0 A...
  • Page 53: I/O Terminals

    User Manual EC1000S 3.3 I/O Terminals 3.3.1 Output terminals (front and rear)  Connection to output outlet (front) The front panel is equipped with universal type output outlets that accommodate power plug types from various countries. These outputs are isolated from the power inputs and case (ground connection).
  • Page 54: Figure 3-5. Output Terminals (Rear)

    User Manual EC1000S  Connection to output terminals (rear) Outputs are isolated from the input power lines and the case (ground connection). When connecting to an output terminal, attach a round crimp contact with a sleeve that has sufficient capacity for the output current.
  • Page 55: Figure 3-6. Connection To Output Terminals

    To ensure safety, always turn the power off before connecting to output cables. ! CAUTION When using the EC1000S as a DC power supply, it is necessary to connect a diode for protection, depending on the load, such as a capacitor and inductor. refer to “4.1.1 Connecting output terminal to load For how to connect a diode for protection, during DC output”.
  • Page 56: Usb Connector

    User Manual EC1000S 3.3.2 USB connector Use a commercial (off-the-shelf) USB cable to connect the rear USB connector to a PC. Attach the accompanying ferrite core to the USB cable to reduce the leakage electric wave. See “6. USB INTERFACE”, for description of using a USB connection for remote control from a USB connector Figure 3-7.
  • Page 57: External Control I/O

    EC1000S 3.3.3 External control I/O The EC1000S is equipped with an (8-bit) external control I/O connector to enable external control using digital signal input and status output using digital signals. The connector is a D-sub 25-pin type. External control I/O connector Figure 3-8.
  • Page 58: Table 3-3. External Control I/O Connectors

    User Manual EC1000S Table 3-3. External Control I/O Connectors Pin No. Function Remark 0: Off Output Power source on/off status 1: On 0: Off Output Output on/off status 1: On 0: Off Note Limiter operation Output 1: On 0: Normal...
  • Page 59: External Signal Input/External Sync Signal Input Terminal

    User Manual EC1000S 3.3.4 External signal input/external sync signal input terminal When using an external input signal as the signal source with external synchronization, connect to the external signal input/external sync signal input terminals (EXT SIG IN/EXT SYNC IN) on the rear panel.
  • Page 60: Figure 3-9. Ext Sig In/Ext Sync In Terminal

    User Manual EC1000S External signal input/ external sync signal input terminal Figure 3-9. EXT SIG IN/EXT SYNC IN Terminal California Instruments...
  • Page 61: Main Operation Example

    User Manual EC1000S 3.4 Main Operation Example Main operations are described below in the following order.  Power-on 3.4.1 Power on/off   Set output mode 3.4.2 Setting output mode   Set output voltage range 3.4.3 Setting output voltage range ...
  • Page 62: Power On/Off

    User Manual EC1000S 3.4.1 Power on/off When the POWER switch is set to “|”, the power goes on, and when it is set to “”, the power goes off. POWER switch Figure 3-10. POWER Switch a) Power-on Set the POWER switch to “|”.
  • Page 63: Figure 3-11. Example Of Main Operations Screen (In Ac-Int Mode)

    User Manual EC1000S  Main operations screen is displayed. Output mode 3.4.2 Output voltage range 3.4.3 Output voltage 3.4.5 Output frequency 3.4.6 Waveform 3.4.4 Limiter function 3.4.7 Select measurement display 3.4.9 Figure 3-11. Example of Main Operations Screen (in AC-INT Mode) The main operations screen that appears immediately after the power-on shows the same settings that were displayed the last time the power was turned off.
  • Page 64: Setting Output Mode

     Before turning off the power, make sure the OUTPUT setting is off. 3.4.2 Setting output mode The EC1000S provides the eight output modes listed in Table 3-4. The output mode cannot be changed while output is on. Before changing the output mode, set output to off.
  • Page 65 User Manual EC1000S  Operation steps Select output mode by selecting in the SET menu. 1. Press the key to move the cursor to the icon. Cursor CANCEL ENTER E NT E R 2. Press the key to display the selection box.
  • Page 66: Setting Output Voltage Range

    User Manual EC1000S CANCEL ENTER 3.4.3 Setting output voltage range Select 100 V or 200 V. The output voltage range cannot be changed while output is on. The setting range varies according to the output voltage range. Table 3-5. List of Setting Ranges for Various Output Voltage Range...
  • Page 67 User Manual EC1000S E NT E R 2. Press the key to display the selection box. Selection box CANCEL ENTER 3. Press the key to move the cursor to the output voltage range to be selected. CANCEL ENTER E NT E R 4.
  • Page 68: Setting Waveform

    EC1000S 3.4.4 Setting waveform The EC1000S allows to select a sine wave, square wave, or arbitrary waveform (16 types in all) as the waveform to be output, by using the control panel. Arbitrary waveforms are generated using data transfers via the USB interface.
  • Page 69: Setting Output Voltage

    User Manual EC1000S 3. Press the key to move the cursor to the waveform to be output. CANCEL ENTER E NT E R 4. Press the key to set the selected waveform. CANCEL ENTER 3.4.5 Setting output voltage For AC-INT mode, the output voltage settings listed below can be selected.
  • Page 70: Setting Output Frequency

    User Manual EC1000S  Operation steps To set the output voltage, set in the SET menu. 1. Press the key to move the cursor to the icon. Cursor CANCEL ENTER E NT E R 2. Press the key to display the modification box.
  • Page 71 User Manual EC1000S Setting Range Resolution Setting Accuracy 1.0 Hz to 550.0 Hz 0.1 Hz ±100 ppm (1.0 Hz to 550.0 Hz, 23 ±5°C) See “3.4.7 Using limiter functions”, when using an output frequency setting range limit.  Operation steps To set the output frequency, set a numerical value labeled in the SET menu.
  • Page 72: Using Limiter Functions

    EC1000S 3.4.7 Using limiter functions a) Using current limiter The EC1000S includes an output peak current limiter function and an output average current limiter function. Limiter settings are retained separately for 100 V and 200 V ranges. Setting range limits can be set for output voltage and output frequency.
  • Page 73 User Manual EC1000S may not be quick enough, so the output overcurrent protection function may be triggered.  Operation steps Select the icon in the SET menu corresponding to the desired setting, then set a numerical value.  Output peak current limiter (positive): ...
  • Page 74: Table 0-2. Voltage And Frequency Setting Range Limiter

    User Manual EC1000S b) Using setting range limiter The EC1000S internal mode (AC-INT, AC+DC-INT) and internal + external mode (AC-ADD, AC+DC-ADD) include setting range limiter functions for output voltage and output frequency. See “5.5.5 Setting range limiter (LIMIT)”, for details.
  • Page 75 User Manual EC1000S CANCEL ENTER E NT E R 3. Press the key to display the LIMIT screen . CANCEL ENTER 4. Press the key to move the cursor to the item to be set. Cursor CANCEL ENTER E NT E R 5.
  • Page 76: Output On/Off

    3.4.8 Output on/off OUTPUT To turn the EC1000S output on or off, simply press the key. When output is on, the output on/off LED goes on, and it goes off when the output is off.
  • Page 77: Using Measurement Functions

    User Manual EC1000S 3.4.9 Using measurement functions The EC1000S is equipped with the following measurement functions.  Voltage (RMS, average DC, peak)  Current (RMS, average DC, peak, peak hold)  Power (effective, reactive, apparent)  Synchronization frequency (external synchronization) ...
  • Page 78 User Manual EC1000S  Operation steps To select a measurement display, select “MEASURE” in the MISC menu 1. Press the key to move the cursor to the icon. CANCEL ENTER Cursor E NT E R 2. Press the key to display the MISC menu.
  • Page 79 User Manual EC1000S 4. Press the key to move the cursor to the desired measurement display. CANCEL ENTER HC1 to HC4 are for measuring harmonic current. See “4.3 Measurement of Harmonic Current”. E NT E R 5. Press the key to set the selected measurement display.
  • Page 80: Output Characteristics

    User Manual EC1000S 3.5 Output Characteristics The following charts show output voltage vs. output current of the EC1000S (resistance load, typical character). Figures 3-13 and 3-14 show charts when AC is input, and Figures 3-15 and 3-16 show when DC is input.
  • Page 81: Figure 0-3. Output Voltage Vs. Output Current Characteristics (200 V Ac Input, Ac-Int)

    User Manual EC1000S 100.0 74.1 Maximum output power = 1000 VA Output range: Rated voltage/Maximum current --------------------------------------------------------------------- 100 V range: 100 Vrms/10 Arms 200 V range: 200 Vrms/5 Arms 100.0 135 140 Output voltage (Vrms)/Rated output voltage (Vrms) × 100 (%) Figure 0-3.
  • Page 82: Figure 0-5. Output Voltage Vs. Output Current Characteristics (200 V Ac Input, Ac+Dc-Int, Ac=0)

