Program Mode
1. Press OPTS
until PROGRAM MODE is
displayed
2. Press ON
.
3. Enter the rate for Step 1 and press OK
4. Enter the volume for Step 1 and press OK
5. Repeat Steps 3 and 4 until all steps are
complete and press OK
.
6. Review the Total VTBD and press OK
7. Press START
.
Ramp Taper Mode
1. Press OPTS
until RAMP TAPER MODE is
displayed
2. Press ON
.
3. Enter the total bag volume to be delivered and
press OK
.
4. Enter the Ramp Up time in hours and minutes
and press OK
.
5. Enter the Taper Down time in hours and minutes
and press OK
.
6. Enter the Total Time for the infusion in hours
and minutes and press OK
.
7. Review the calculated Max Rate and
press OK
.
8. Press START
.
Alarm Troubleshooting
Air In Line!
.
.
Battery
Too Low!
.
Door Open!
Dwnstream
Occlusion!
Pump
Stopped!
Silence alarms by pressing
Refer to Instructions for Use for complete
operating instructions.
• Check for air in air-in-line
detector.
• Check that tubing is properly
inserted in air-in-line detector.
• Remove air from line.
• Plug pump into wall outlet.
• Close door.
• Open clamps.
• Check for kinked tubing.
• Check IV site.
• Check occlusion pressure
setting.
• Pump is stopped.
• Either start infusion or press
to extend hold time
3 minutes.
.
B. Braun Medical Inc.
824 Twelfth Avenue
Bethlehem, PA 18018
1-800-227-2862
www.bbraunusa.com
™
Vista
basic
Infusion Pump
Quick
Reference
Rx only
Vista™ and Working Together for Excellence in Safety
are a trademarks of B. Braun Medical Inc.
©2002, B. Braun Medical Inc.
950853-Rev B 06/02
Need help?
Do you have a question about the Vista basic and is the answer not in the manual?
Questions and answers