Signal Measurements
Push the MEASURE button to display the automated measurement
menu. The new measurements are in the Type menu. The TDS2MM
Module automatically determines the 10%, 50%, and 90% waveform
points.
Type
Rise Time
Fall Time
Pos Width
Neg Width
TDS 200-Series Extension Module Instructions
TDS2MM Measurement Module
Definition
Measures the time between 10% and 90% of the first
rising edge of the waveform. Rising edge must be
displayed to measure.
Measures the time between 90% and 10% of the first
falling edge of the waveform. Falling edge must be
displayed to measure.
Measures the time between the first rising edge and the
next falling edge at the waveform 50% level. Rising and
falling edges must be displayed to measure.
Measures the time between the first falling edge and the
next rising edge at the waveform 50% level. Falling and
rising edges must be displayed to measure.
29