    User Manual EC1000S 100.0 52.7 Maximum output power = 1000 W Output range: Rated voltage/Maximum current --------------------------------------------------------------------- 100 V range: 100 V/10 A 200 V range: 200 V/5 A 100.0 190 200 Output voltage (V)/Rated output voltage (V) × 100 (%) Figure 0-5.
  • Page 83: Advanced Operation Example

    User Manual EC1000S 4. ADVANCED OPERATION EXAMPLE Use as a DC Power Source ..................... 80 4.1.1 Connecting output terminal to load during DC output ........... 80 4.1.2 Selecting an output mode (AC+DC-INT mode) ..............81 4.1.3 Setting the output voltage range ..................83 4.1.4...
  • Page 84: Use As A Dc Power Source

    EC1000S 4.1 Use as a DC Power Source When using the EC1000S as a DC voltage source, select DC - Internal signal mode (AC+DC-INT), then enter the settings listed in Table 4-1. AC components (sine wave, square wave, and arbitrary waveform) can be superimposed on the DC power source.
  • Page 85: Selecting An Output Mode (Ac+Dc-Int Mode)

    EC1000S A block diode is required to prevent current from flowing from the load to the output terminal of the EC1000S even if the output voltage is lowered. Connect the diode and the load serially as shown in Figure 4-1.
  • Page 86 User Manual EC1000S Cursor CANCEL ENTER E NT E R Press the key to display the selection box. Selection box CANCEL ENTER key to move the cursor to “AC+DC-INT”. Press the Cursor CANCEL ENTER E NT E R Press the key to set the AC+DC-INT mode.
  • Page 87: Setting The Output Voltage Range

    User Manual EC1000S 4.1.3 Setting the output voltage range Select 100 V or 200 V as the output voltage range. As is shown in Table 4-2, the range that can be set for the output voltage differs according to the selected output voltage range.
  • Page 88 User Manual EC1000S Selection box CANCEL ENTER Press the key to move the cursor to 100 V or 200 V. CANCEL ENTER E NT E R Press the key. Then the icon appears in the status window . CANCEL ENTER ...
  • Page 89: Setting The Output Voltage

    User Manual EC1000S 4.1.4 Setting the output voltage The ranges that can be set are listed below. Table 4-3. Output Voltage Setting Ranges When AC+DC Mode Is Selected Setting resolution Setting range 200.0 to +200.0 V 0.1 V 100 V range DC voltage 400.0 to +400.0 V...
  • Page 90 User Manual EC1000S Modification box CANCEL ENTER Press the keys to move the cursor to the digit to be changed, turn the MODIFY dial and set a numerical value. The set value will be reflected in output immediately. Decrease Increase...
  • Page 91: Setting Output Frequency Of Superimposed Ac

    User Manual EC1000S 4.1.5 Setting output frequency of superimposed AC To set the output frequency, set a numerical value under in the SET menu.  Operation steps Press the key to move the cursor to the icon. Cursor CANCEL ENTER...
  • Page 92: Setting Superimposed Ac Waveform

    User Manual EC1000S 4.1.6 Setting superimposed AC waveform Selects among 16 types of sine waves (SIN), square waves (SQU), and arbitrary waveforms (ARB). Arbitrary waveforms are generated using data transfers via the USB interface. See “4.6 Output of Arbitrary Waveforms”, for details of arbitrary waveforms.
  • Page 93: Use Measurement Function

    User Manual EC1000S E NT E R 4. Press the key to set the selected waveform. CANCEL ENTER 4.1.7 Use measurement function Measured values are displayed in the measured value window MEASURE screen, which enables the output status to be monitored.
  • Page 94 User Manual EC1000S b) Select measurement display Select one form among RMS, AVG, or PEAK as the measured current and voltage values to be displayed. To select a measurement display, select “MEASURE” in the MISC menu  Operation steps Press the key to move the cursor to the icon.
  • Page 95 User Manual EC1000S Press the key to move the cursor to the desired measurement display. Select “AVG” to measure only the DC power source. CANCEL ENTER E NT E R Press the key to set the selected measurement display. CANCEL...
  • Page 96: Measurement Of Inrush Current

    Since the rated current is 3.8 A, this represents a fourfold increase in the current flow. The EC1000S can supply four times as large peaked current as the rated current supporting such a usage. There is no need to use a power source with an extra large capacity in order to supply the inrush current (for a capacitor input circuit of CF = 4 or less).
  • Page 97: Set Phase At Output On

    User Manual EC1000S 4.2.2 Set phase at output on “Phase at output on” in the SET menu and set a numerical To set the phase at output on, select value. Before changing the phase at output on, set output off.
  • Page 98: Set Measurement Display To Peak Value

    User Manual EC1000S 4.2.3 Set measurement display to peak value Set the displayed current measurement value to the peak value (PEAK). To change the displayed current measurement value, select “MEASURE” in the MISC menu  Operation steps E NT E R...
  • Page 99: Reset The Peak Current Hold Value

    Peak hold value 4.2.4 Reset the peak current hold value The EC1000S includes a function that resets the peak current hold value. A reset sets the peak current hold value to 0 Apk, and the maximum peak value of current is updated until the next reset is performed.
  • Page 100 User Manual EC1000S  The MISC menu is displayed. E NT E R key to move the cursor to “RESET Ipkh…”, then press the Press the key.  The EXEC box is displayed. EXEC box CANCEL ENTER E NT E R When “EXEC”...
  • Page 101: Measurement Of Harmonic Current

    The EC1000S includes a function for measuring harmonic current. This measurement does not comply with the IEC standards. This measurement can be performed only in AC-INT mode and when the fundamental wave (panel frequency setting) is either 50 Hz or 60 Hz.
  • Page 102 User Manual EC1000S CANCEL ENTER Selection box Press the key to move the cursor to the harmonic order to be displayed, E NT E R then press the key. CANCEL ENTER  When HC1 to HC4 have been selected as the displayed measurements, the following harmonic current measured values are displayed.
  • Page 103: Figure 4-4. Example Of Harmonic Measurement Window Display

    User Manual EC1000S Figure 4-4. Example of Harmonic Measurement Window Display Press the key to switch the HC1, HC2, HC3, and HC4 screens. Press the key to end harmonic measurement and return to the TOP MENU (normal display). California Instruments...
  • Page 104: Using The Sequence Function

    4.4 Using the Sequence Function 4.4.1 Sequence operations When the sequence function of the EC1000S is used, the pattern to be output can be freely programmed. For example, if a series of voltage change patterns that will be used in voltage variation tests for car electronics are programmed in advance, the output will change according to the sequence, so the test can be completed in one series of operations.
  • Page 105: Table 4-4. Step Execution Parameters

    User Manual EC1000S a) Sequence A sequence is comprised of a combination of at least two steps (the minimum unit is one step). b) Step Step execution parameters and step transition parameters are specified for each step. When sequence execution is not performed (i.e., during normal output mode), these parameters are specified in the same way.
  • Page 106: Table 4-5. Step Transition Parameters

    User Manual EC1000S The step transition parameters specify the next step and its timing. Table 4-5. Step Transition Parameters [Step time] Specified in time [Step end phase] Sets step end phase as enabled or disabled. The following parameters are provided to specify the action at the end of the step.
  • Page 107: Figure 4-5. Example Of Basic Step Transitions

    User Manual EC1000S An example of basic step transitions is shown below. Output parameters Steps 1, 2 Step 3 Normal output Step 1 Step 2 Step 3 Time step <START> Figure 4-5. Example of Basic Step Transitions c) Sequence mode Figure 4-6 shows a mode transition diagram for sequence operations.
  • Page 108: Figure 4-7. Example Of Sequence Operation From Hold Mode To

    User Manual EC1000S Table 4-6. Sequence Modes Idle mode Mode during normal output. Run mode Mode during sequence execution. Sequence pause mode. Hold mode Maintains output at time when <Hold> was set.  d) Sequence control The following describes how sequence operations are controlled. Sequence control operations are indicated in triangular brackets (<...
  • Page 109: Figure 4-8. Example Of Sequence Operation At

    User Manual EC1000S Step 2 settings • Final value • Step time • Branch step 5 Output parameter <Branch> Step 5 settings • Final value from previous step is used Step 1 Step 2 Step 5 Time Figure 4-8. Example of Sequence Operation at <Branch>...
  • Page 110: Figure 4-9. Loop Specification Method And Step Transition Example

    User Manual EC1000S Step 4  [jump step]  [jump times] NoCare : Specified times (10 times) Step 5 Changes to [jump step]  [jump step]  [jump times] NoCare : Specified times (10 times) Changes to next step Step 6 number after jump ...
  • Page 111: Figure 4-10. Example Of Step Synchronized Output

    User Manual EC1000S f) Step synchronized output The code that is specified in the step being executed is output to the external control I/O connector. However, “0” is output during a normal output step. Step synchronized output Bit 0 Step 1...
  • Page 112: Sequence Settings

