14.7 Troubleshooting

14.7.1 Error Messages

See "Error Messages" on page 73 for information relating to error
messages.

14.7.2 Common Problems and Solutions

Analog Output Incorrect
Make certain of the type of Analog Output (voltage or current)
that the channel is equipped with; see "Identifying the Output
Type" on page 68.
Use the DIAGNOSTICS -> DAC FULL SCALE and DIAGNOSTICS -> DAC
ZERO SCALE menu items to force the Analog Output to a known
output. Then, adjust your readout device, panel meter, PLC or
data acquisition system to match.
Check the output selection jumpers; see "Output Selection" on
page 68.
Check the "DAC SETUP -> DAC. ZERO-SCALE" and "DAC SETUP ->
DAC FULL-SCALE" menu items.
Sensitivity to EMI/RFI
To obtain maximum immunity to electromagnetic or radio fre-
quency interference, make certain that the shield of the cable is
connected to ground/earth only at the readout device. The shield
of the cable should not be connected to the instrument.
Honeywell Sensing and Control
69