    Sequence settings are made as described below.  Input per step via control panel  Programmed from PC, with USB transfer to EC1000S  Sequence data is edited and transferred using accompanying control software See “4.4.3 Programming sequences” below, for further description of the program messages for sequence operations.
  • Page 113 User Manual EC1000S E NT E R Press the key to display the SEQUENCE screen. CANCEL ENTER Press the key to select the operation to be performed.  To control sequence operations: select “CONTROL…”. See “4.4.4 Control of sequence operations”.) Note This cannot be selected when output is off.
  • Page 114: Programming Sequences

    User Manual EC1000S 4.4.3 Programming sequences When “PROGRAM” is selected in the SEQUENCE screen, the PROGRAM screen is displayed. The sequence is programmed according to the various settings in the PROGRAM screen. To change to the SEQUENCE screen, the output mode must first be switched to AC-INT or AC+DC-INT.
  • Page 115: Table 4-9. Setting Items In Program Screen (2/3)

    User Manual EC1000S Table 4-9. Setting Items in PROGRAM Screen (2/3) Setting Name Item Operation Sets the AC frequency. AC frequency Setting range: 1.0 Hz to 550.0 Hz Setting resolution: 0.1 Hz Sets the method used to change to the specified frequency.
  • Page 116: Table 4-10. Setting Items In Program Screen (3/3)

    User Manual EC1000S Table 4-10. Setting Items in PROGRAM Screen (3/3) Setting Name Item Operation Sets operation after end of step. Select one of the following three options.  CONT: STEP TERM Step end Sequence continuation  STOP: Change to Idle mode ...
  • Page 117 User Manual EC1000S Open the SEQUENCE screen and display the PROGRAM screen. See “4.4.2 Sequence settings”, for steps for displaying the PROGRAM screen. key to move the cursor to “TIME”. Press the Cursor CANCEL ENTER E NT E R Press the key to display the modification box.
  • Page 118 User Manual EC1000S  Operation step <2> The settings for DC voltage operation type (AC+DC mode only), AC voltage operation type, frequency operation type, waveform (WAVE), waveform operation type, phase operation type, step synchronized output (SYNC CODE), step synchronized output operation type, step end phase enable/disable (STOP PHS), and step end (STEP TERM) are made by selecting settings for the various items in the order described below.
  • Page 119: Control Of Sequence Operations

    User Manual EC1000S E NT E R Press the key to set the operation at the step end. 4.4.4 Control of sequence operations The CONTROL item in the SEQUENCE screen is used to control sequence operations. A sequence can be started only when output is on. Consequently, the sequence control selection box can be selected only when output is on.
  • Page 120: Figure 4-12. Sequence Shortcut Keys

    User Manual EC1000S b) When operating with the shortcut keys  Operation step The control of sequence operations, can also be performed with the shortcut keys at the bottom of the control panel. The START, STOP, and HOLD keys under “SEQUENCE” correspond to items in the CONTROL selection box.
  • Page 121: Clear Sequence Memory

    User Manual EC1000S 4.4.5 Clear sequence memory Data is saved to sequence memory automatically. Separate sequence data can be saved for 100 V range and 200 V range in AC-INT mode and for 100 V range and 200 V range in AC+DC-INT mode.
  • Page 122: Example Of Sequence Operation Settings

    User Manual EC1000S 4.4.6 Example of sequence operation settings The following describes the steps for setting a new sequence operation, based on the sequence See “5.6.5 Time unit transition example described in Table 4-12. The time unit is seconds (s). ( setting”...
  • Page 123 User Manual EC1000S The operation steps are described below.  Operation steps using control panel Open the PROGRAM screen. “4.4.2 Sequence The steps for opening the PROGRAM screen are described in settings”.  Setting step 1 key to move the cursor to “DC VOLT”.
  • Page 124 User Manual EC1000S Decrease Increase E NT E R CANCEL Press the key to close the modification box. key to move the cursor to the line displaying “CONST” (fourth line from the top). Press the CANCEL ENTER E NT E R Press the key to display the selection box.
  • Page 125 User Manual EC1000S E NT E R Press the key to confirm the setting. The settings for all other items are the same as the default settings, so they are not changed.  Setting step 2 key to move the cursor to “STEP”.
  • Page 126: Execution Of Sequence Operations

    User Manual EC1000S 4.4.7 Execution of sequence operations Once sequence programming has been completed, start the sequence. The execution steps for the sequence operations are described below. a) When using the MISC menu  Operation steps Press the TOP MENU key (shortcut key) to return to the top menu. In the top menu, “SET” is shown in the upper left corner of the menu window.
  • Page 127 User Manual EC1000S CANCEL ENTER E NT E R 4. Press the key to display the selection box. CANCEL ENTER key to move the cursor to “ON”. Press the CANCEL ENTER E NT E R Press the key to set output on, at which time the output on/off LED goes on.
  • Page 128 User Manual EC1000S CANCEL ENTER E NT E R key to select “CONTROL”, then press the Press the key to display the selection box. Selection box CANCEL ENTER E NT E R Select “START”, then press the key to start the sequence.
  • Page 129: Figure 4-14. Lcd Screen During Sequence Execution

    User Manual EC1000S c) Operations after starting sequence When a sequence is started, the status icon ( ) is displayed in the status window to indicate that the sequence is being executed. A message box is also displayed to indicate the step number being executed.
  • Page 130: Control Using External Control I/O Connector

    The sequence step synchronized output signals are output from the external control I/O connector. The status of the EC1000S can be monitored by reading the output signals from the external control input/output connector. See “3.3.4 External signal input/external sync signal input terminal”.
  • Page 131: Output Of Arbitrary Waveforms

    Arbitrary waveform memory The EC1000S includes arbitrary waveform memory as an internal signal source, and this memory can be written to via the USB interface. Arbitrary waveform memory cannot be written to from the panel screen.
  • Page 132 User Manual EC1000S CANCEL ENTER Selection box Press the key to move the cursor to the arbitrary waveform to be E NT E R selected (“ARB1” to “ARB16”), then press the key.  This sets the selected arbitrary waveform. CANCEL...
  • Page 133: Synchronization Of Output To External Signal

    EC1000S 4.7 Synchronization of Output to External Signal The EC1000S is equipped with an externally synchronized oscillation function. The software PLL function synchronizes the output frequency to the frequency of the external synchronization TTL signal. The output frequency can be synchronized to any frequency from 40 to 500 Hz. The synchronization phase difference cannot be set.
  • Page 134: External Signal Synchronization

    User Manual EC1000S 4.7.1 External signal synchronization This function synchronizes the output frequency to a signal that is input to the rear panel external sync signal input (BNC connector). The output frequency can be synchronized to any frequency from 40 to 500 Hz.
  • Page 135 User Manual EC1000S E NT E R Press the key to select the mode to be set. CANCEL ENTER E NT E R Press the key to move the cursor to the icon, then press the key to display the selection box.
  • Page 136: Line Synchronization

    User Manual EC1000S 4.7.2 Line synchronization This function synchronizes the output frequency to the frequency of the AC power source line. The output frequency can be synchronized to power source line frequency (50 Hz/60 Hz).  Operation steps Select in the SET menu to select external synchronization mode, then set the sync signal source to line synchronization (“LINE”).
  • Page 137: Using Memory Functions

    User Manual EC1000S 4.8 Using Memory Functions Various settings are retained in memory with battery backup, and the settings when the power source was last turned off are restored at power-on. Among these various settings, the basic settings (output mode, output range, DC settings, AC settings, output current limiter, and setting range limit) can be stored in store/recall memory No.
  • Page 138 User Manual EC1000S  Operation steps and then “MEMORY” in the MISC menu to store, recall, or clear memory or to set the Select target memory number. Press the key to move the cursor to the icon. CANCEL ENTER Cursor E NT E R 2.
  • Page 139 User Manual EC1000S Press the key to select the operation to be performed. CANCEL ENTER E NT E R Press the key to display the selection box. CANCEL ENTER Press the key to move the cursor to the target memory number.
  • Page 140: Amplification Of External Signal

    EC1000S 4.9 Amplification of External Signal The EC1000S can be used as an amplifier to which a signal is input from the external signal input terminals on the rear panel, by selecting the external signal source mode (AC-EXT or AC+DC-EXT).
  • Page 141 User Manual EC1000S a) Possible voltage settings for output  Operation steps Select the icon in the SET menu corresponding to the desired setting, then set a numerical value. “Select output mode”:  Select external mode (AC-EXT or AC+DC-EXT). “Output voltage range”: ...
  • Page 142 User Manual EC1000S 4. Press the key to move the cursor to the output voltage range to be set, then press E NT E R key.  This sets the selected output voltage range. CANCEL ENTER E NT E R...
  • Page 143 User Manual EC1000S b) Signal input The setup for external input is completed once the above operations and the settings are done. Now connect the signal source and input the signal. If the input signal exceeds the maximum voltage that can be output, the output overvoltage protection function will operate, in which case output may be turned off and the waveform may become distorted.
  • Page 144: Adding External Signals And Internal Signals

    EC1000S 4.10 Adding External Signals and Internal Signals When in the EC1000S internal + external mode (AC-ADD, AC+DC-ADD), the internal signal source signal can be added to the external signal input prior to output. When using a signal input from an external source as a signal source, connect to the rear panel external signal input/external sync signal input terminal (EXT SIG IN/EXT SYNC IN).
  • Page 145 User Manual EC1000S CANCEL ENTER California Instruments...
  • Page 147: Menus

    User Manual EC1000S 5. MENUS Screen Configuration ..................... 145 5.1.1 Status icons ........................147 5.1.2 Warnings and error messages ..................148 5.1.3 Sequence display ......................148 Basic Operations ......................149 5.2.1 Menus ..........................149 5.2.2 Navigating the menu tree ....................154 5.2.3...
  • Page 148 User Manual EC1000S 5.5.7 Reset of peak current hold value ..................177 5.5.8 Output on/off ........................177 System Menu ........................179 5.6.1 Keylock ..........................181 5.6.2 Beep sound ........................182 5.6.3 LCD contrast adjustment ....................183 5.6.4 LCD display color setting ....................184 5.6.5...
  • Page 149: Screen Configuration

    User Manual EC1000S 5.1 Screen Configuration The LCD on the control panel splits in multiple windows, as is shown in Figure 5-1 and Figure 5-2. Figure 5-1. LCD Screen (When Normal) Figure 5-2. LCD Screen (When Warning Is Displayed) California Instruments...
  • Page 150 User Manual EC1000S The following information is displayed in various parts of the screen. A: Status window: This window shows the status of various devices. B: Message box: When an error has occurred, a message box appears near the bottom of the status window, and the cause is displayed.
  • Page 151: Status Icons

    User Manual EC1000S 5.1.1 Status icons The status window displays status icons that indicate the status of the EC1000S. These status icons are listed below in Table 5-1. Table 5-1. Status Icons List Icon Item Description When a backup memory checksum error occurs, this icon Battery almost depleted flashes to indicate that the battery is almost depleted.
  • Page 152: Warnings And Error Messages

    See “7 TROUBLESHOOTING”, for further description of error messages. 5.1.3 Sequence display The EC1000S is equipped with a function that indicates the progress of the sequence being executed. The information listed in Table 5-2 is displayed in the status window.
  • Page 153: Basic Operations

    The EC1000S basic operations are described below. 5.2.1 Menus The EC1000S uses a hierarchical menu system for setting various parameters. Menu operations are performed using the menu window in the LCD screen. Figure 5-3 illustrates the EC1000S menu tree. (1) SET...
  • Page 154: Table 5-3. Items In Set Menu

    User Manual EC1000S Table 5-3. Items in SET Menu Icon Item Reference See “5.3.1 Output mode setting”. Select output mode See “5.3.2 Output voltage range Select output voltage range setting”. See “5.3.3 Setting output voltage ”. Set AC voltage to be output See “5.3.3 Setting output voltage ”.
  • Page 155: Table 5-4. Items In Misc Menu

    User Manual EC1000S Table 5-4. Items in MISC Menu Setting Name Setting Reference See “5.5.1 Sequence (SEQUENCE)”. SEQUENCE Open sequence menu See “5.5.2 Memory (MEMORY)”. MEMORY Open memory menu See “5.5.3 Remote (REMOTE)”. REMOTE Open remote menu See “5.5.4 System (SYSTEM)” and...
  • Page 156: Table 5-5. Items In Sequence Menu

    User Manual EC1000S Table 5-5. Items in SEQUENCE Menu Setting Name Setting Reference CONTROL... See “4.4.4 Control of sequence Control sequence operation operations”. PROGRAM STEP Step number TIME Step time DC voltage DC VOLT DC operation type AC voltage AC VOLT...
  • Page 157: Table 5-7. Items In Remote Menu

    User Manual EC1000S Table 5-7. Items in REMOTE Menu Setting Name Setting Reference See “5.5.3 Remote (REMOTE)”. LOCAL... Switch to local mode Table 5-8. Items in SYSTEM Menu Setting Name Setting Reference See “5.6.1 Keylock”. KEYLOCK Keylock See “5.6.2 Beep sound”.
  • Page 158: Navigating The Menu Tree

    User Manual EC1000S 5.2.2 Navigating the menu tree Use the UP and DOWN keys to move among items displayed in the menu windows. Press the ENTER key to go to a lower-level menu in the menu tree and press the CANCEL key to go to the higher-level menu.
  • Page 159: Numerical Value Input Operations

    User Manual EC1000S 5.2.3 Numerical value input operations When input of numerical values is required, the modification box is displayed. When the modification box is opened, the digit cursor is displayed in one column. Use the DIGIT key to move the digit cursor and use the MODIFY key to increase or decrease values.
  • Page 160: Exec Input

    When the ENTER key is pressed, the item is actually executed, and the EXEC box is closed. If the EXEC box is closed by pressing the CANCEL key, nothing is executed. Figure 5-6. EXEC Box 5.2.6 Shortcut keys The EC1000S shortcut keys can be used to perform the following operations. Table 5-10. Shortcut Keys Key No. Indication on Control Panel...
  • Page 161: Set Menu

    User Manual EC1000S 5.3 SET Menu The SET menu in the menu window is used to set basic functions. The available settings differ according to the output mode. Table 5-11. SET Menu Items and Output Modes Output Mode AC+DC Item...
  • Page 162 User Manual EC1000S Use of other functions and detailed settings         See “ Table 5-4. Items in MISC ”. Menu California Instruments...
  • Page 163: Output Mode Setting

    The DC component is removed.  AC+DC: In this mode, the EC1000S is used as a DC constant voltage source. The voltage supplied to the load is proportionate to the set value or the input signal.
  • Page 164: Output Voltage Range Setting

    User Manual EC1000S Figure 5-7 shows a block diagram of the EC1000S signal sources. SYSTEM CONTROL EXT INPUT GAIN EXT SIG IN EXT SYNC To AMP CONTROL LINE LINE SYNC SYNC INT OSC Figure 5-7. Block Diagram of Signal Sources ...
  • Page 165: Setting Output Voltage

    User Manual EC1000S 5.3.3 Setting output voltage The output voltage settings listed below can be made for the two operation modes. Table 5-13. Output Voltage Settings Operation Mode Setting Setting Range Resolution 0.1 Vrms 0.0 to 140.0 Vrms AC mode 100 V range 0.1 Vp-p...
  • Page 166: Setting Output Frequency

    User Manual EC1000S 5.3.4 Setting output frequency This sets the AC voltage frequency. When either AC-SYNC or AC+DC-SYNC mode is selected, the AC voltage frequency cannot be set. Table 5-14. Output Frequency Setting Setting Range Resolution Setting Accuracy ±100 ppm(1.0 Hz to 550.0 Hz,23 ±5C)...
  • Page 167: Output Peak Current Limiter Settings

    User Manual EC1000S See “3.4.4 Setting waveform”, for description of waveform settings. See the control software user’s manual, for description of arbitrary waveform transfers. 5.3.7 Output peak current limiter settings The positive value and the negative value of the peak current limiter can be set independently.
  • Page 168: Output Average Current Limiter Settings

    User Manual EC1000S CAUTION ! If the output current or frequency suddenly increases, due, for example, to an output short circuit, the limiter response may not be quick enough, so the output overcurrent protection function may be triggered. 5.3.8 Output average current limiter settings This limits the output average current.
  • Page 169: Measure Screen

    User Manual EC1000S 5.4 MEASURE Screen The MEASURE screen in the measured value window displays measured values for items selected as measurement display items (some of which depend on the output mode). The measured values for output voltage and output current are displayed as values set via “MEASURE” in the MISC menu.
  • Page 170 User Manual EC1000S “5.4.4 Load power factor Always Load power factor None displayed measurement” “5.4.5 Load crest factor Always Load crest factor None displayed measurement” California Instruments...
  • Page 171: Output Voltage Measurement

    User Manual EC1000S 5.4.1 Output voltage measurement When “MEASURE” has been selected in the MISC menu, RMS value (RMS), average DC (AVG), or peak value (PEAK) is displayed. The initial setting (factory setting) is to display the RMS value (RMS).
  • Page 172: Output Current Measurement

    User Manual EC1000S 5.4.2 Output current measurement When “MEASURE” has been selected in the MISC menu, RMS value (RMS), average DC (AVG), or peak value (PEAK) is displayed. The initial setting (factory setting) is to display the RMS value (RMS).
  • Page 173: Output Power Measurement

    User Manual EC1000S 5.4.3 Output power measurement The output power values described below are always displayed. a) Effective power  Full scale: 1200 W  Display resolution:  Measurement accuracy: 2% of the full scale (at 23 5°C) (when output is 100 VA or above) b) Apparent power Calculated as output voltage RMS value ×...
  • Page 174: Output Harmonic Current Measurement

    User Manual EC1000S 5.4.6 Output harmonic current measurement Output harmonic current measurement is enabled only in the AC-INT mode and when the frequency setting is 50 or 60 Hz. When HC1, HC2, HC3, or HC4 is selected under “MEASURE” in the MISC menu, the harmonic measurement window is displayed.
  • Page 175: Misc Menu

    User Manual EC1000S 5.5 MISC Menu This menu is used to enter detailed settings for various functions. These settings can be entered regardless of whether output is on or off. 5.5.1 Sequence (SEQUENCE) A sequence menu is displayed. This menu is used to control sequences, set up programming, and clear sequence memory.
  • Page 176: Table 5-20. Program Screen Items (1/3)

    User Manual EC1000S b) PROGRAM (sequence programming) This “PROGRAM” screen is used to program sequences. The items listed below in Table 5-20 to Table 5-22 are set for individual steps. Sequence programming is stored automatically, and the data for AC mode with a 100 V range or a 200 V range is stored separately from the data for AC+DC mode with a 100 V range or a 200 V range.
  • Page 177: Table 5-21. Program Screen Items (2/3)

    User Manual EC1000S Table 5-21. PROGRAM Screen Items (2/3) Setting Name Item Operation Sets the AC frequency. Setting range: 1.0 Hz to 550.0 Hz AC frequency Setting resolution: 0.1 Hz Sets the method used to change to the specified frequency.
  • Page 178: Table 5-22. Program Screen Items (3/3)

    User Manual EC1000S Table 5-22. PROGRAM Screen Items (3/3) Setting Name Item Operation Sets operation after end of step. Select one of the following three options.  CONT: Sequence continuation STEP TERM Step end  STOP: Change to Idle mode ...
  • Page 179: Memory (Memory)

    User Manual EC1000S 5.5.2 Memory (MEMORY) The “MEMORY” screen is used to control memory. Settings saved (stored) to store/recall memory No. 1 to No. 30 can be recalled and used. Recall of store/recall memory contents can be executed only when output is off.
  • Page 180: System (System)

    CANCEL ENTER 5.5.4 System (SYSTEM) The EC1000S system settings are displayed in the “SYSTEM” screen. See “5.6 System Menu”, for detailed description of the system menu. 5.5.5 Setting range limiter (LIMIT) The setting range limiter indicates the LIMIT screen, enabling the setting range of output voltage and output...
  • Page 181: Selection Of Measurement Display (Measure)

     The same setting range limit is used for all output modes. Not only the currently selected output mode setting, but the settings for all other modes are restricted by this limit. (The EC1000S stores the setting for each output mode.) The following describes an example where the selected output mode is AC-INT and the limit is set to prevent the output voltage from exceeding 100.0 Vrms.
  • Page 182 User Manual EC1000S  Operation steps key to move the cursor to “OUTPUT”. Press the CANCEL ENTER E NT E R Press the key to display the selection box. CANCEL ENTER key to move the cursor to “ON”. Press the...
  • Page 183: System Menu

    User Manual EC1000S 5.6 System Menu This menu is used to enter the following system settings for the EC1000S.  KEYLOCK: Keylock  BEEP: Beep  CONTRAST: LCD contrast  COLOR: LCD display color  TIME UNIT: Time unit  INIT OUT: Output on/off at power-on ...
  • Page 184 User Manual EC1000S CANCEL ENTER E NT E R Press the key to display the SYSTEM screen . CANCEL ENTER Press the key to select the operation to be performed. California Instruments...
  • Page 185: Keylock

    EC1000S 5.6.1 Keylock The EC1000S is equipped with a function that prevents key input errors during operation. When “keylock” is on, all keys become inactive except for the keylock off key. Although it is possible to switch among menus when keylock is on, basically no changes can be made except for turning keylock off. However, output can still be turned off via the OUTPUT key as an emergency response measure.
  • Page 186: Beep Sound

    User Manual EC1000S E NT E R CANCEL Press the key to complete the setting. Press the key to cancel the setting. 5.6.2 Beep sound When the beep sound is on, a beep sounds when an incorrect operation is made, a key is pressed, or a modify operation is made.
  • Page 187: Lcd Contrast Adjustment

    User Manual EC1000S CANCEL ENTER E NT E R CANCEL 4. Press the key to complete the setting. Press the key to cancel the setting. 5.6.3 LCD contrast adjustment The LCD screen contrast level can be set to any of 100 levels to suit the operation environment.
  • Page 188: Lcd Display Color Setting

    User Manual EC1000S Press the keys to move the cursor to the column to be changed, turn the MODIFY dial and set a numerical value. The set value will be displayed immediately. Decrease Increase E NT E R CANCEL 4. After setting all of the columns, press the key to close the modification box.
  • Page 189: Time Unit Setting

    User Manual EC1000S Selection box CANCEL ENTER key to move the cursor to “BLUE” for blue based or “WHITE” for white based. Press the CANCEL ENTER E NT E R CANCEL 4. Press the key to complete the setting. Press the key to cancel the setting.
  • Page 190: Output On/Off Setting At Power-On

    5.6.6 Output on/off setting at power-on The EC1000S can be set to automatically turn output on at power-on. However, when restarting after a system lock, the EC1000S will start with output off regardless of this setting.  Operation steps key to move the cursor to “INIT OUT”.
  • Page 191: External Control Input Enable/Disable Setting

    5.6.7 External control input enable/disable setting External control input can be set as enabled or disabled. However, the EC1000S status will be output even when external control input is set as disabled. See “4.5 Control Using External Control I/O Connector”, for details of external control input.
  • Page 192: Reset Function

    This function resets all basic settings to their initial (factory) settings. The EC1000S cannot be reset when output is on. Always turn output off before resetting the EC1000S. See “3.2 Display and Initial Settings at Power-on”, for description of the settings after a reset.
  • Page 193: System Information

    CANCEL ENTER EXEC box E NT E R After selecting “EXEC”, press the key to reset the EC1000S. 5.6.9 System information Select “INFORMATION” in the “SYSTEM” screen to view system information such as the firmware version number. See “8.4 Checking the Version Number”, for description of the operation steps for viewing this version number.
  • Page 195: Usb Interface

    User Manual EC1000S 6. USB INTERFACE Setup for Use ........................192 Command List ........................ 193 Command Descriptions ....................199 6.3.1 Overview of Programming Language ................199 6.3.2 Detailed command descriptions ..................203 Command Tree ....................... 234 Status System ......................... 236 6.5.1...
  • Page 196: Setup For Use

    EC1000S 6.1 Setup for Use The EC1000S can be controlled under the USB Test and Measurement Class (USBTMC) standard. Almost all control functions can be operated from the control panel, and internal status information such as settings and error information can be read.
  • Page 197: Command List

    EC1000S 6.2 Command List The EC1000S command list is shown in Table 6-1 to Table 6-9 and IEEE488.2 (SCPI) common commands supported by the EC1000S are listed in Table 6-10. The symbols used in Table 6-1 to Table 6-10 are explained below. The lowercase letters in each keyword can be omitted.
  • Page 198: Table 6-2. Command List (Measure Subsystem)

    User Manual EC1000S DC output voltage [SOURce:]VOLTage:OFFSet[:IMMediate] (during AC+DC-INT, AC+DC-ADD, or AC+DC-SYNC mode) Output voltage range [SOURce:]VOLTage:RANGe Table 6-2. Command List (MEASure Subsystem) Function Command Output current RMS value MEASure[:SCALar]:CURRent[:AC]? + current peak value MEASure[:SCALar]:CURRent:HIGH?  current peak value MEASure[:SCALar]:CURRent:LOW?
  • Page 199: Table 6-4. Command List (Status Subsystem)

    User Manual EC1000S Table 6-4. Command List (STATus Subsystem) Function Command Operation condition register STATus:OPERation:CONDition? Operation event enable register STATus:OPERation:ENABle Operation event register STATus:OPERation[:EVENt]? STATus:OPERation:NTRansition Operation transition filter STATus:OPERation:PTRansition Warning event condition register STATus:WARNing:CONDition? Warning event enable register STATus:WARNing:ENABle Warning event register...
  • Page 200: Table 6-8. Command List (System Subsystem)

    User Manual EC1000S Table 6-8. Command List (SYSTem Subsystem) Function Command Beep sound on/off SYSTem:BEEPer:STATe External control input enable/disable SYSTem:CONFigure:EXTIO Error query SYSTem:ERRor? Output at power-on SYSTem:PON[:OUTPut] Time unit for sequence step execution SYSTem:TUNit Clear warning SYSTem:WRELease Table 6-9. Command List (PROGram Subsystem)
  • Page 201: Table 6-10. Common Command List (Common Commands And Queries)

    User Manual EC1000S Table 6-10. Common Command List (Common Commands and Queries) Common Commands and Name Function Queries *CLS Clear command Clears standard event register, etc. Standard event enable *ESE Sets standard event enable register. command *ESR? Standard event register query Queries standard event register.
  • Page 202 User Manual EC1000S a) Input buffer  The number of commands that can be sent at one time is limited by the input buffer capacity (10,000 characters).  The transferred commands are first stored in the input buffer, then successively interpreted and executed.
  • Page 203: Command Descriptions

    Refer to the following for general information on SCPI. http://www.scpiconsortium.org Standard Commands for Programmable Instruments (SCPI) VERSION 1999.0 b) SCPI data format All command messages and acknowledge messages sent to the EC1000S are in “ASCII” format. Table 6-11. Numerical Value Data Format Symbol Data Format <NR1>...
  • Page 204: Figure 6-1. Binary Block Data

    User Manual EC1000S Non-zero Numerals 8-bit data byte numerals Figure 6-1. Binary Block Data The “non-zero numeral” element value indicates the number of “numeral” elements that follow afterward. The “numeral” element value indicates the number of “8-bit data byte” elements that follow afterward.
  • Page 205: Figure 6-2. Partial Command Tree

    User Manual EC1000S Figure 6-2. Partial Command Tree  Root level The command tree top level is the root. With SCPI, the current path is set to the root, as shown below. ○ Immediately after the device power source goes on ○...
  • Page 206 User Manual EC1000S 3) Abbreviation of command string The command syntax enables each command (including certain parameters) to be written as a combination of uppercase and lowercase letters. The uppercase letters in the command can be used as an abbreviation of the command.
  • Page 207: Detailed Command Descriptions

    A triangular symbol is used to explicitly indicate that a space exists. b) Remark  Commands and queries are both included as EC1000S commands, but for simplicity both are referred to as commands. Any command that includes a question mark (?) with a keyword is a query.
  • Page 208 User Manual EC1000S  [SOURce:]CURRent:LIMit:PEAK:HIGH This sets the upper limit of the output peak current (positive polarity). The output peak current limiter (positive polarity) setting ranges from 10.0 to 42.0 A (for 100 V range) or from 5.0 to 21.0 A (for 200 V range).
  • Page 209 User Manual EC1000S  [SOURce:]FREQuency:LIMit:HIGH This sets the upper limit of the frequency that can be set. However, a value that is at or below the currently set frequency cannot be set. ([5, “Out of Limiter”] error occurs.) The limit value setting ranges from 1.0 to 550.0 Hz.
  • Page 210 User Manual EC1000S  [SOURce:]MODE This sets the output mode. There are eight output modes, each of them is a combination of one operation mode (AC or AC+DC) and one signal source mode, (INT, EXT, ADD, or SYNC). Consequently, the output mode setting is AC-INT, AC-EXT, AC-ADD, AC-SYNC, AC+DC-INT, AC+DC-EXT, AC+DC-ADD, or AC+DC-SYNC.
  • Page 211 User Manual EC1000S  [SOURce:]PHASe[:IMMediate] This sets the phase at the start of output. The phase setting range is 0.0 to 359.9. However, when output is on, the phase setting cannot be changed. ([1, “Invalid with output on”] error occurs.) When in AC-EXT or AC+DC-EXT mode, the output waveform setting cannot be set.
  • Page 212 User Manual EC1000S Parameters: None Query example: SEQ:LEN? Response format: <NR1> Response example:  [SOURce:]SEQuence:COUNt? This returns the jump times of the step specified by SEQ:STEP. The step jump times response ranges from 1 to 999. 1 is returned in the idle mode.
  • Page 213 User Manual EC1000S For the sine wave (SIN) and square wave (SQU), it is Vrms, and for the arbitrary waveform (ARB) it is Vp-p. When in AC-EXT, AC-ADD, AC-SYNC, AC+DC-EXT, AC+DC-ADD, or AC+DC-SYNC mode, step execution parameters cannot be set or queried. ([3, “Invalid in this mode”] error occurs.) Syntax: [SOURce:]SEQuence:EPARameter <NR2>, <NR1>..
  • Page 214 User Manual EC1000S Range 0 (CONST) or 1 (KEEP) 11th parameter (step synchronized output) Type <NR1> Range 0 (00), 1 (01), 2 (10), or 3 (11) 12th parameter (step synchronized output operation type) Type <NR1> Range 0 (CONST) or 1 (KEEP) Command example: SEQ:EPAR 0, 0, 100.0, 0, 50.0, 0, 1, 0, 90.0, 0, 0, 0...
  • Page 215 User Manual EC1000S  [SOURce:]SEQuence:TPARameter This sets the step transition parameters. The target sequence step must be set in advance by the SEQ:STEP. ( “[SOURce:]SEQuence:STEP” on page 212.) See “ SYSTem:TUNit” on page 6-36) in advance to set the unit for the Use SYST:TUN ( sequence step execution time or query the unit.
  • Page 216 User Manual EC1000S  [SOURce:]SEQuence:STEP Specifies the target step number for sequence editing. The step number specification ranges from 1 to 255. Syntax: [SOURce:]SEQuence:STEP <NR1> Parameters: 1 to 255 Command example: SEQ:STEP 1 Query example: SEQ:STEP? Response format: <NR1> Response example: ...
  • Page 217 User Manual EC1000S Response example: 200.0 California Instruments...
  • Page 218 User Manual EC1000S  [SOURce:]VOLTage:LIMit:LOW This sets the upper limit of the output voltage that can be set. However, a value that is at or below the currently set cannot be specified. ([5, “Out of Limiter”] error occurs.) For the 100 V range, the limit value setting ranges from 200.0 to 5.0 V, and for the 200 V range, it ranges from 400.0 to 10.0 V.
  • Page 219 User Manual EC1000S d) MEASure subsystem  MEASure[:SCALar]:CURRent[:AC]? This returns the output current RMS value (AC+DC RMS value) in Arms units. The output current RMS value measurement full scale is ±15.00 Arms. If the measured value exceeds the full scale, 99.99 Arms is returned as a measurement error.
  • Page 220 User Manual EC1000S  MEASure[:SCALar]:CURRent:AMPLitude:MAX? This returns the output current peak value (hold) in Apk units. The held output current peak value full scale is 45.0 Apk. If the measured value exceeds the full scale, 99.9 Arms is returned as a measurement error.
  • Page 221 User Manual EC1000S Query example: MEAS:CURR:CRES? Response format: <NR2> Response example: 1.41  MEASure[:SCALar]:CURRent:HARMonic[:AMPLitude]? This returns the harmonic current in Arms units within the specified harmonic order range. The harmonic current measurement full scale is 15.00 Arms. If the measured value exceeds the full scale, 99.99 Arms is returned as a measurement error.
  • Page 222 User Manual EC1000S Syntax: MEASure[:SCALar]:FREQuency? Parameters: None Query example: MEAS:FREQ? Response format: <NR2> Response example: 50.4 California Instruments...
  • Page 223 User Manual EC1000S  MEASure[:SCALar]:POWer:AC:APParent? This returns the apparent power in VA units. The apparent power measurement full scale is 1400 VA. If the measured value exceeds the full scale, 9999 VA is returned as a measurement error. Syntax: MEASure[:SCALar]:POWer:AC:APParent?
  • Page 224 User Manual EC1000S Parameters: None Query example: MEAS:POW:AC? Response format: <NR1> Response example:  MEASure[:SCALar]:VOLTage[:AC]? This returns the output voltage RMS value (AC+DC RMS value) in Vrms units. The output voltage RMS value measurement full scale is 450.0 Vrms. If the measured value exceeds the full scale, 999.9 Vrms is returned as a measurement error.
  • Page 225 User Manual EC1000S  MEASure[:SCALar]:VOLTage:AVErage? This returns the output voltage average value (AC+DC average value) in V units. The output voltage average value measurement full scale is ±450.0 V. If the measured value exceeds the full scale, 999.9 V is returned as a measurement error.
  • Page 226 User Manual EC1000S f) STATus subsystem  STATus:OPERation:CONDition? This returns the operation condition register (OPCR). The response ranges from 0 to 20482. Syntax: STATus:OPERation:CONDition? Parameters: None Query example: STAT:OPER:COND? Response format: <NR1> Response example: 16384  STATus:OPERation:ENABle This sets the operation event enable register (OPEE). The setting ranges from 0 to 20482.
  • Page 227 User Manual EC1000S  STATus:OPERation:NTRansition STATus:OPERation:PTRansition This sets the operation transition filter (OPTF). The setting ranges from 0 to 20482.  When the NTR filter is set to 1, OPER is set to 1 when OPCR changes from 1 to 0.
  • Page 228 User Manual EC1000S  STATus:WARNing:ENABle This sets the warning event enable register. The setting ranges from 0 to 31999. Syntax: STATus:WARNing:ENABle <NR1> Parameters: 0 to 31999 Command example: STAT:WARN:ENAB 1 Query example: STAT:WARN:ENAB? Response format: <NR1> Response example:  STATus:WARNing[:EVENt]? This returns the warning event register value.
  • Page 229 User Manual EC1000S g) OUTPut subsystem  OUTPut[:STATe] This controls on/off settings of the output relay. Syntax: OUTPut[:STATe] <bool> Parameters: OFF|0 / ON|1 Command example: OUTP ON Query example: OUTP? Response format: <NR1> Response example: h) INPut subsystem  INPut:GAIN This sets the external input gain used in the external mode (AC-EXT or AC+DC-EXT) and internal + external mode (AC-ADD or AC+DC-ADD).
  • Page 230 User Manual EC1000S i) TRACe subsystem  TRACe:CATalog? This returns a list of defined waveform names. This command returns a list of all waveform names that can be selected by the user. Syntax: TRACe:CATalog? Parameters: None Query example: TRAC:CAT? Response format: <CRD>[, <CRD>]...
  • Page 231 User Manual EC1000S  TRACe:DELete[:NAME] This clears the specified arbitrary waveform memory. When this command is executed, a sine wave is written to ARB1 to ARB8, a square wave is written to ARB9 to ARB12, and a triangle wave is written to ARB13 to ARB16.
  • Page 232 User Manual EC1000S  SYSTem:ERRor? This reads one error from the error queue (FIFO), and clears the read error in the error queue. The maximum number of errors that can be held in the error queue is 20. If 21 or more errors occur, [-350, “Too Many Errors”] is inserted as the 20th error.
  • Page 233 User Manual EC1000S  SYSTem:WRELease This clears a warning. When a warning is cleared, the SET screen is displayed and key input is enabled. In order for the warning to be cleared, it is necessary that all the factors of the warning has been removed and it is ready to be cleared.
  • Page 234 User Manual EC1000S l) Common commands  CLS This clears the following registers.  Standard event register  Operation event register  Warning event register  Status byte  Error queue CLS Syntax: Parameters: None CLS Command example: Response format:...
  • Page 235 Syntax: Parameters: None IDN? Query example: Response format: <SRD> “NF Corporation, EC1000S, 000001, Ver1.00” Response example:  OPC ※ When all operations have been completed, this sets the standard event register OPC bit (BIT0). OPC? returns 1 when all operations have been completed. However, the standard event register OPC bit is not cleared to zero when OPC? is executed.
  • Page 236 User Manual EC1000S  RST This resets the device to its factory settings. However, reset cannot be executed when output is on. ([1, “Invalid with output on”] error occurs.) RST Syntax: Parameters: None RST Command example: Response format: None Response example: None ...
  • Page 237 User Manual EC1000S  TST? This returns the self test results. In this device, “0” is always returned. TST? Syntax: Parameters: None TST? Query example: Response format: <NR1> Response example:  WAI ※ When all operations are completed, the device is not able to execute any subsequent commands.
  • Page 238: Command Tree

    User Manual EC1000S 6.4 Command Tree The command tree for each of the EC1000S subsystem is shown in Figure 6-3. California Instruments...
  • Page 239: Figure 6-3. Command Tree

    User Manual EC1000S Figure 6-3. Command Tree California Instruments...
  • Page 240: Status System

    User Manual EC1000S 6.5 Status System The EC1000S includes the minimum set of status groups defined by IEEE488.2. The EC1000S status system is outlined below in Figure 6-4. Not used (always 0) Warning status summary Not used (always 0) Not used (always 0)
  • Page 241: Status Byte

    User Manual EC1000S 6.5.1 Status byte The status byte register definitions are listed in Table 6-13. Bits in the status byte register become valid when 1 is set in the service request enable register, then the ORed result of the valid bits is taken to issue a service request.
  • Page 242: Standard Event Status

    User Manual EC1000S 6.5.2 Standard event status The structure of the standard event status register is shown in Figure 6-5 below. ESR (standard event register) ESE (standard event enable register) Power on (PON) User request (URQ) Command error (CME) Execution error (EXE)
  • Page 243: Table 6-14. Standard Event Status Register Definitions

    User Manual EC1000S Table 6-14. Standard Event Status Register Definitions Weight Description Power on bit 1 is set to this bit when the power is turned on. PON (7) When this register is read, this bit is cleared to 0, and it remains 0 until the power is turned on again.
  • Page 244: Operation Status

    User Manual EC1000S 6.5.3 Operation status The structure of the operation status register is shown in Figure 6-6 below. OPCR (operation condition register) OPTF (operation transition filter) Always 0 Under sequence execution Not used Under sequence hold Not used Not used...
  • Page 245: Warning Status

    User Manual EC1000S 6.5.4 Warning status The structure of the warning status register is shown in Figure 6-7 below. WRCR (warning condition register) WRTF (warning transition filter) Always 0 Peak current limiter operation Average current limiter operation Power limiter operation...
  • Page 246: Error Message List

    User Manual EC1000S 6.6 Error Message List USB error messages are listed in Table 6-15 below. Table 6-15. USB Error Message List Message Cause “No error” No error. Command was recognized as invalid due to any cause. 100 “Command error”...
  • Page 247: Programming Precautions

    User Manual EC1000S 6.7 Programming Precautions a) Arbitrary waveform data Arbitrary waveform data comprises a block of binary data. The preceding ASCII program message command field and the binary data block arbitrary waveform data should be transferred at the same time.
  • Page 249: Troubleshooting

    User Manual EC1000S 7. TROUBLESHOOTING Protection Function ....................... 246 Error Messages and Responses .................. 248 7.2.1 Error at power-on ......................249 7.2.2 Protection function-related errors .................. 251 7.2.3 Panel operation errors ..................... 252 7.2.4 Warning messages ......................254 7.2.5 USB-related external control errors ................256 When Fault Is Suspected....................
  • Page 250: Protection Function

    See “7.2.4 Warning messages”, for description of warning messages. See “5.6.2 Beep sound”, for description of the beep sound. The EC1000S checks for errors in the order those listed in Table 7-1 and only the message of the error detected first will be displayed.
  • Page 251: Table 7-1. Protection Function

    User Manual EC1000S Table 7-1. Protection Function Action When Detected Protection Displayed Message System Output off Lock Internal power source error message Internal power   “7.2.2 Protection function-related errors”. source error Internal Internal communication error message 1  ...
  • Page 252: Error Messages And Responses

    EC1000S 7.2 Error Messages and Responses The EC1000S performs a self test at power-on, and outputs an error message if any abnormalities are detected. An error message is also displayed when a setting or operation is performed incorrectly or when an abnormality is detected in output or internally.
  • Page 253: Error At Power-On

    User Manual EC1000S 7.2.1 Error at power-on The EC1000S performs the following fault checks at power-on. If an error occurs, NG is displayed on the system check screen. Nothing is displayed when all check items are OK. ROM CHECK...... OK RAM READ/WRITE CHECK ...
  • Page 254 User Manual EC1000S Instruments or its representative should be contacted. Backup memory, sequence memory, and VERSION CHECK Version number check waveform memory are initialized, then the system is started. California Instruments...
  • Page 255: Protection Function-Related Errors

    User Manual EC1000S 7.2.2 Protection function-related errors When a protection function-related error has occurred, an error message is displayed in the message window, and operation is stopped. Figure 7-2. Screen Display When Protection Function-Related Error Has Occurred When a protection function-related error message is displayed, turn off the power switch. ( “7.1 Protection Function”.)
  • Page 256: Panel Operation Errors

    User Manual EC1000S 7.2.3 Panel operation errors Panel operation error messages are displayed in the status window. The message box closes automatically after 1.5 seconds. Message box Figure 7-3. Screen Display When Panel Operation Error Has Occurred Table 7-4. Panel Operation Errors (1/2)
  • Page 257: Table 7-5. Panel Operation Errors (2/2)

    User Manual EC1000S output mode. See “3.4.2 ”.) Setting output mode Table 7-5. Panel Operation Errors (2/2) Error Message Cause and Action Required Cannot set (execute) due to BUSY status. Wait until the BUSY display disappears, then try operation again.
  • Page 258: Warning Messages

    User Manual EC1000S 7.2.4 Warning messages When output is turned off due to an output error, internal temperature error, or external sync signal See “7.1 frequency out of range error, a warning is displayed in the message window. ( Protection Function”.) Figure 7-4.
  • Page 259: Table 7-7. Warning Messages (2/2)

    User Manual EC1000S Table 7-7. Warning Messages (2/2) Message Cause and Action Required Output off due to the output peak current protection Output OFF. function. caused by: Press the ENTER key or CANCEL key. The error is cleared and the system becomes available. Check OVER PEAK CURRENT settings and load conditions before using.
  • Page 260: Usb-Related External Control Errors

    User Manual EC1000S 7.2.5 USB-related external control errors Table 7-6 lists USB-related external control error messages. Table 7-8. USB Error Message List Message Text Cause “No error” No error. Command was recognized as invalid due to any cause. 100 “Command error”...
  • Page 261: When Fault Is Suspected

    7.3 When Fault Is Suspected If a problem that might be due to a fault occurs when using the EC1000S, check the following table to determine whether the problem might be due to an operation, use method, connection error, etc.
  • Page 262: Table 7-11. When Fault Is Suspected (Problem Related To Output Voltage Or Output Voltage Range Setting)

    User Manual EC1000S Table 7-11. When Fault Is Suspected (Problem Related to Output Voltage or Output Voltage Range Setting) Symptom Possible Cause Required Action, etc. Turn power switch off. Output voltage Fault in internal power source If the output voltage range cannot be...
  • Page 263: Table 7-13. When Fault Is Suspected (Problem Related To Output Error)

    User Manual EC1000S See “3.4.2 ”. Setting output mode See “4.7.2 ”. Line synchronization See “5.3.1 Output mode setting”. Table 7-13. When Fault Is Suspected (Problem Related to Output Error) Symptom Possible Cause Required Action, etc. Check the output voltage setting.
  • Page 264: Table 7-14. When Fault Is Suspected (Problem Related To Measurement Functions)

    User Manual EC1000S Something is blocking the air Make sure the EC1000S complies with the flow near the air intake on the installation requirements described in “2.2 front panel or the air exhaust ”. Installation Environment on the rear panel.
  • Page 265: Table 7-16. When Fault Is Suspected (Problem Related To Memory Function)

    DC voltage is not generated, but Set AC (AC) mode. due to control limitation, the When in AC-INT mode, the EC1000S DC Transformer is used voltage cannot be reduced all offset voltage is ±...
  • Page 267: Maintenance

    User Manual EC1000S 8. MAINTENANCE Introduction ........................264 Routine Maintenance ..................... 264 Storage, Repackaging, and Transport ................. 266 Checking the Version Number ..................267 Backup Battery ....................... 269 California Instruments...
  • Page 268: Introduction

    If these operation checks are not passed, request calibration and/or repair from California Instruments or its representative. 8.2 Routine Maintenance Be sure to install the EC1000S in a location that meets the installation requirements. See “2.2 Installation Environment”, for description of the installation requirements. a) When panel or case is dirty Wipe clean with a soft cloth.
  • Page 269: Figure 8-1. Air Filter Cleaning Steps

    User Manual EC1000S b) When fan filter is dirty Two air filters are provided to filter out dust that gathers through the air inlets in the front panel. When too much dirt accumulates in the filters, it can clog them and reduce air flow, causing higher internal temperatures which can reduce reliability.
  • Page 270: Storage, Repackaging, And Transport

    User Manual EC1000S 8.3 Storage, Repackaging, and Transport Store the EC1000S in a location that meets the installation requirements. See “2.2 Installation Environment”, for description of installation requirements. a) Storage for long periods of non-use  Remove the power cord from the outlet and the main unit.
  • Page 271: Checking The Version Number

    EC1000S 8.4 Checking the Version Number The EC1000S firmware version is displayed in the bottom right of the screen for a few seconds after power-on. At other times, the version information can be checked via the control panel system information screen, as described below.
  • Page 272: Figure 8-2. System Information Screen

    User Manual EC1000S CANCEL ENTER key to move the cursor to “INFORMATION”. Press the CANCEL ENTER E NT E R key to display the “SYSTEM INFORMATION” screen. Press the Figure 8-2. SYSTEM INFORMATION Screen Press the TOP MENU key to return to the SET menu screen.
  • Page 273: Backup Battery

    See “4.8 Using Memory Functions”) and other required parameters The contents of memory ( and data are backed up by a lithium battery in the EC1000S. The backup period varies according to the temperature and use conditions, but it is approximately five years when stored with the power off.
  • Page 275: Specifications

    User Manual EC1000S 9. SPECIFICATIONS Output..........................272 Current Limiter ....................... 276 Setting Range Limits ..................... 276 Signal Sources ....................... 277 Measurement Functions ....................278 Sequence Function ......................281 Arbitrary Waveform Memory ..................282 Setting Memory ......................282 Protection Functions ..................... 282 9.10...
  • Page 276: Output

    User Manual EC1000S Unless otherwise noted, the following conditions are adopted: Power input: 200 V AC, 50 Hz Output mode: AC-INT (alternate current - internal signal source) mode Output frequency: 50 Hz Output voltage: 100 Vrms (200 Vrms when in 200 V range)
  • Page 277 User Manual EC1000S 270 Vrms (AC) / 380 V (DC) or lower (200 V range)  Output terminals Terminal block with M4 screws (rear panel) AC receptacle (universal type, front panel) Note: Use AC receptacle for AC (0 V AC to 250 V AC). When DC is included, use screw terminal block on rear panel.
  • Page 278 User Manual EC1000S the maximum output current. Note 3: When output frequency is 40 Hz or lower, or it is 400 Hz or higher, the maximum output current may decrease. Note 4: The RMS value for AC + DC is within the maximum output current.
  • Page 279 User Manual EC1000S +1 dB, 3 dB 100 Hz to 550 Hz:  Output voltage distortion factor 0.5% or less (50 Hz or 60 Hz)  Load regulation 0.5% or less (0 to 100%, via output terminal)  Line regulation 0.2% or less...
  • Page 280: Current Limiter

    User Manual EC1000S 9.2 Current Limiter  Output peak current limiter When limiter is operating, output voltage is clipped. Positive current setting range 100 V range: +10.0 A to +42.0 A (Initial value: +42.0 A) 200 V range: +5.0 A to +21.0 A (Initial value: +21.0 A)
  • Page 281: Signal Sources

    User Manual EC1000S 9.4 Signal Sources The signal sources that can be selected are internal (INT), external signal input (EXT), internal + external (ADD), and external synchronization (SYNC).  Internal signal source (Internal source mode, internal + external source mode) See “9.1 Output”.
  • Page 282: Measurement Functions

    User Manual EC1000S Measurement Functions Measurement accuracy applies to 50Hz-400Hz output unless otherwise noted.  Output voltage measurement RMS value: RMS value of AC+DC Full scale: 225.0 Vrms: 100 V range 450.0 Vrms: 200 V range Display resolution: 0.1 Vrms Measurement accuracy: ±0.5% of the full scale (at 23 ±5°C)
  • Page 283 User Manual EC1000S Measurement accuracy: ±3% of the full scale (at 23 ±5°C, sine wave)  + peak current  and   peak current |maximum value Peak value hold: Full scale: 45.0 Apk Display resolution: 0.1 Apk Measurement accuracy: 3% of the full scale (at 23 5°C, sine wave) ...
  • Page 284 User Manual EC1000S Display resolution: 0.01 Arms and 0.1% Measurement accuracy: 1% of full scale at RMS value current (at 23 ±5°C, to 20th harmonic) ±1.5% of full scale at RMS value current (at 23 ±5°C, from 20th to 40th) Note: This measurement does not comply with the IEC standards.
  • Page 285: Sequence Function

    User Manual EC1000S 9.6 Sequence Function Output parameters can be changed abruptly or swept in sequence. Signal source is internal signal source only. Sequence function settings are stored in memory with battery backup.  Number of sequences 1 sequence per operation mode (AC or AC+DC mode) and output range (100 V or 200 V) ...
  • Page 286: Arbitrary Waveform Memory

    User Manual EC1000S 9.7 Arbitrary Waveform Memory The contents of arbitrary waveform memory are retained by battery backup.  Waveform memory count  Waveform length 4096 words  Waveform data 15 bits  Arbitrary waveform memory cannot be written to from the panel screen. USB interface is used for writing.
  • Page 287: General

    User Manual EC1000S 9.10 General  LCD settings Contrast: Adjustable Display color: Blue background or white background  Beep sound On or off When on, a beep sound will be output if a key input error or other operation error occurs. When a protection function-related error occurs, an alarm sound will be output regardless of the beep sound setting.
  • Page 288: External Control I/O

    User Manual EC1000S 9.11 External Control I/O  External control operation mode Enabled or disabled (status output is continually output)  Control input Input level High level: +4.0 V or above, Low level: +1.0 V or below Nondestructive maximum input: +10 V/5 V Input impedance +5 V pull-up with 47 k...
  • Page 289: Withstand Voltage And Insulation Resistance

    User Manual EC1000S 0.1 Ω or less (not including the power cord set) Protective ground impedance: 9.14 Withstand Voltage and Insulation Resistance Between input power line vs. output/case or input power line/case vs. output  Withstand voltage 1500 V AC ...
  • Page 290: Figure 9-1. Temperature And Humidity Ranges

    User Manual EC1000S Storage condition Performance guarantee Operation guarantee 10 50 C Figure 9-1. Temperature and Humidity Ranges 9.17 External Dimensions and Weight  External dimensions Width: 258 mm Height: 176 mm Depth: 440 mm (Projections are not included.) ...
  • Page 291: Figure 9-2. External Dimensions

    User Manual EC1000S Figure 9-2. External Dimensions California Instruments...